Repository navigation
Membrane Control 0.1.21 - Live sensor zero reference
Pre-release
Pre-release
Membrane Control 0.1.21 adds a pressure and flow zero reference to the live sensor display.
Developed for Dr. Onita Basu's lab at Carleton University.
Download MembraneControl-Setup-0.1.21-windows-x64.exe from Assets. If your board already shows live sensor inputs, install the desktop app and keep the firmware you have.
With the pump stopped, no flow, and the reference gauge at zero, click Zero pressure and flow under the Dashboard's live sensor readings. The app shows how many millivolts each reading has changed since that reference. Use Clear zero to remove it. Reconnecting clears the reference.
The factory firmware and calibrated profile continue to handle the actual outputs and pressure limit. See the sensor guide.