Connection

connect

Connect to a UNI-T UTS analyzer over auto (the default: VXI-11 when the analyzer offers it, else raw SCPI/TCP), VXI-11, or raw SCPI/TCP. Verifies the link with *IDN? and returns the transport in use and the identification string:

Connected to 192.168.1.100
Transport: VXI-11 (device inst0), selected by auto
Device: UNI-T,UTS3036B,<serial>,<firmware>
Parameter Type Required Default Description
name string (1-32 letters, digits, _, -) No the default instrument, or the only one configured Which instrument to connect. A new name is registered by this call (a first connect that fails leaves it unregistered). The host and port default to the target declared for that name in UTS_INSTRUMENTS; UTS_IP/UTS_PORT belong to default only. Connecting a name again replaces only that instrument's link. See list_instruments
host string No UTS_IP env var (default) or the UTS_INSTRUMENTS target of name Hostname or IP address (letters, digits, ., -, _, : only), or auto to find the analyzer with discover_instruments first: it connects if exactly one UNI-T analyzer answers on the local subnet(s), and otherwise returns an error that lists what was found (or says none was) and asks for an explicit host. The result says the host was auto-selected and shows the model and firmware (quoted) so you can check it; discovery is unauthenticated
port integer 1-65535 No UTS_PORT env var, then 5025 TCP port of the raw transport. VXI-11 finds its own port through the portmapper (TCP port 111), so an explicit port together with transport: "vxi11" is an error; a port (or UTS_PORT) without any transport choice selects raw
transport raw, vxi11 or auto No UTS_TRANSPORT env var, then auto (raw when only a port is given) raw: raw SCPI over TCP. vxi11: VXI-11 core channel; fails if it cannot be opened. auto: VXI-11 if the portmapper on port 111 registers the VXI-11 core program and the link opens, otherwise raw (the reason for the fallback is logged to stderr and shown in the Transport: line)

The VXI-11 device name comes from UTS_VXI11_DEVICE (default inst0; 1 to 64 characters from letters, digits and . _ , -). An invalid transport, UTS_TRANSPORT or UTS_VXI11_DEVICE is an error that names the offending value or variable and sends nothing. Error and timeout messages name the transport when it is not raw, for example Query timeout after 5000ms for command: :FOO? (over VXI-11). Over VXI-11 a blocking *OPC? read is used to wait for a sweep (see single_sweep), a query timeout is reported about 0.5 s after the requested one, and the link stays up. Large replies are not truncated over VXI-11 (see scpi_query and screenshot).

UTS_IP=auto works the same way (also for the startup auto-connect, which logs the auto-selected host). An explicit host parameter always wins over UTS_IP=auto and never triggers discovery. With several instruments, discovery leaves out analyzers that are already connected under another name (they are not probed again or counted), and UTS_IP=auto never picks a host that UTS_INSTRUMENTS names; if the only analyzer found is already connected the error says so.

If UTS_IP is set, the server also connects automatically at startup, using UTS_TRANSPORT and UTS_VXI11_DEVICE; a failed auto-connect is logged to stderr and does not stop the server.

Front-panel lock. The connect and disconnect results say that the analyzer's touch screen and buttons are locked while it is in remote mode and that the user returns it to local mode with the front-panel Esc key. On a UTS3036B V1.04.0040 over raw TCP, connecting alone, *IDN? and :SYSTem:LOCK? did not lock the panel, while :KEY:LOCK?, :KEY:LED? and :SENSe:FREQuency:CENTer? did (VXI-11 was not separately isolated). See verified-hardware.md "Front-panel lock".

discover_instruments

Find UNI-T / VXI-11 analyzers on the local network without knowing their IP address. Read-only: it only identifies, it never changes instrument state. The server broadcasts an ONC RPC portmapper request (UDP port 111, GETPORT for the VXI-11 core program 395183, TCP) on each local IPv4 interface (the directed broadcast address of each subnet), collects the answers for a short window, then asks each responder *IDN? over VXI-11 (falling back to raw SCPI on TCP 5025) and checks whether port 5025 accepts a connection. The analyzer registers VXI-11 with its portmapper, which is why it answers. Works on Windows, macOS and Linux; no dependencies.

Parameter Type Required Default Description
timeout_ms integer 100-10000 No 1500 How long to wait for replies

Returns JSON: instruments (each with host, vxi11, vxi11Port, raw, analyzer, idn with manufacturer, model and firmware, and a note when it could not be identified), notes (sockets that could not be opened, skipped hosts) and a hint. The serial number is deliberately not returned (use identify after connecting). analyzer is true for a UNI-T maker or a UTS model. Finding nothing is not an error.

Scope and trust. Discovery never scans: it sends one request per local interface to that subnet's broadcast address (never a range of hosts or ports) and talks only to hosts that answered. It does not cross routers: an analyzer on another subnet is not found by broadcast (give its address to connect). The portmapper protocol is unauthenticated, so replies are accepted only from ordinary host addresses inside the subnet of the interface the request left from (not loopback, multicast, broadcast, network or the computer's own address) and only when they name an unprivileged port (1024-65535); the identification text is accepted only if it consists of plain identifier characters and is shown quoted. Anything on your LAN can still pretend to be an analyzer, so check the model and address that discovery or host: "auto" reports. The host this server is already connected to is listed but not probed again. Optional mDNS/DNS-SD (_lxi._tcp) is not implemented (the analyzer was not verified to advertise it). The tool does not use the shared instrument connection, so it neither waits for nor blocks other tools.

disconnect

Disconnect from the analyzer. Safe to call when not connected. Optional name: which instrument to disconnect (default: the default one); the others stay connected and the name stays registered, so connect with that name reconnects it. An unknown name is an error listing the configured ones. The result reminds the user that the analyzer is not known to return to local mode on its own: if its front panel is still locked, press Esc on the analyzer.

list_instruments

List the instruments this server knows by name: the default one (UTS_IP), those from UTS_INSTRUMENTS and any added by connect with a name. Read-only, reads the registry only (it never waits for a running sweep), and shows no serial number. No parameters.

