Skip to content

Releases

Membrane Control 0.1.27 - Bluetooth pairing and flow display

Choose a tag to compare

@regular030 regular030 released this 10 Nov 17:04

Membrane Control 0.1.27 updates Bluetooth pairing and the live flow display.

Developed for Dr. Onita Basu's lab at Carleton University.

  • To pair the controller again, use the new USB Bluetooth pairing shortcut with the updated firmware, then select the outgoing Windows Bluetooth COM port.
  • The Dashboard shows the lab's corrected default flow estimate. It labels values outside the checked range and shows raw millivolts where a usable estimate is unavailable. Existing reference fits, experiment records, and controller limits keep their current calibration.
  • The Windows Setup includes the desktop app, compiled ESP32 firmware, and flashing and pairing shortcuts.

Installation instructions

Membrane Control 0.1.26 - Certificate-based sensor estimates

Choose a tag to compare

@regular030 regular030 released this 09 Nov 09:15

Membrane Control 0.1.26 uses the pressure and flow calibration certificate points for live Dashboard estimates.

Developed for Dr. Onita Basu's lab at Carleton University.

  • With the rig stopped and zero pressure and flow verified, click Zero pressure and flow to capture a baseline for the current connection. The Dashboard shows live estimates wherever the certificate points and ADC input range support them.
  • The zero baseline shows changes in raw voltage. Near-zero pressure can fall below the usable ADC range, and the flow certificate starts at 20 mL/min. The Dashboard labels those ranges.
  • Sensor setup still accepts two independent references when available. The commissioned firmware profile drives run data and pressure protection.

Sensor setup instructions

Membrane Control 0.1.25 - Valve commands and sensor setup

Choose a tag to compare

@regular030 regular030 released this 09 Nov 08:01

Membrane Control 0.1.25 incorporates the tested Bluetooth reconnection fix and improves live sensor estimates and manual valve commands.

Developed for Dr. Onita Basu's lab at Carleton University.

  • Install MembraneControl-Setup-0.1.25-windows-x64.exe. If the board is still on the v0.1.24 Bluetooth image, use Flash normal controller (Bluetooth) over wired USB with the valve and pump loads isolated, then use the existing outgoing Bluetooth port.
  • The Dashboard shows a labeled live estimate for each sensor in range. With two measured points in Sensor setup, it uses your fit within the captured range for that connection. Commissioned readings continue to drive the run and its protection.
  • The serial queue sends the opening Bluetooth heartbeat before a subsequently queued manual valve command. Use the normal Dashboard toggles to hold a valve open and Stop valves to release all valves. A lost controller link still releases them.

Bluetooth controller instructions · Sensor setup instructions

Membrane Control 0.1.24 - Bluetooth pairing recovery

Choose a tag to compare

@regular030 regular030 released this 06 Nov 18:44

Membrane Control 0.1.24 lets the normal ESP32 Bluetooth firmware use the saved bench pairing and adds a shareable connection report.

Developed for Dr. Onita Basu's lab at Carleton University.

  • Install MembraneControl-Setup-0.1.24-windows-x64.exe and run Flash normal controller (Bluetooth) with the valve and pump loads isolated. Flash over USB and try the existing outgoing Bluetooth COM port first.
  • If the controller reports a missing bond on its wired boot log, use Flash ESP32 Bluetooth test to enroll the lab PC and verify that the board saves one bond before reflashing the normal Bluetooth firmware.
  • Under Connections and settings, Copy connection diagnostics collects the selected and detected COM ports, current status and recent ESP32 connection events for troubleshooting.

Bluetooth controller instructions · Sensor setup instructions

Membrane Control 0.1.23 - Bluetooth controller and sensor setup

Choose a tag to compare

@regular030 regular030 released this 06 Nov 18:07

Membrane Control 0.1.23 adds normal ESP32 controller operation over Bluetooth and a Sensor setup window for live inputs and reference checks.

Developed for Dr. Onita Basu's lab at Carleton University.

  • Install MembraneControl-Setup-0.1.23-windows-x64.exe. With the loads isolated, use Flash normal controller (Bluetooth) to upload over the wired USB-to-UART port. Keep the existing pairing from Flash ESP32 Bluetooth test, then connect to its outgoing Bluetooth COM port in the app. The normal build uses real manual valve outputs and the same hardware profile as the wired controller.
  • Open Sensor setup on Dashboard or under Connections and settings to see live pressure, flow and pump feedback pin voltages. Capture a per-sensor zero and two measured reference points, then copy the readings and calculated fit for the rig's profile.
  • Firmware discards the previous control session when Bluetooth disconnects. Wired firmware writes each reply as one complete serial frame. Hardware testing on the lab rig remains pending.

