Skip to content

OpenOSC 0.22.2

Pre-release
Pre-release

Choose a tag to compare

@bve bve released this 26 Nov 23:49
· 2 commits to main since this release

Warning

OpenOSC is in active development. This pre-release changes the multimeter display and adds waveform file management. Hardware testing used the ZT703S. The ZT706S was checked through builds and host tests. Please report regressions with your model, settings and steps to reproduce. Return to the stock firmware.

What's new

Multimeter reading layout

The unit sits to the right of the main digits on the same baseline. The digital reading keeps a fixed right edge across ranges, and both displays center OL. MIN, MAX and AVG share the main unit when all three use it. A capacitance reading can display 90 pF on the 9.999 nF range while keeping its full resolution. Issue #185 · Multimeter guide

Manage saved waveforms

Open Settings → Files & USB → Waveforms to browse recordings by filename and size. Press F3 Delete and confirm to remove one file. Clear waveforms asks before clearing waveform files in /openosc/CAPTURES, including earlier plain CSV and RAW exports. During the operation, F4 Stop ends the cleanup. Issue #186 · Managing waveforms

Waveform export menu

Analyze offers Save waveform CSV.gz and Save waveform RAW.gz. Both stream compressed data and check available space as they write. Decompress the archive to recover the CSV or RAW data. Issue #187 · SCPI file download

Testing

Both firmware and Recovery packages were built and verified. On the ZT703S, hardware tests covered saving and reading both gzip formats, deleting individual test captures and canceling a bulk clear. The bulk deletion and low-space cases were tested on FAT images. Host tests covered the multimeter layout, waveform browser and menu, and SCPI file ownership. Full changelog

Files

Device Recovery bootloader Firmware
ZT703S OpenOSC-Recovery-ZT703S.ZTK OpenOSC-Firmware-ZT703S.OSC
ZT706S OpenOSC-Recovery-ZT706S.ZTK OpenOSC-Firmware-ZT706S.OSC

Download the files for your model. Both Recovery packages have the same SHA-256 checksums as in 0.22.1.

Updating

  • From 0.3.0 or newer — Hold F2 at startup to enter Recovery, copy the matching .OSC file to /openosc/UPDATE/, select UPDATE FIRMWARE and confirm the file. Select START APPLICATION after the update finishes.
  • For a new installation or from 0.2.0 or older — Follow the installation instructions.

Known issues

  • USB data uses the 200/100 profile. USB hubs can cause disconnects.
  • Keep the CDC port open when switching from Scope to DMM until the instrument changes profiles. Closing it immediately can lose the SYST:MODE DMM command. Issue #177
  • A CDC file listing was interrupted during a CSV.gz save on the ZT703S bench. The saved archive was read and verified after reconnecting.
  • On the test ZT703S, Scope frames stopped updating at 100 µs/div with both AWG and the built-in DAC sine output. Concurrent capture and AWG were verified at 1 ms/div.
  • CH1 interleaving is experimental on the ZT703S.
  • Settings are written on orderly power-off, restart or Recovery entry.

SHA-256

0cdeab5ca76bb2777323fa510d01407edb9fd1fee074b087c5ea8688ba83af03  OpenOSC-Firmware-ZT703S.OSC
3bc8691d72b042f102588db4789e2614f5809942f5f96dc38cc2d1d6c707c518  OpenOSC-Firmware-ZT706S.OSC
a9b9a8c120b0a045fc14ed698d75bc3b826242c7a816dfe6112c5606d8ff5113  OpenOSC-Recovery-ZT703S.ZTK
51ed387d80c88c9d36bf34b495b3cef10c90b1cea03c7eb55bed55dc3aa151cc  OpenOSC-Recovery-ZT706S.ZTK