Returns JSON: instruments (each with name, connected, isDefault, and when connected connection (host and port) and transport; configured shows the host[:port] declared for it) and a hint. isDefault marks the instrument a call without a name uses: default if it exists, else the only one. With several instruments and none called default, name the one you mean (name for connect, disconnect, identify, and the optional instrument argument on every other tool; see Selecting an instrument). UTS_INSTRUMENTS is name=host[:port],name2=... or JSON ({"sa1":"192.168.1.100:5026"}); an invalid value is logged at startup and ignored.

identify

Query *IDN? (manufacturer, model, serial number, firmware version). Read-only. Optional name: which instrument to query (default: the default one). The output ends with a Transport: line naming the transport in use (raw SCPI over TCP or VXI-11). An empty reply or one without four comma-separated fields is an error that quotes the reply.

Raw SCPI

Escape hatches for commands without a dedicated tool. Syntax is in programming-guide.md.

scpi_query

Send one SCPI query (must end in ?) and return the reply. Read-only: anything that is not a single query is refused (use scpi_command). Unsupported queries get no reply from the instrument and time out (over raw the instrument then stalls for about 2 s; over VXI-11 nothing is lost and the -113 is in the error queue at once). Over VXI-11 the screen image works too: :DISPlay:DATA? with reply: "block" returns the whole 4 MB bitmap (the first 64 KiB are shown, the rest marked truncated) and :DISPlay:ZBmp? the full 200 KB stream, while over raw both stop at 144,800 bytes and time out.

Parameter Type Required Default Description
command string Yes e.g. *IDN?, :SWEep:POINts?
timeout_ms integer 1-120000 No 5000 (text), 30000 (block) Reply timeout
reply text | block No text block reads an IEEE 488.2 # binary block and returns it as base64 (first 64 KiB shown; larger replies are marked truncated)

Only one query per call: on the UTS3036B a ;-joined line of queries is answered with a single reply, which would be misread as the answer to the next query.

scpi_command

Send a SCPI command with no reply expected. Queries are refused (use scpi_query). Like every setter, it clears the error queue (*CLS), sends the command synced with *OPC? and then checks :SYSTem:ERRor?; an error reported by the instrument (for example -131 Invalid suffix) is returned as an error result naming the command.

Parameter Type Required Default Description
command string Yes e.g. :FREQuency:CENTer 1 GHz
confirm boolean No false Required to send a command on the safety denylist: *RST, *SAV, :SYSTem:DEFault, :SYSTem:LOCK, :KEY:*:LOCK, :SYSTem:COMMunicate:LAN:* setters, :SYSTem:CONFigure, :MMEMory:STORe. Only set with the user's approval

Commands containing CR/LF, control characters or non-ASCII text, unterminated quotes, or malformed # blocks are refused.

Frequency

Units are Hz. Command syntax: programming-guide.md section 2. Query forms and Hz/GHz/bare-number setters were verified on the UTS3036B.

get_frequency

Read-only. Returns JSON with center_hz, span_hz, start_hz, stop_hz, center_step_hz, center_step_auto and offset_hz. No parameters.

set_frequency

Frequency parameters are coupled, so choose one group per call; the settings are re-read afterwards and anything the instrument changed is listed in warnings (including center_step_auto and, with zero_span, a span that is not 0).

Parameter Type Description
center_hz number > 0 Center frequency (may be combined with span_hz; span is set first)
span_hz number > 0 Span (use zero_span for zero span)
start_hz number >= 0 Start frequency (alone, or with stop_hz)
stop_hz number > 0 Stop frequency; with start_hz it must be greater than start_hz. The edge that will not cross the other is moved first
full_span boolean Full span; cannot be combined with center/span/start/stop
zero_span boolean Zero span; cannot be combined with center/span/start/stop
center_step_hz number > 0 Center step (turns auto step off)
center_step_auto boolean Automatic center step (not with center_step_hz)
offset_hz number Frequency offset

center/span and start/stop cannot be mixed in one call. Before applying, the instrument's error queue is cleared with *CLS; afterwards it is checked and any error is returned as an error result (settings may be partially applied).

Amplitude

Command syntax: programming-guide.md section 2 (amplitude). Verified on the UTS3036B: see verified-hardware.md.

get_amplitude

Read-only. Returns JSON with reference_level (in the current power_unit), power_unit (dBm, dBmV, dBuV, V, W), scale_per_div_db, scale_type (logarithmic or linear), reference_level_offset_db, attenuation_db, attenuation_auto, preamp and input_impedance_ohm. No parameters. The input impedance is read with the :SENSe:CORRection:IMPedance:INPut:MAGNitude? alias because :INPut:IMPedance? gets no reply.

set_amplitude

All parameters are optional, but at least one is required. The settings are re-read afterwards; values the instrument clamped are listed in warnings, and so are a power_unit, scale_type, preamp or attenuation_auto it did not take.

Parameter Type Description
power_unit dBm | dBmV | dBuV | V | W Power unit (applied first)
reference_level number Reference level in the unit in effect after this call. In dBm the instrument accepts -100 to 30 and clamps the rest; the maximum also depends on attenuation and preamp
scale_per_div_db number 0.1-20 Scale per division
scale_type logarithmic | linear Y scale type
reference_level_offset_db number Reference level offset in dB (clamped to +-327.6); it is included in the reference_level readback
attenuation_db integer 0-51 Manual attenuation; turns automatic attenuation off
attenuation_auto boolean Automatic attenuation (not with attenuation_db). The automatic value follows the reference level and may lag a moment
preamp boolean Preamplifier
input_impedance_ohm 50 | 75 Input impedance (set through the :SENSe:CORRection alias; the :INPut:IMPedance setter is not accepted)

Commands are sent in dependency order (unit, preamp, attenuation, impedance, scale, offset, then reference level) because the reference level is limited by the attenuation and preamp. Numbers are sent without unit suffixes. Before applying, the error queue is cleared with *CLS; afterwards it is checked and any error is returned as an error result (settings may be partially applied).