Bluetooth controller instructions · Sensor setup instructions

Membrane Control 0.1.22 - Bluetooth connection test

Choose a tag to compare

@regular030 regular030 released this 06 Nov 05:21

Membrane Control 0.1.22 adds an optional Bluetooth Classic connection test for the ESP32-WROOM relay board.

Developed for Dr. Onita Basu's lab at Carleton University.

Install MembraneControl-Setup-0.1.22-windows-x64.exe, then run Flash ESP32 Bluetooth test with the wired USB-to-UART COM port and the loads isolated. Pair with MembraneControl-Bench in Windows and select its outgoing Bluetooth COM port in the app. The firmware reports simulated data and outputs.

To return to the rig's wired controller, run Flash normal controller with the wired port. See the Bluetooth test instructions.

Membrane Control 0.1.21 - Live sensor zero reference

Choose a tag to compare

@regular030 regular030 released this 05 Nov 23:50

Membrane Control 0.1.21 adds a pressure and flow zero reference to the live sensor display.

Developed for Dr. Onita Basu's lab at Carleton University.

Download MembraneControl-Setup-0.1.21-windows-x64.exe from Assets. If your board already shows live sensor inputs, install the desktop app and keep the firmware you have.

With the pump stopped, no flow, and the reference gauge at zero, click Zero pressure and flow under the Dashboard's live sensor readings. The app shows how many millivolts each reading has changed since that reference. Use Clear zero to remove it. Reconnecting clears the reference.

The factory firmware and calibrated profile continue to handle the actual outputs and pressure limit. See the sensor guide.

Membrane Control 0.1.20 - Valve receipts and sensor sample time

Choose a tag to compare

@regular030 regular030 released this 05 Nov 22:39

Membrane Control 0.1.20 recovers some lost manual-valve replies and shows an advancing sample time on the live sensor panel.

Developed for Dr. Onita Basu's lab at Carleton University.

Download MembraneControl-Setup-0.1.20-windows-x64.exe from Assets. Use Flash normal controller from the updated setup for valve receipts.

  • A fresh ESP32 telemetry frame can confirm a valve command using the exact pending request ID and matching requested states. An unconfirmed command still triggers Stop. This recovery does not resend valve commands or change the watchdog.
  • Dashboard → Live sensor check shows the ESP32's sample time so steady readings are distinguishable from a frozen display. Test voltages require no manual configuration on factory firmware.
  • The guides describe how to investigate a complete serial dropout and how to measure two reference points for a calibrated profile.

See the serial recovery guide and sensor guide.

Membrane Control 0.1.19 - Live sensors and ESP32 connection log

Choose a tag to compare

@regular030 regular030 released this 05 Nov 22:16

Membrane Control 0.1.19 adds live sensor checks and a persistent ESP32 connection log.

Developed for Dr. Onita Basu's lab at Carleton University.

Download MembraneControl-Setup-0.1.19-windows-x64.exe from Assets. The new app records serial events with compatible controller firmware. Use Flash normal controller from the updated setup for live sensor voltages and ESP32 reset reporting.

  • Dashboard → Live sensor check shows GPIO34 pressure, GPIO36 flow, and optional GPIO39 pump feedback in millivolts while idle or using manual valves. Estimates assume the documented divider and sensors. Calibrated readings and experiment protection still require a reviewed profile.
  • Connections and settings → Copy ESP32 log copies recent valve commands, acknowledgements, link errors, timeouts and reset information. The log survives app restarts and omits the routine sample stream.
  • The pump profile template records the lab-confirmed bit 15, LOW for STOP and HIGH for RUN. The factory firmware leaves the pump output off, with manual V3 available as before.

See serial recovery and the sensor test guide.

Membrane Control 0.1.18 - ESP32 recovery and sensor voltage check

Choose a tag to compare

@regular030 regular030 released this 05 Nov 21:46

Membrane Control 0.1.18 improves normal controller recovery after telemetry loss and adds an idle sensor voltage check.

Developed for Dr. Onita Basu's lab at Carleton University.

Install

Download MembraneControl-Setup-0.1.18-windows-x64.exe from Assets. The app-only update improves serial recovery. Use Flash normal controller from the new setup for the sensor check.

Serial recovery

An unanswered handshake and an idle stale connection now release the COM port. The app pauses heartbeats while waiting for Stop and keeps an unconfirmed Stop fault visible. After reconnecting to a faulted ESP32, the operator can clear the fault after checking the rig. Old valve commands are never replayed.

See the serial recovery guide.

Read sensor inputs

Stop the controller, then open Connections and settings → Read sensor inputs. Factory normal firmware samples the protected ADC inputs and reports millivolts for pressure (IO34), flow (SVP / GPIO36), and optional pump feedback (SVN / GPIO39). The check does not populate calibrated PSI or mL/min or enable experiments.