OpenOSC 0.21.5

Choose a tag to compare

@bve bve released this 23 Nov 13:34
· 2 commits to main since this release

Warning

Work in progress, for enthusiasts only. ZT703S is the main development and hardware test device.
Hardware coverage for the ZT706S is limited.
Install at your own risk. Return to the stock firmware.

What's new

Scope panning and zoom (#106, #109, #111)

  • Panning follows the next frame when a running capture is reusing the previous frame's memory. At 512K points, this removes the misleading No record to move popup during normal acquisition.
  • Zoom & Pan reports a zoom limit when the requested span cannot fit in the record. If a later capture is shorter than the existing zoomed-out view, the screen shows ZOOM all.
  • The ZOOM and PAN indicators remain visible after leaving Zoom & Pan with Back. Successful Auto Setup returns the view to its default zoom and pan.

Battery display calibration (#107)

Settings → Device → Power has Battery gain and Battery offset controls. They let you align the displayed voltage with an external meter and also adjust the estimated charge percentage and bars. The low-battery shutdown threshold uses the factory reading.

Tooltip controls (#108)

Choose which tooltip categories to show in Settings → Display → Tooltips. Usage tips, unavailable actions, progress, completions and advisory warnings each have a switch. Critical alerts are always shown.

Analog multimeter scale (#112)

Scale labels use the size of the displayed digits to clear the arc, tick marks and needle. This also keeps longer Type K labels readable.

Thanks to @art-thunder, @OlegMedvedcuk, @Fyfar and @LeonIndman for the reports and suggestions.

Testing

Verified the 512K panning fix, Zoom & Pan limits, Auto Setup, tooltip controls, battery calibration and analog multimeter scales on the ZT703S during development. The reported 0–1500 °C dial was also checked in a rendering test. The 0.21.5 firmware was installed and checked in Scope and DMM modes on the ZT703S. The native DMM model, settings layout, scope controller and rendering suites passed. Both models passed the application and Recovery builds, package validation and stack checks.

Full changelog

Files

Device Recovery bootloader Firmware
ZT703S OpenOSC-Recovery-ZT703S.ZTK OpenOSC-Firmware-ZT703S.OSC
ZT706S OpenOSC-Recovery-ZT706S.ZTK OpenOSC-Firmware-ZT706S.OSC

Download the files for your model.

Updating

  • From 0.3.0 or newer — Hold F2 at startup to open Recovery. Copy the matching .OSC into /openosc/UPDATE/. Select UPDATE FIRMWARE, choose the image, select CONFIRM, and press MENU. When the update completes, select START APPLICATION.
  • Both Recovery packages match 0.21.4 byte for byte. If Recovery is already installed, update the firmware alone.
  • From 0.2.0 or older, or for a new installation — Follow the installation instructions.

Known issues

  • CH1 interleaving is experimental on the ZT703S. Extra noise can still be visible when both ADCs are active at 280 MS/s. On-device offset and gain/timing calibration is available.
  • Dense two-channel captures with waveform density enabled can reduce the display frame rate.
  • In single-slot configurations, an armed capture can reuse the previous record's memory. In WAIT, panning is unavailable until a new record is captured. Completed records can be panned in HOLD.
  • Settings are saved at an orderly power-off, restart or Recovery entry. An unexpected power loss can discard settings changes made since the last save.
  • USB data requires the 200/100 performance profile. USB hubs can cause disconnects.

SHA-256

7bc3b5663bfc7af2a759c72d879bd8b66971fd72541fb543f153955372842e8f  OpenOSC-Firmware-ZT703S.OSC
89e0aa023f9061b4da1e6c0d5a2fe9fc0ce236c2ce504af3ac1deac1a0d0ed4b  OpenOSC-Firmware-ZT706S.OSC
a389038d3aeec07168e0cc07d5f311815ca1054bf2f7c45a2c7b168680862d81  OpenOSC-Recovery-ZT703S.ZTK
6f1db91d198f5c406676c9e5751b1e73a69557c21dd63721c6a79180eb876fd2  OpenOSC-Recovery-ZT706S.ZTK