Bandwidth, sweep and detector

Command syntax: programming-guide.md section 2. Verified on the UTS3036B: see verified-hardware.md. All setters clamp silently, so every setter re-reads the state and lists adjusted values in warnings (including requested auto flags, enums and trigger source).

get_bandwidth / set_bandwidth

get_bandwidth (read-only) returns rbw_hz, rbw_auto, vbw_hz, vbw_auto, vbw_rbw_ratio and filter_shape (gaussian or flattop).

set_bandwidth parameters (all optional, at least one): rbw_hz, rbw_auto, vbw_hz, vbw_auto, vbw_rbw_ratio, filter_shape. A manual value turns the matching auto flag off, so rbw_hz with rbw_auto=true (and likewise for VBW) is rejected. RBW/VBW snap to 1-3-10 steps and clamp to 1 Hz..3 MHz (UTS3036B); the VBW follows RBW times the ratio while VBW is auto. Order sent: shape, RBW, ratio, VBW.

get_sweep / set_sweep

get_sweep (read-only) returns sweep_time_s, sweep_time_auto, sweep_time_rule (normal/accuracy), sweep_type (sweep/fft), sweep_type_auto, sweep_points, trigger_source (immediate/video) and trigger_video_level_dbm.

set_sweep parameters (all optional, at least one): sweep_points (integer 11-40001; this is the trace length), sweep_time_s (1 ms-4000 s, turns auto off), sweep_time_auto, sweep_time_rule, sweep_type, sweep_type_auto, trigger_source, trigger_video_level_dbm (-170..30). External trigger is rejected by the instrument and not offered. Order sent: points, type, rule, auto flags, trigger level then source (setting the level selects the video trigger), and the manual sweep time last. Changing the span or RBW re-enables or recomputes automatic sweep time, so re-read after changing them.

get_detector / set_detector

Uses :SENSe:DETector:TRACe<n> (the plain :DETector[:FUNCtion] is for measurements). get_detector (read-only) takes an optional trace (1-6) and returns {traces: [{trace, detector, auto}]}. set_detector takes trace plus detector (sample, positive, negative, normal, average; turns auto off for that trace) or auto.

Acquisition

Command syntax: programming-guide.md sections 2 and 9. Verified on the UTS3036B: see verified-hardware.md. Important: while a single sweep is running the instrument ignores every command (and *OPC? does not block), so these tools stay silent for the sweep time and then poll *OPC? (src/acquisition.ts, waitForSweepComplete). :INITiate:STOP/PAUSe/RESume are rejected by the instrument and are not offered.

get_acquisition / set_acquisition

get_acquisition (read-only) returns continuous (true = continuous sweep, false = single), sweep_time_s and sweep_time_auto. set_acquisition takes continuous (boolean; sends :INITiate:CONTinuous ON|OFF), re-reads, and reports a mismatch in warnings.

single_sweep

Switches to single mode, runs one sweep and waits until it is complete, so the trace can be read. Optional timeout_s (default sweep time + 10 s; a timeout returns an error result). Leaves the instrument in single mode. Over the raw transport it stays silent for the sweep time and polls *OPC? (the instrument ignores input while sweeping); over VXI-11 (which queues commands) it writes *OPC? once and does one blocking read that returns when the sweep ends.

start_sweep / wait_for_sweep_complete

start_sweep sends :INITiate:IMMediate (method: "immediate", default) or :INITiate:RESTart. In single mode wait (default true) waits for completion; with wait: false call wait_for_sweep_complete next (nothing else can be sent meanwhile). In continuous mode the sweep is restarted, not awaited, and a warning says so. timeout_s is optional on all three tools.

Trace

Command syntax: programming-guide.md section 3 (ambiguities 8 and 9). Trace data is read by get_trace_data (below). Behavior verified on the UTS3036B: see verified-hardware.md. Traces are numbered 1-6 on the UTS3036B; the tools use the per-trace forms :TRACe<n>:....

get_trace

Read-only. Optional trace (1-6; all six when omitted). Returns selected_trace, the global average_count and average_type (voltage, power, log), and traces: [{trace, type, update, display}] where type is write, average, max_hold or min_hold, update is true for a live trace and false for view/hold (frozen), and display is whether the trace is shown. Trace math cannot be read: :CALCulate:MATH? is not supported and stalls the instrument, so it is never queried.

set_trace

All parameters optional, at least one: selected_trace (1-6, :TRACe:SELect); trace (1-6) with type, update and/or display for that trace (type, update and display require trace); average_count (integer 1-999, global) and average_type (voltage, power, log, global). Sent in this order: select, type, update, display, average type, average count. Everything is re-read; a value the instrument did not take is a warnings entry. The average count is validated to 1-999 before sending because the instrument accepts 0 (becomes 1) and 1000 (kept as 1000, not clamped). Changing a trace type, the update state or the average count restarts that trace's hold/average (see restart_trace).

restart_trace

Restarts max hold, min hold or averaging. The instrument has no clear-trace command; this tool re-sends :TRACe<n>:TYPE <current type>, which was verified to clear and restart the accumulation. :INITiate:RESTart and :INITiate:IMMediate (start_sweep) do not restart a hold. Optional trace (1-6); without it every trace that is not in write type is restarted (and the result says when there was nothing to restart). Returns restarted: [{trace, type}] and warnings.

set_trace_math

Sends :CALCulate:MATH TRACE<n>,<op>,TRACE<a>,TRACE<b>,<offset>. trace (1-6), operation (off, power_difference, power_sum, log_difference, log_offset), operand_a and operand_b (1-6; required unless off), offset_db (-100..100, only for the two log operations, default 0). The setting cannot be read back (no query), so the result carries trace_state (the trace's type/update/display) and a warning; verify through the trace data. Enabling math switches the trace on (update and display true) and off does not switch it off again, so restore that with set_trace.

get_trace_data

Read-only. Reads the amplitude values of one trace with the computed axis. Parameters (all optional): trace (1-6, default the selected trace from :TRACe:SELect?; anything else is rejected before anything is sent), format (ascii default, real32, real64), fresh (default false), max_points (integer >= 2), include_axis (default true).

Returns compact JSON: trace, trace_type (write, average, max_hold, min_hold), trace_updating, trace_displayed, unit (dBm, dBmV, dBuV, V, W from :UNIT:POWer?), format, points (:SENSe:SWEep:POINts?), axis, summary, decimation, warnings, values and, with include_axis, frequencies_hz (or times_s in zero span).

