Membrane Control 0.1.20 recovers some lost manual-valve replies and shows an advancing sample time on the live sensor panel.

Developed for Dr. Onita Basu's lab at Carleton University.

Download MembraneControl-Setup-0.1.20-windows-x64.exe from Assets. Use Flash normal controller from the updated setup for valve receipts.

  • A fresh ESP32 telemetry frame can confirm a valve command using the exact pending request ID and matching requested states. An unconfirmed command still triggers Stop. This recovery does not resend valve commands or change the watchdog.
  • Dashboard → Live sensor check shows the ESP32's sample time so steady readings are distinguishable from a frozen display. Test voltages require no manual configuration on factory firmware.
  • The guides describe how to investigate a complete serial dropout and how to measure two reference points for a calibrated profile.