GalaxyBench

barriap edited this page Jul 5, 2023 · 40 revisions

Table of Contents

BEFORE STARTING THE MACHINE

  • Check if the compressor is ON (otherwise turn it ON: compressor room next to the workshop)

  • Check if the compressor switch in the electical cabinet is on ON (otherwise turn it ON)

gal1

  • Start supply from the compressor to the machine (black valve on the wall behind the machine)

gal2

MACHINE START

  • Start the machine from the switch on the cabinet on the machine left
    • IMPORTANT: ONLY AFTER STARTING COMPRESSOR SUPPLY!!!

gal3

  • When the light on the controller are stable, press and hold “MACH.START” button (Only lights on: Jog Mode on MACH, SLOW, MOVEMENTS; PROBES ENABLE)

  • If the machine doesn’t start, check the safety buttons are unlocked.

    • There are two of them: one on the cabinet on the left, one on the controller
  • Place the grid and the probe on the right position

image

  • Mount the arrays on the grid (barcode on the right bottom of the upper surface)

image

MEASURMENT ROUTINE SELECTION

  1. Start PC-DMIS 2019
  2. In “Seleziona file tastatore” window, leave “ROBERTA” and press ok
  3. Press ok in the window “Avviare macchina se necessario”
  4. Close the “PCDLRN” window
  5. File → Apri → Find the routine (File name: LYSO_array_PREIRR.PRG/LYSO_array_POSTIRR.PRG/LYSO_array_TEST.PRG Directory:C:\Users\Public\Documents\Hexagon\PC-DMIS\2019 R2)
  6. Check both “Apri on-line” (/”Macchina1”) and “Includi CAD”
  7. Check speed on the controller is at maximum

STARTING MEASUREMENT

  • Select everything (“Seleziona tutto” or Maiusc+F3)
  • Select “No” on the window “OK a contrassegnare gli elem. Dell’allineam. manuale?” if the grid is positioned in the right position.
    • If the grid is in another position, select “Yes” and follow the instructions for the manual movements after the Run
  • Select “Esegui” or Ctrl+Q
  • Select "Ok" in the window "Avviare movimento del polso"
  • The calibration starts and then the probing is automatic
  • Before the actual probing, there will be 12 windows to be filled: one window to be filled with the array code (LAST 6 DIGITS) and the other one with its type, this couple of windows for each array

gal4

The order is (as shown in the picture):

  • array1=array in the position (row,column)=(3,2)
  • array2=array in the position (3,1)
  • array3=array in the position (2,2)
  • array4=array in the position (2,1)
  • array5=array in the position (1,2)
  • array6=array in the position (1,1)

gal5

COPY RAW DATA IN WORKSTATION HPZ4

Raw data are saved as "barcode_day-time.TXT" format in the directory C:\Users\Angelo\Desktop\Roberta\MisureDimensionali_LYSO\Array

  • Open Windows command prompt
  • Go to the directory Array
cd C:\Users\Angelo\Desktop\Roberta\MisureDimensionali_LYSO\Array
  • Copy raw data in the workstation hpz4: directory ArrayData_tmp
pscp -P 3333 *.TXT cmsdaq@pccmsdaq01.roma1.infn.it://data/cmsdaq/DimensionBench/Arrays/ArrayData_tmp
  • Copy raw data from dir Array to dir Processati
copy *.TXT C:\Users\Angelo\Desktop\Roberta\MisureDimensionali_LYSO\Processati
  • Delete raw data from dir Array
del *.TXT

AFTER THE MEASUREMENT

  • Bring the probe manually to a safe position
  • Turn the machine OFF
  • Turn the compressor supply OFF from the valve (ONLY AFTER TURNING OFF THE MACHINE)