  • Axis. axis.kind is frequency with start_hz, stop_hz, span_hz, step_hz; the point axis is linear, f_i = start + i*(stop-start)/(points-1) (one point sits at the start). In zero span (start equals stop) there is no frequency axis: axis.kind is time with zero_span: true, center_hz and sweep_time_s, times_s runs from 0 to the sweep time, the summary has peak_time_s instead of peak_frequency_hz, and a warning says so.
  • Summary covers the full trace even when the response is decimated: points, finite_points, non_finite_points, min, max, mean (plain mean of the values in the trace's unit, so a mean of dBm values is a log average), peak_index, peak_level and peak_frequency_hz. Non-finite values (NaN or infinity) are returned as null, left out of the statistics and reported in warnings.
  • Format. ascii is the default: no state change, about 14 KB, 7 significant digits, and the fastest read on the unit. real32 is a #4 block of big-endian IEEE 754 floats (about 4 KB) with the same 7 digits as ASCII (REAL32 and ASCII gave identical values on hardware); real64 (about 8 KB) is the full double. For the binary formats the instrument's :FORMat is set for the read and put back in a finally (as ASCii, the setter form, although the query answers ASC8; if something else had set REAL32 or REAL64 before, that format is restored). If the restore fails the error says "Could not restore the data format" and names the command to send with scpi_command; a read error and a restore error are both reported.
  • fresh. Runs one single sweep first (the same method as single_sweep, valid on both transports) so the values are a complete, current sweep, reads, and switches continuous mode back on if it was on (fresh.continuous_restored). If that restore fails the error says so and points to set_acquisition. Without fresh the values are whatever the trace holds now, which in continuous mode can be a sweep in progress, and a frozen (view/hold) or hidden trace is reported in warnings.
  • max_points. When the trace has more points, peak-preserving min/max decimation is applied: the trace is cut into floor(max_points/2) equal buckets and the minimum and maximum of each are kept in order, so the global peak and minimum are always in the result. decimation says decimated: true, original_points, returned_points and the method; indices gives each returned point's index in the full trace; warnings repeats it. Nothing is decimated silently.
  • Errors. A malformed, empty, short or long reply is an error that names the command and quotes the start of the reply (or of the block); a value count different from :SENSe:SWEep:POINts? is an error naming both counts; a timeout names the command.

Markers

Command syntax: programming-guide.md section 4 (markers). Markers are numbered 1 to 10; every tool validates the number before it touches the instrument. Verified on the UTS3036B: see verified-hardware.md.

get_marker

Read-only. Optional marker (1-10, default 1). Returns marker, mode (off, normal, delta, fixed), trace (1-6), and the marker's frequency x (Hz; seconds in zero span) and level y (current power unit, dBm by default) together. A delta marker also returns reference_marker, and its x and y are relative to that marker. An off marker has x and y of null: the instrument still answers X?/Y? for it, but with stale values.

set_marker

marker (1-10) plus at least one of mode (off, normal, delta, fixed), trace (1-6), reference_marker (1-10, another marker; used by delta mode) and x_hz (position in Hz, 0 or more). Sent in the order mode, trace, reference, x, then the marker is re-read and returned as marker with warnings for anything the instrument did not take. The instrument does not clamp x_hz to the sweep span (a position outside it is accepted and its level is meaningless), and a position set on a marker that is off is ignored (a warning says so): use mode: "normal" first or marker_search. Setting reference_marker on a marker that is off turns it on in delta mode (verified). For a delta marker the position is not compared because the readback is relative.

marker_search

marker and action: peak (:MAXimum), next_peak (:MAXimum:NEXT), next_left / next_right (:MAXimum:LEFT / :RIGHt), minimum (:MINimum) or peak_to_peak (:PTPeak). Returns the marker (frequency and level together) after the search. A search turns a marker that is off on in normal mode; peak_to_peak leaves the marker in delta mode (and says so in note). Peaks must satisfy the criteria of get_peak_criteria. The search acts on the trace the marker is assigned to and the currently displayed data: use single_sweep first for a reproducible result.

marker_to_instrument

marker (must be on) and target: center, step (center frequency step; this turns automatic step off), start, stop or reference_level, sent as :CALCulate:MARKer<n>:SET:<target>. Returns target, the new value and its unit (Hz, or the current power unit for reference_level). A marker that is off is refused before anything is sent. The instrument rejects a center that would put the span outside its range with -224, which is returned as an error.

markers_all_off

No parameters. Sends :CALCulate:MARKer:AOFF and reads every marker's mode to confirm; returns all_off and warnings for any marker still on. The instrument's marker positions and traces keep their last values.

get_peak_criteria / set_peak_criteria

Peak search criteria: threshold_dbm (a peak must be above it; -200 to 30) and excursion_db (a peak must rise this far above its valleys; 0 to 100), each with a threshold_auto / excursion_auto flag. In automatic mode the instrument computes the value from the trace and reports it. get_peak_criteria is read-only and takes no parameters. set_peak_criteria needs at least one of the four parameters; giving a value turns that criterion's automatic mode off (verified), *_auto: true switches it back on and cannot be combined with a value for the same criterion. Values outside the documented ranges are refused before anything is sent; the readback is compared and clamped values are warnings.

get_marker_function / set_marker_function

A marker's function: noise (noise level per Hz at the marker), band_power (power in a band around the marker), band_density (power density in that band) or off. get_marker_function is read-only (marker 1-10, default 1) and returns function and band_span_hz. set_marker_function needs marker and at least one of function and band_span_hz (Hz, above 0); the span is sent first. The function's result is the marker's level: read it with get_marker (y). Values the instrument did not take are warnings.

get_marker_ndb / set_marker_ndb

The N-dB bandwidth measurement of a marker: the width in Hz of the trace around the marker where it is within N dB of the marker level. get_marker_ndb (read-only) returns enabled, ndb (negative dB) and bandwidth_hz (null while off, because the result query is only sent while on, and null when the instrument answers nan because the trace does not fall N dB on both sides of the marker). set_marker_ndb takes marker and at least one of enabled and ndb (-140 to -0.01, checked before anything is sent); N is sent first. Readback differences are warnings.

get_frequency_counter / set_frequency_counter

The marker frequency counter: enabled, gate_time_s and frequency_hz (the counted frequency of the selected marker via :CALCulate:MARKer:FCOunt:X?, only queried while on; null while off or when the instrument reports its invalid value 9.91e37). get_frequency_counter is read-only and takes no parameters. set_frequency_counter takes enabled and/or gate_time_s (1e-6 to 0.5 s, checked before anything is sent). Readback differences are warnings.

