Manual testing

If you have a UNI-T UTS analyzer, the manual testing guide walks you through checking every tool group, once over raw SCPI and once over VXI-11. It lists the prerequisites, a safe setup (no real address or serial number in anything you share), a step-by-step checklist with the expected result and what to record for each step, how to put the analyzer back as you found it, a reporting template (model, firmware, results per group) and a short troubleshooting table.

Things to know before you start:

  • The first ordinary SCPI query puts the analyzer in remote mode, which locks its touch screen and buttons until you press the physical Esc key on it. On firmware V1.04.0040, :SYSTem:LOCK 0, VXI-11 device_local and :KEY:ESC do not return it to local mode.
  • Never send :KEY: commands or queries. Which other headers trigger remote mode is unknown.
  • SRQ is not available on this firmware; the tools wait for sweeps with *OPC?.
  • A test in the repository fails when a tool has no checklist step, so the guide always covers the full tool list.

See also Troubleshooting and Instrument Behavior, and report results as an issue in the repository.