OpenOSC 0.21.6

Choose a tag to compare

@bve bve released this 23 Nov 16:58
· 1 commit to main since this release

Warning

Work in progress, for enthusiasts only. ZT703S is the main development and hardware test device.
Hardware coverage for the ZT706S is limited.
Install at your own risk. Return to the stock firmware.

What's new

Multimeter diode-test scale (#116)

The analog dial and digital bar now use a 0–3.3 V scale on both models. The scale appears as soon as diode mode reports a reading or OL. The needle and bar track the measured voltage.

Thanks to @LeonIndman for the report.

Boot credits

Added RAPO to the donors list. Thank you for supporting the project.

Testing

Built the firmware and Recovery for both models. Native DMM model tests verified the diode scale, normal readings and overload on both models. The ZT703S screen rendering, update validation and stack checks passed. This release was verified without flashing a physical device.

Full changelog

Files

Device Recovery bootloader Firmware
ZT703S OpenOSC-Recovery-ZT703S.ZTK OpenOSC-Firmware-ZT703S.OSC
ZT706S OpenOSC-Recovery-ZT706S.ZTK OpenOSC-Firmware-ZT706S.OSC

Download the files for your model.

Updating

  • From 0.3.0 or newer — Hold F2 at startup to open Recovery. Copy the matching .OSC into /openosc/UPDATE/. Select UPDATE FIRMWARE, choose the image, select CONFIRM, and press MENU. When the update completes, select START APPLICATION.
  • Both Recovery packages match 0.21.5 byte for byte. If Recovery is already installed, update the firmware alone.
  • From 0.2.0 or older, or for a new installation — Follow the installation instructions.

Known issues

  • CH1 interleaving is experimental on the ZT703S. Extra noise can still be visible when both ADCs are active at 280 MS/s. On-device offset and gain/timing calibration is available.
  • Dense two-channel captures with waveform density enabled can reduce the display frame rate.
  • In single-slot configurations, an armed capture can reuse the previous record's memory. In WAIT, panning is unavailable until a new record is captured. Completed records can be panned in HOLD.
  • Settings are saved at an orderly power-off, restart or Recovery entry. An unexpected power loss can discard settings changes made since the last save.
  • USB data requires the 200/100 performance profile. USB hubs can cause disconnects.

SHA-256

effc9421a2568c573938bdd10bfb8aa8481858c84ccc7a9f472f96a75d5359ee  OpenOSC-Firmware-ZT703S.OSC
2b6759fcf6bb2e11e054fcd631e01b6358e02eaf35b0972cfbc025444980effc  OpenOSC-Firmware-ZT706S.OSC
a389038d3aeec07168e0cc07d5f311815ca1054bf2f7c45a2c7b168680862d81  OpenOSC-Recovery-ZT703S.ZTK
6f1db91d198f5c406676c9e5751b1e73a69557c21dd63721c6a79180eb876fd2  OpenOSC-Recovery-ZT706S.ZTK