Limit lines

Command syntax: programming-guide.md section 5 (limit lines). Lines are numbered 1 to 6 (1 to 4 on the 1000-series models). Verified on the UTS3036B: see verified-hardware.md. There is no pass/fail tool: the manual documents no pass/fail read-back for limit lines. To judge a trace against a line, read both (get_limit_line and get_trace_data) or take a screenshot.

get_limit_line

Read-only. Takes line (1-6, default 1) and returns displayed, type (upper or lower), trace (the trace the line is tested against), margin_db, margin_enabled and points (a list of frequency_hz, amplitude_dbm, connected; the instrument always reports the first point as connected). The analyzer sends no reply at all to the point query of a line without points, so the tool reads it last with a 1.5 s wait and takes that silence as an empty list; a failure of any of the other queries is an error, never an empty line.

set_limit_line

Create or edit a limit line. line (1-6, default 1) plus at least one of displayed, type (upper: the trace must stay below the line, lower: above), trace (1-6), margin_db (-40 to 40), margin_enabled and points. points replaces the whole list: 1 to 10 entries of {frequency_hz, amplitude_dbm, connected?} in ascending frequency order (equal frequencies make a vertical step; connected defaults to true and is ignored for the first point). All of this is checked before anything is sent. The analyzer keeps at most 10 points and silently ignores a longer list, so the tool refuses more than 10. Setting points also turns the line's display on, so displayed is sent last; setting the margin switches its state on, so margin_enabled is sent after it. The line is read back; anything the instrument did not take is listed in warnings.

delete_limit_line

Delete the points of one limit line (line) or of all of them (all: true); type, trace and margin settings stay. The analyzer finishes a delete a moment after it acknowledges it, so the tool re-reads the points (up to four times, 0.3 s apart) until the line is empty; for all it checks line 1. A line that still has points is reported in warnings. Returns deleted ([n] or "all").

Tracking generator

Command syntax: programming-guide.md section 6 (the manual calls the tracking generator the "trace source"). Verified on the UTS3036B, which has the option: see verified-hardware.md. On an analyzer without the option the first query gets no reply, and both tools then return an error naming :OUTPut:STATe? and saying the option is probably not installed; set_tracking_generator sends nothing in that case. The message also names the other cause of no answer: a measurement is active. The generator is unavailable in the measurement modes (entering one switches the output off and :OUTPut:STATe is undefined there). When this server entered the measurement (configure_measurement), both tools say so at once without sending anything: "the tracking generator cannot be changed while a measurement is active ... call exit_measurement first" (not a missing option); exit_measurement restores the output that was on.

get_tracking_generator

Read-only. Returns enabled (output on/off), level_dbm, level_offset_db, reference_trace, normalize and normalize_ref_level_db.

set_tracking_generator

Takes at least one of enabled, level_dbm (-40 to 0), level_offset_db (-200 to 200), reference_trace, normalize and normalize_ref_level_db (-100 to 100); ranges are checked before anything is sent (the instrument clamps silently outside them). Level and offset are set before the output is switched on. reference_trace only takes while the output is on (otherwise a warning says so). Switching the output off also re-asserts automatic RBW and VBW when they were automatic: the instrument otherwise leaves the RBW at a stale 300 kHz (auto sweep time 264 ms instead of 2.6 ms) with both auto flags still reading on; the tool waits 0.6 s, re-sends the auto commands and warns if the RBW did not return. With the output on the auto sweep time is about 2.15 s. Readback differences are warnings. On some benches the output is wired to the analyzer input: switch it off when done.

Measurement function module

Command syntax: programming-guide.md section 1 (:INSTrument[:SELect] {SA|EMI|AD}). The owner's UTS3036B has the spectrum analyzer (SA), EMI and analog demodulation (AD) modules licensed: see verified-hardware.md. No SCPI query reports which modules are licensed; switching to an unlicensed one is refused silently by the firmware (no SCPI error, module unchanged), which is indistinguishable here from a link glitch, so select_instrument_module reports a mismatch as a warning, never a guess at the cause.

get_instrument_module

Read-only, safe in any module. Returns module: spectrum_analyzer, emi or analog_demod. A reply other than SA/EMI/AD is an error quoting it.

select_instrument_module

module: sa, emi or ad. Sends the select and reads it back; does not restore the previous module afterwards (the caller chose the switch) and does not change the front-panel lock behavior beyond that of an ordinary command (see "Front-panel lock" in verified-hardware.md). If the readback does not match what was requested, that is reported as a warnings entry, not an error - this is expected when the module is not licensed. A timeout reading the module back names the analyzer's own documented quirk right after a mode switch (a reply can be silently dropped on the same connection even though the switch took); call get_instrument_module again, ideally after reconnecting.

Every other tool that talks to the instrument is spectrum-analyzer-only (#178, #191): the same-named command often means something different per module (for example :SENSe:FREQuency:CENTer under EMI), so before sending anything each one checks the module fresh (no caching across calls) and refuses with an actionable message naming select_instrument_module if it is not SA. get_instrument_module, select_instrument_module, identify, scpi_query, scpi_command, connect, disconnect, discover_instruments and list_instruments are exempt, since they must work in any module or send none of their own.

Analog demodulation module

Command syntax: programming-guide.md section 1; state save/restore follows docs/research/analog-demodulation.md section 1 and the same read-back pattern exit_measurement uses. get_instrument_module/select_instrument_module remain the general-purpose module tools; these two add only the spectrum-analyzer state round trip around :INSTrument:SELect AD/SA.

enter_analog_demod

Switches to the AD module and reads it back. From spectrum analyzer mode it saves the frequency, bandwidth, sweep time and continuous-sweep settings first (exit_analog_demod restores them); calling it again while already in AD keeps whatever was saved the first time rather than overwriting it. If the readback does not match (an unlicensed module is refused silently by this firmware, no SCPI error), that is a warnings entry, not an error - the same convention as select_instrument_module. Leaves the instrument in AD on purpose. Does not select or preset an AM/FM demodulation type (:CONFigure:AM/:CONFigure:FM reset that type's parameters and are out of scope here) and does not save or restore the tracking generator (unlike leaving a measurement, nothing confirms the AD module switch touches it).

exit_analog_demod

Switches back to spectrum analyzer mode and, unless restore_settings is false, restores the settings enter_analog_demod saved, verified by read-back with one resend before giving up (same as exit_measurement). Always sends the switch, even if this server never called enter_analog_demod (for example after select_instrument_module); a clear warnings entry says so when there is nothing saved to restore, rather than guessing at settings this server never saw.

