Dear Protocentral,
I bought the spo2 shield recently and followed the instructions given in this git. Processing graph output values always shows values in range of 60 to 70. A normal healthy person like I should get values in 94+ range. I came across this section after a brief look at processing code:
` AvgYdata[arrayIndex] = (float)receivedVoltage_RED;
AvgZdata[arrayIndex] = (float)receivedVoltage_IR;
value1 = (float)( AvgYdata[arrayIndex] - averageValue(AvgYdata));
value2 = (float)( AvgZdata[arrayIndex] - averageValue(AvgZdata));
ydata[arrayIndex] = value1;
zdata[arrayIndex] = value2;
float RedDC = (float) averageValue(AvgYdata);
float IrDC = (float) averageValue(AvgZdata);
arrayIndex++;
if (arrayIndex == pSize)
{
arrayIndex = 0;
time = 0;
RedAC = s.SPO2_Value(ydata);
IrAC = s.SPO2_Value(zdata);
float value = (RedAC/abs(RedDC))/(IrAC/abs(IrDC));
/******** Emprical Formalae *********/
//float SpO2 = 10.0002*(value)-52.887*(value) + 26.817*(value) + 98.293;
// float SpO2 =((0.81-0.18*(value))/(0.73+0.11*(value)));
float SpO2=110-25*(value);
SpO2 = (int)(SpO2 * 100);
SpO2 = SpO2/100;
oxygenSaturation.setText(SpO2+"");`
- Is there any issue in the code or the hardware? Please rectify this issue.
- Further, the values are not consistent. Sometimes there is big difference between two consecutivve values. Is there any special care needed for finger position.