OpenOSC 0.22.0
Pre-releaseWarning
OpenOSC is in active development. This pre-release makes substantial changes to the oscilloscope, measurement display, USB controller and idle power management. Please test it on your own setup before relying on it and report regressions with your model and settings. Hardware testing for this build used the ZT703S. The ZT706S was checked through builds and host tests. Return to the stock firmware.
What's new
Compressed waveform exports
Analyze now offers Save waveform CSV.gz and Save waveform RAW.gz. Both stream gzip files to /openosc/CAPTURES using a shared 16 KiB workspace. Decompress a file on a computer to recover the original export bytes. If the disk fills during a compressed save, OpenOSC discards the temporary file and reports the space error. Issue #166 · SCPI file download
Scope trace labels and frequency measurements
- The Math trace now shows MATH and its channel operation. The label moves below the counter when the overlay is visible, making the extra trace easier to identify with two inputs enabled. Issue #167
- Analyze → Measurements → Short window f/T can use a recent, settled hardware count when a short measurement window has no usable frequency of its own. SCPI
WAV:MEAS?reportsCNTorCNT?for a value from the counter. Issue #171 - Analyze → Display → Frequency counter → No edges counted → Trace f (~) can show an approximate waveform frequency in the overlay and FC readout when the comparator counts zero edges. SCPI counter queries still report the hardware count. Issue #172
- In the same section, Timed edges improves the resolution of slow signals on the physical CH1 input. CH2 keeps its gate counter. The new frequency options are opt-in. Issue #173
Multimeter units and measurement wake
Resistance and capacitance use familiar prefixes that follow the measurement resolution in the reading, REL and statistics. Examples include 670.0 kΩ, 90 pF and 22000 µF. Issue #168
Settings → Device → Power → Wake on measurement offers Off, DMM, Scope and Both. A qualifying DMM reading change or a genuine scope trigger restores the backlight and restarts the idle power-off timer. Issue #169
Automatic USB controller
Settings → Files & USB → USB controller → Auto probes for a host and turns the controller off when no computer is detected. With the USB power signal still high, it retries periodically. Enabled is the default. Issue #170
Testing
Firmware and Recovery packages were built and checked for both models. Host tests covered gzip round trips at maximum depth, full-volume cleanup, settings persistence and the hardware counter state machine. ZT703S testing covered compressed exports and a full-volume refusal, the Math label, DMM display and CSV output, Scope backlight wake, USB host detection and retries, and all three frequency options. The 32,768-point CSV went from 1.29 MB to 187 kB and the 258,048-point RAW went from 516 kB to 110 kB. The reported ZT706S AC-coupling sequence still needs hardware confirmation. Full changelog
Files
| Device | Recovery bootloader | Firmware |
|---|---|---|
| ZT703S | OpenOSC-Recovery-ZT703S.ZTK |
OpenOSC-Firmware-ZT703S.OSC |
| ZT706S | OpenOSC-Recovery-ZT706S.ZTK |
OpenOSC-Firmware-ZT706S.OSC |
Download the files for your model.
Updating
- From 0.3.0 or newer — Hold F2 at startup to enter Recovery. Copy the matching
.OSCto/openosc/UPDATE/. Select UPDATE FIRMWARE, choose the file and confirm. After the update, select START APPLICATION. - For a new installation or from 0.2.0 or older — Follow the installation instructions.
- Before downgrading from USB Auto — Select Enabled or Disabled and save with an orderly restart. Older firmware may reject a settings file containing Auto.
Known issues
- USB data requires the 200/100 performance profile. USB hubs can cause disconnects.
- PNG screenshots and compressed waveform exports share temporary acquisition memory. A save can be temporarily unavailable while another task uses that memory.
- Compression ratio depends on the data. A deep waveform can still exceed the available space.
- Continuous envelope transfers and dense two-channel waveform density can reduce the display update rate. At slow timebases, allow more than the record duration plus one second for a fresh-frame SCPI query.
- CH1 interleaving remains experimental on the ZT703S. Offset, gain and timing calibration are available on the instrument.
- Settings are saved on orderly power-off, restart or entry into Recovery. Unexpected power loss can discard changes made since the last save.
SHA-256
e6fccc44044b194466a6e039b1dacf13ff0bd2dc76b28b074f6890412320d3b6 OpenOSC-Firmware-ZT703S.OSC
a774d5ace42c4919720e66e1e3476b6a1e1512917074ff69c11ac05a492fd484 OpenOSC-Firmware-ZT706S.OSC
a9b9a8c120b0a045fc14ed698d75bc3b826242c7a816dfe6112c5606d8ff5113 OpenOSC-Recovery-ZT703S.ZTK
51ed387d80c88c9d36bf34b495b3cef10c90b1cea03c7eb55bed55dc3aa151cc OpenOSC-Recovery-ZT706S.ZTK