get_ad_settings / set_ad_frequency / set_ad_demod_timing

Research-stage: field names and ranges come from the programming manual via docs/research/analog-demodulation.md sections 2-3. The AM/FM prefix each AD command needs is read fresh from :CONFigure? on every call, rather than assumed, since the demodulation type can change outside this server's own calls. A handful of the getters have been live-verified (:AM:FREQuency:SPAN?, :AM:BANDwidth:CHANnel?/:RESolution?, the AF spectrum start/stop, :AM:DWSWeep:TIME?, input impedance - see verified-hardware.md); :CONFigure? itself, the demod-time/averaging getters, every FM-prefixed form, and every setter in this group are not yet live-verified - clamp-vs-error behavior is UNKNOWN until probed with the owner present.

get_ad_settings is a read-only snapshot: demod_type, center frequency and step, RF span, resolution bandwidth, demodulation bandwidth, AF spectrum start/stop and bandwidth, demod window sweep time, demod time and averaging, and input impedance. A silent or malformed reply to any query is an error naming that command - no partial snapshot is ever returned. Filters, audio and display/marker settings are separate tool groups.

set_ad_frequency configures center_hz (50 Hz to the instrument's max minus 50 Hz, per the manual), span_hz (100 Hz-12.8 MHz), rbw_hz/rbw_auto (100 Hz-1 MHz), demod_bandwidth_hz (400 Hz-25 MHz), center_step_hz/center_step_auto. set_ad_demod_timing configures window_sweep_time_s (1 us-500 ms), demod_time_s/demod_time_auto, average_state, average_count (1-999). Both require at least one parameter, reject ..._auto: true combined with its matching explicit value, re-read everything requested afterwards, and report any instrument-adjusted value as a warnings entry. Neither restores the previous state (the caller chose the change). Both require the AD module.

set_ad_filters

Sets one AD audio filter family: low_pass_filter (OFF, LPF300, LPF3K, LPF15K, LPF30K, LPF80K, LPF100K, LPF300K), high_pass_filter (OFF, HPF20, HPF50, HPF300, HPF400), or band_pass_filter/weighting (OFF, AWEighting, CWEighting, CCIR1k, CCIR2k, CUNWeighting) - low_pass_filter and high_pass_filter can be set together in one call (live-verified: the instrument does not turn either off for the other); band_pass_filter is mutually exclusive with both (live-verified: setting it turns off an active low-/high-pass filter and vice versa) and is refused locally if requested together with either. This tool reads all three before and after the change and reports whichever family the instrument turned off that was not explicitly requested as a warnings entry. The AM/FM prefix is read fresh from :CONFigure? each call. Requires the AD module. Live-verified (2026-11-12).

set_ad_audio

Sets the AD module's speaker (on/off) and/or volume_pct (0-100), both read from and shared with the rest of the instrument. Turning the speaker on, or raising the volume while it is already on, requires confirm: true - :SPEaker ON plays the demodulated signal on the loudspeaker and headphones immediately, which would be an unexpected tone in a shared lab (research doc, risk 3). By default, once the requested change is confirmed by read-back, this tool restores the speaker and volume to what they were immediately before the call (restored: true in the result); pass keep: true to leave the new state in place instead. Requires the AD module. The read path is live-verified (2026-11-12); the speaker-on/volume-raise write path remains intentionally not exercised against real hardware outside a confirmed, supervised live session.

get_demod_result

Read-only numeric result window: center frequency, carrier power, carrier frequency error, modulation rate, SINAD, distortion/total Vrms, THD, and peak+/peak-/(pk-pk)/2/RMS for both the current and max-hold rows, without a screenshot. Wraps the undocumented bare :FETCh:AM?/:FETCh:FM? query (the 15-value metrics line), confirmed live on this firmware (see verified-hardware.md, "Installed options and AD results readback", and research doc section 6a). The AM/FM type is read fresh from :CONFigure? each call. Does not expose the RF spectrum trace, demodulated waveform or AF spectrum views (other :FETCh suffixes), which remain screenshot-only. Requires the AD module. Units beyond what the live example confirmed (Hz, dB) are plausible, not independently verified.

set_ad_display

Configures one display window (1 RF spectrum, 2 demodulation waveform, 3 AF spectrum)'s Y-axis reference_level and/or per_division, and (window 3 only) af_spacing (logarithmic/linear). The unit of reference_level/per_division depends on the window and the current demod type (read fresh from :CONFigure?): dBm for window 1, modulation depth as a fraction (AM) or Hz deviation (FM) for window 2, and an unconfirmed unit for window 3 (the programming manual's wording is ambiguous there). Requires at least one parameter and the AD module; af_spacing is rejected outside window 3. Re-reads everything requested and reports any instrument-adjusted value as a warnings entry. Live-verified for windows 1 and 3 (2026-11-12); window 2 not separately exercised.

ad_marker / ad_marker_search / get_ad_marker / ad_markers_all_off

AD's marker tools, separate from the spectrum-analyzer module's set_marker/marker_search/get_marker/markers_all_off: the AD module has six markers per demod type (not SA's ten), and AD-specific trace names (rf_spectrum, demod, demod_average, demod_max, demod_min, af_spectrum) rather than SA's numeric 1-6 trace slots. x's unit is Hz for the two spectrum traces and seconds for the four demodulation-waveform traces; y's unit is dBm on the RF spectrum trace, modulation depth (AM) or Hz deviation (FM) on the waveform traces, and unconfirmed on the AF spectrum trace (the same manual ambiguity as set_ad_display's window 3).

ad_marker sets a marker's mode (off/normal/delta/fixed), trace, reference_marker (delta mode) and/or x, applied in that order, then re-reads and reports every instrument-adjusted value in warnings; a position set while the marker is off is a no-op, reported as a warning rather than silently ignored. ad_marker_search runs a peak/next_peak/next_left/next_right/minimum/peak_to_peak search, turning an off marker on. get_ad_marker reads one marker's mode, trace, position and value together with their units. ad_markers_all_off turns every marker off and confirms it by reading all six. All four require the AD module. Live-verified (2026-11-12) against a real external signal, including delta mode and the all-off action.

