Skip to content

OpenOSC 0.21.8

Choose a tag to compare

@bve bve released this 24 Nov 12:52
· 32 commits to main since this release

Warning

OpenOSC is in active development. ZT703S is the main development and hardware test device.
Hardware coverage for the ZT706S is limited.
Install at your own risk. Return to the stock firmware.

What's new

Scope panning and trigger level

  • F2 panning returns to the original trigger position after either end of the record. The PAN indicator clears at the original position, even when successive captures differ slightly in length. #127 · #129
  • The trigger readout and Level editor show 0 V at the nearest comparator code after a vertical position change. An explicitly selected zero stays aligned as position or V/div changes. SCPI reports the calibrated comparator threshold. #127

Deep memory

Analyze → Acquisition → Deep memory now offers two ways to use memory above the default 64K budget. Longer record is the default and gives you more waveform to pan through at the selected time/div. Higher sample rate uses the extra memory to increase the sample rate up to the hardware limit. Longer records take more time to capture. #130

ZT703S DAC stability

On the 200 MHz performance profile, the DAC now updates at up to 6.25 MS/s to give DMA transfers more time during simultaneous scope capture and measurements. The maximum for Sine and Positive sine on this profile is 223.2 kHz. Frequency quantization can be larger near that limit. For example, a 200 kHz request produces about 198.4 kHz. DAC limits on the 280 and 360 MHz profiles and the PWM square frequency range are unchanged. #128

Auto Setup

When a short trial follows a faster trigger signal, Auto Setup keeps the vertical range already found for a slower signal on the other channel. This prevents repeated range adjustments that can stall Auto Setup on mixed-frequency inputs.

Thanks to @Nikonist, @TzOk83 and @Fyfar for the reports and reproduction details.

Testing

On a ZT703S, the 10 kHz Low sine ran for over two minutes at each of 25 and 50 MS/s with screen measurements enabled. The DAC square was checked with both channels on. An induced DMA fault still stopped the output and returned -240. Bench checks also covered panning from both ends, returning to the trigger, the 0 V readout, and longer captures with the same time/div and sample rate. At 2 ms/div with one channel, switching from 64K to 512K extended the measured record from about 20 ms to 161 ms. Auto Setup was checked with mixed-frequency inputs.

Firmware and Recovery builds passed for both models, along with host regressions for acquisition, controls, DAC behavior, and firmware update validation. Hardware testing used the ZT703S.

Full changelog

Files

Device Recovery bootloader Firmware
ZT703S OpenOSC-Recovery-ZT703S.ZTK OpenOSC-Firmware-ZT703S.OSC
ZT706S OpenOSC-Recovery-ZT706S.ZTK OpenOSC-Firmware-ZT706S.OSC

Download the files for your model.

Updating

  • From 0.3.0 or newer — Hold F2 at startup to open Recovery. Copy the matching .OSC file into /openosc/UPDATE/. Select UPDATE FIRMWARE, choose the image, select CONFIRM, and press MENU. When the update completes, select START APPLICATION.
  • The Recovery packages match 0.21.7 byte for byte. If Recovery is already installed, update the firmware alone.
  • From 0.2.0 or older, or for a new installation — Follow the installation instructions.

Known issues

  • CH1 interleaving is experimental on the ZT703S. Extra noise can still be visible when both ADCs are active at 280 MS/s. On-device offset and gain/timing calibration is available.
  • Dense two-channel captures with waveform density enabled can reduce the display frame rate.
  • A single-slot capture can reuse the previous record's memory. If this happens while waiting for a trigger, panning becomes available once a new record is captured.
  • Settings are saved at an orderly power-off, restart or Recovery entry. Unexpected power loss can discard changes made since the last save.
  • USB data requires the 200/100 performance profile. USB hubs can cause disconnects.

SHA-256

c1e408f596344705ed5f15d7fe3006083b85a3e0f6f9f662ffe70d6081684c76  OpenOSC-Firmware-ZT703S.OSC
2a9f2fb7781052b808128871d2c689e596d8e785eb5b4d3512d3a36d307a9426  OpenOSC-Firmware-ZT706S.OSC
a389038d3aeec07168e0cc07d5f311815ca1054bf2f7c45a2c7b168680862d81  OpenOSC-Recovery-ZT703S.ZTK
6f1db91d198f5c406676c9e5751b1e73a69557c21dd63721c6a79180eb876fd2  OpenOSC-Recovery-ZT706S.ZTK