@xbtu2 xbtu2 commented Nov 25, 2025

ATC is supposed to have 3 temperature zones instead of 1. Not specifying all 3 temperatures causes early termination of protocols

@rickwierenga rickwierenga merged commit 86bbc46 into PyLabRobot:main Nov 25, 2025
9 checks passed
@xbtu2 xbtu2 deleted the fix-atc-num-zones branch November 25, 2025 05:05
@BioCam
Copy link
Collaborator

BioCam commented Nov 25, 2025

@xbtu2, How often are people using the 3 temperature zones separately compared to using them all together?

Do you think it would be worth autofilling the temperature zones if only 1 is given?

e.g. if one uses a 96-wellplate and only declares a single temperature at each step (which seems to me the "status quo"), then it will fill the remaining 2 zones with the same temperature.

I think this would facilitate machine agnosticity for thermocycling protocols.

@xbtu2
Copy link
Contributor Author

xbtu2 commented Nov 25, 2025

@BioCam you are probably right. the 3 temperature zones are not even shown in the vendor demo software. i think i will add a param for multi-zone for run_protocol in the backend and just give a warning if temps are autofilled