EMI function

Research-stage: field names, ranges and groupings come from the programming manual via docs/research/emi-function.md; none of it has been verified live yet (EMI-12, #187, needs the owner present at the analyzer). Requires the EMI module (select_instrument_module); every tool here refuses with an actionable message otherwise.

get_emi_settings

Read-only snapshot, grouped so one call can stay short: standard, sequence, scan_table (10 bands), meters, detectors, traces, search, frequency, amplitude, trigger. The optional groups argument selects which to read; by default scan_table is omitted (10 bands x ~11 queries each, by far the most expensive group) and every other group is included. A silent or malformed reply to any query is an error naming that command - no partial snapshot is ever returned.

set_emi_scan_table

Edit the EMI module's 10-band scan table. bands: an array of per-band edits (band 1-10, plus any of enabled, start_hz, stop_hz, rbw_hz/rbw_auto, time_s/time_auto, attenuation_db/attenuation_auto, points, preamp). presets: load a CISPR band into a band (band, cis: A 9 kHz-150 kHz, B 150 kHz-30 MHz, C 30 MHz-300 MHz, D 300 MHz-1 GHz, CD 30 MHz-1 GHz, E 1 GHz-fmax); applied before any bands edit for the same band number. Validated before sending: stop_hz must be at least 100 Hz above start_hz, attenuation_db an even number 0-50, time_s between 1 ms and 4000 s, points 11-40001 (enforced by the schema). The whole 10-band table is re-read after the edit (before and after, to catch coupling) and the result reports: any field the instrument adjusted away from what was requested, and any band that was not requested but whose start/stop moved anyway - the neighbor-band coupling the user manual describes when bands overlap or cross. Requires the EMI module.

set_emi_measurement

Configure the EMI module's measurement. All parameters optional, at least one required: standard (none/cispr; CISPR fixes RBW to 200 Hz/9 kHz/120 kHz/1 MHz and the EMI 6 dB filter, none allows 1-3-10 RBW steps 1 kHz-3 MHz), sequence (scan, search, scan_search_measure, scan_search, search_measure, measure), remeasure (current, all, marked), trace_detectors/meter_detectors (arrays of {index, detector}, trace/meter index 1-3), meter_dwell_s (0.001-100 s, one value for all meters), final_detectors (array of {index 1-3, detector?, dwell_s? 0.001-60 s, limit_line? 1-6}), average_count (1-999), average_type (voltage, power, log), search_mode (peak_only, peak_and_limit, segment_and_limit), peak_count/segment_count (1-50). detector is one of positive, quasi_peak, average_continuous, average, negative; the user manual documents quasi-peak, EMI average (average_continuous) and average as mutually exclusive, at most two together, and this is checked per detector group before anything is sent. Everything requested is re-read afterwards and any instrument-adjusted value is a warnings entry. Leaves the new state (nothing is restored). Requires the EMI module.

get_emi_limit_line / set_emi_limit_line / delete_emi_limit_line

The EMI module's limit lines, reusing the SA limit-line tool's design (same parameters, same :CALCulate:LLINe<n> commands). Requires the EMI module. The firmware's own command-callback audit (private uts3036b-firmware repo) found that the EMI-prefixed :CALCulate:FSCan:LLINe<n>... commands and these plain :CALCulate:LLINe<n>... ones share the exact same callback functions as the SA limit-line implementation. Despite the shared callback, the point cap is not the same: live-verified at 12 points (2026-11-12, EMI-12, #187, owner present), not SA's 10 — 10, 11 and 12 all round-trip exactly; 15 and 20 are both silently truncated to the same 12. set_emi_limit_line adds two EMI-specific actions over set_limit_line, sent before any other setting: copy_from (copies another line's points via :LLINe<n>:COPY LLINE<m>; must name a different line) and build_from_trace (builds points from a trace's current data via :LLINe<n>:BUILd TRACE<m>, trace 1-3 since EMI has only 3 traces). Either action producing no points (an empty source) is a warnings entry. delete_emi_limit_line is unchanged from delete_limit_line beyond the module requirement. See Limit lines above for the rest of the parameters and behavior.

get_emi_correction / set_emi_correction

The EMI module's 10 correction (antenna/cable) factor slots ([:SENSe]:CORRection:CSET<n>, programming manual pp.125-127). Requires the EMI module. get_emi_correction (slot 1-10) reads whether the slot is enabled and its point list (frequency_hz, gain_db). set_emi_correction takes slot plus any of points (replaces the whole list, 1-10 points in ascending frequency order), enabled (on/off) or delete (clears the slot's points; not combined with points); or all_off/delete_all alone, without slot, to turn off or clear every slot at once. The 10-point cap was live-verified (2026-11-12, this unit, EMI licensed): writing 10 points round-trips exactly; writing 11 is silently rejected, with no error on *ESR?, *STB? or :SYSTem:ERRor?, and empties the slot's data the same way an over-cap limit-line write does — set_emi_correction refuses a request over 10 points locally before sending anything, rather than relying on that silent rejection. The slot is re-read after every change; anything the instrument did not take is a warnings entry.

get_emi_marker / set_emi_marker / emi_marker_search / emi_markers_all_off / get_emi_peak_criteria / set_emi_peak_criteria

EMI module markers (6 markers, 3 traces; programming manual pp.94-99). :CALCulate:FSCan:MARKer<n>:MODE/:TRACe/:REFerence set mode (off/normal/delta/fixed), trace and the delta-mode reference marker; marker position and level readback also uses the FSCan-prefixed :CALCulate:FSCan:MARKer<n>:X?/:Y? form, not the plain, non-prefixed form the manual's own text suggests — live-verified 2026-11-12 (EMI-12, #187, owner present): the plain form gets no reply at all on a query and -113 Undefined header with no effect on a set, while the FSCan-prefixed form works correctly for both. get_emi_marker (marker 1-6) reads mode, trace, frequency and level together; a delta marker's values are relative to its reference marker, and an off marker reads x/y as null. set_emi_marker sets mode/trace/reference_marker/frequency_hz (applied in that order), re-reads, and reports anything the instrument did not take as a warnings entry; frequency_hz on an off marker has no effect, a warnings entry says so. emi_marker_search moves a marker with action: peak, next, left, right, minimum (turns an off marker on) and returns its new position. emi_markers_all_off turns every marker off and confirms by re-reading all 6. get_emi_peak_criteria/set_emi_peak_criteria (:CALCulate:FSCan:MARKer:PEAK:THReshold/:EXCursion, each with a :STATe auto flag) are the EMI module's own peak-search criteria, separate from the SA get_peak_criteria/set_peak_criteria tools. Requires the EMI module.

