- BEFORE opening the box, check if the HV is OFF, otherwise turn it OFF
Open firefox
Go to page localhost:8888
IMPORTANT DO THIS ONLY IF THERE IS NO OPEN FIREFOX WITH A GUI ALREADY RUNNING. 2 GUIS CONFLICT!!
DO NOT UNPLUG/PLUG ARDUINOS OR OTHER USB CABLES UNLESS YOU KNOW WHAT YOU'RE DOING, OTHERWISE PORT NAMES WILL BE
CHANGED!
Currently:
/dev/sensor_1 External temperature+humidity sensors
/dev/sensor_0 Internal temperature+humidity sensors
/dev/ttyUSB0 Lauda Chiller
/dev/usbtmc0 LED Pulser
To check temperature graph:
https://thingspeak.com/channels/781520
Remember: Temperature should be ~22.3 degree when external temperature is 26 (and chiller is at 18.5).
Chiller lies below the box (on the right side from the door).
The chiller temperature is set at 18.5 degrees. Pump flow config is set at 3
The status of the chiller is reported in the GUI. Normally the cooling should be COOLING OK and green
Every 2 weeks please check the level of the water (please use only demineralised water to refill, which should be close to the chiller)
The chiller should be kept ON unless there will be no measurements for more than 2 weeks. To turn ON/OFF simply press the COOLING button, which should become yellow (COOLING OFF). To turn back ON simply press the button again, which after few seconds should become green (COOLING OK)
IMPORTANT DO NOT CHANGE PARAMS UNLESS YOU KNOW WHAT YOU'RE DOING