GalaxyBench

barriap edited this page Feb 16, 2023 · 40 revisions

BEFORE STARTING THE MACHINE

  • Check if the compressor is ON (otherwise turn it ON: compressor room next to the workshop)

  • Check if the compressor switch in the electical cabinet is on ON (otherwise turn it ON)

gal1

  • Start supply from the compressor to the machine (black valve on the wall behind the machine)

gal2

MACHINE START

  • Start the machine from the switch on the cabinet on the machine left
    • IMPORTANT: ONLY AFTER STARTING COMPRESSOR SUPPLY!!!

gal3

  • When the light on the controller are stable, press and hold “MACH.START” button (Only lights on: Jog Mode on MACH, SLOW, MOVEMENTS; PROBES ENABLE)

  • If the machine doesn’t start, check the safety buttons are unlocked.

    • There are two of them: one on the cabinet on the left, one on the controller
  • Place the grid and the probe on the right position

  • Mount the arrays on the grid (barcode on the right bottom of the upper surface)

MEASURMENT ROUTINE SELECTION

  1. Start PC-DMIS 2019
  2. In “Seleziona file tastatore” window, leave “ROBERTA” and press ok
  3. Press ok in the window “Avviare macchina se necessario”
  4. Close the “PCDLRN” window
  5. File → Apri → Find the routine (File name: LYSO_array.PRG Directory:C:\Users\Public\Documents\Hexagon\PC-DMIS\2019 R2)
  6. Check both “Apri on-line” (/”Macchina1”) and “Includi CAD”

STARTING MEASUREMENT

  • Select everything (“Seleziona tutto” or Maiusc+F3)
  • Select “No” on the window “OK a contrassegnare gli elem. Dell’allineam. manuale?” if the grid is positioned in the right position.
    • If the grid is in another position, select “Yes” and follow the instructions for the manual movements after the Run
  • Select “Esegui” or Ctrl+Q
  • Using the controller, manually touch the upper point of the calibration sphere and than click DONE
  • The calibration starts and then the probing is automatic
  • Before the actual probing, there will be 12 windows to be filled: one window to be filled with the array code and the other one with its type, this couple of windows for each array

gal4

The order is (as shown in the picture):

  • array1=array in the position (row,column)=(3,2)
  • array2=array in the position (3,1)
  • array3=array in the position (2,2)
  • array4=array in the position (2,1)
  • array5=array in the position (1,2)
  • array6=array in the position (1,1)

gal5

COPY RAW DATA IN WORKSTATION HPZ4

Raw data are saved as "barcode_day-time.TXT" format in the directory C:\Users\Angelo\Desktop\Roberta\MisureDimensionali_LYSO\Array

  • Open Windows command prompt
  • Go to the directory Array
cd C:\Users\Angelo\Desktop\Roberta\MisureDimensionali_LYSO\Array
  • Copy raw data in the workstation hpz4
pscp -P 3333 *.TXT cmsdaq@pccmsdaq01.roma1.infn.it://data/cmsdaq/DimensionBench/Arrays/ArrayData
  • Copy raw data from dir Array to dir Processati
pscp *.TXT C:\Users\Angelo\Desktop\Roberta\MisureDimensionali_LYSO\Processati
  • Delete raw data from dir Array
del *.TXT

AFTER THE MEASUREMENT

  • Bring the probe manually to a safe position
  • Turn the machine OFF
  • Turn the compressor supply OFF from the valve (ONLY AFTER TURNING OFF THE MACHINE)