emi_signal_list / get_emi_signal_list_sort / set_emi_signal_list_sort

The EMI module's signal table (peak table; programming manual pp.96-104). The manual documents no command to read the table's contents, and almost none of these actions has a read-back - emi_signal_list can only confirm the command was accepted (no instrument error), never what it actually did; the result says so and names the screenshot tool (over VXI-11) as the visual verification path. action is one of: add_marker (a marker's frequency to the list, needs marker 1-6), add_marker_measurement (the measurement marker's value to the list), replace_with_meter/append_meter (a row from a meter reading; replace_with_meter needs meter 1-3, append_meter takes none per the manual's own listing), delete/mark/clear_mark (need signal 1-1000), mark_all/clear_marks, set_meter (couple a marker to a meter, needs meter), zoom_in/zoom_out (signal optional). Commands whose argument the manual writes with a leading space (:SLISt:DELete:SIGNal <n>, :SLISt:REPLace:METer <n>, etc.) are sent as a header plus a separate value, not a numeric header suffix - inferred from the manual's own notation, not independently live-verified. get_emi_signal_list_sort/set_emi_signal_list_sort are the one documented queryable part: sort type (frequency, or amplitude/delta relative to trace 1-3) and order (ascending/descending); set_emi_signal_list_sort reads both back and reports anything the instrument did not take as a warnings entry. Requires the EMI module.

emi_run / emi_stop

Start and stop an EMI scan sequence (:INITiate:IMMediate/:INITiate:FSCan:CLEar:IMMediate/:INITiate:STOP, programming manual pp.114-115). Live-verified 2026-11-12 with the owner present: a cold scan-only sequence left the instrument silent to every command, including :SYSTem:ERRor?, for 25 s or more while it ran, then answered again on its own once it stopped - matching the manual's note that a scan-only sequence finishes by itself. There is no documented "finished" query, so emi_run infers completion the same way wait_for_sweep_complete infers an SA sweep's end: it sends :INITiate2:CONTinuous OFF (cannot read the prior state first - :INITiate2:CONTinuous? got no reply at all in EMI, live-verified, unlike SA where it answers; this is said in the result's warnings), then :INITiate:IMMediate or, with clear_list: true, :INITiate:FSCan:CLEar:IMMediate (clears the signal list and starts), then polls *OPC? until anything answers again or timeout_s (default 45, 0-600) passes. On success the result is {started: true, finished: true, waited_s, error_queue, warnings} - error_queue is whatever :SYSTem:ERRor? reports afterward, reported as information rather than thrown as a failure (the scan's own commands were live-verified not to raise an instrument error). On a timeout the result is still a non-error: {started: true, finished: false, waited_s, warnings} with a warning that the scan may still be running, since silence during a scan is expected, not a fault. Reading the resulting signal list is not possible through documented commands; use emi_signal_list actions or the screenshot tool over VXI-11. emi_stop sends :INITiate:STOP, live-verified to be accepted without an instrument error in EMI (unlike SA, where it gives -113); there is no read-back to confirm it actually halted anything, and the result says so. :INITiate:PAUSe/:INITiate:RESume were live-verified accepted the same way but are not exposed as tools (outside this issue's scope; no independent confirmation that pausing actually paused anything was taken). Requires the EMI module.

Measurements

Channel power, time-domain power, occupied bandwidth (OBW), adjacent channel power (ACPR), third-order intercept (TOI), carrier-to-noise ratio (C/N) and harmonics. The instrument has no query for the active measurement, and :CONFigure:<meas> presets the span, RBW and sweep time (for example span 3 MHz) which :CONFigure:SANalyzer does not put back, so these tools save and restore that state and never scan for the active mode (queries the instrument does not answer stall it, and over VXI-11 can drop the link). Spectrum analyzer mode is told by :SENSe:DETector:TRACe1?, which only it answers; one measurement is told by :SENSe:<meas>:AVERage:STATe?. While a measurement is active most spectrum analyzer queries (detector, trace type) get no reply. A probe that gets no reply costs about 4.5 s (the 1.5 s timeout plus a 3 s pause while the instrument recovers). Measurement names: channel_power, time_power, obw, acpr (:SENSe:ACPower), toi, cn (CNRatio), harmonics.

Parameters (configure_measurement and measure; * = common to all measurements): average* (on/off), average_count* (integer 1-999), average_control* (exponential, repeat), average_type* (voltage, power, log); integration_bandwidth_hz (channel_power, acpr or cn, greater than 0; channel_power is clamped by the instrument to 100 Hz-3.61 GHz); left_limit_s, right_limit_s (time_power only, 0-0.01 s; the instrument clamps the right limit to 10 ms); percent (obw only, above 0 up to 1; clamped to 0.9999) and xdb (obw only, -100 to -0.1 dB); carrier_bandwidth_hz (acpr, greater than 0), offset_hz (acpr: adjacent channel offset; cn: noise offset; 0 or more) and offset_side (acpr: both, negative, positive); noise_bandwidth_hz (cn, greater than 0); fundamental_hz (harmonics, greater than 0), harmonics_count (harmonics, integer 2-10, counts the fundamental; the preset is 2) and sweep_time_s (harmonics, 0.001-100 s). TOI has only the averaging parameters. A parameter for another measurement, or an out-of-range value, is refused before anything is sent. The parameters are read back after setting and anything the instrument changed is a warnings entry.

get_measurement

Read-only. Without arguments it asks the one query only spectrum analyzer mode answers and returns {"measurement": "spectrum_analyzer"}; if that gets no reply it is an error saying a measurement is active (it does not guess which). With measurement set it checks that measurement and returns its parameters; an error means it is not active.

configure_measurement

measurement plus optional parameters (above) and preset (default true: :CONFigure:<meas>; false: :CONFigure:<meas>:NDEFault; both change the span, RBW and sweep time). Enters the measurement, sets the parameters, reads them back and leaves the instrument in the measurement. When entered from spectrum analyzer mode the center, span, RBW, VBW, sweep time (each with its auto flag), continuous-sweep state and the tracking generator output (plus its level and offset when it was on; entering a measurement switches it off) are saved for exit_measurement (saved_settings: true). Returns measurement, parameters, warnings, saved_settings.

read_measurement

Read-only. measurement (must be active, else an error naming the silent query), timeout_s (default 30, 1-600) and the flag fetch, false by default. READ starts the measurement and waits for it (over VXI-11); fetch: true returns the latest result at once. Over raw TCP the instrument ends a READ reply without a line terminator, so FETCh is used instead and a warning says so. Results: channel_power gives channel_power_dbm and power_density_dbm_per_hz; time_power gives power_dbm; obw gives obw_hz, frequency_error_hz, xdb_bandwidth_hz and total_power_dbm (queried as separate documented sub-results; the combined OBW reply has seven values of which the manual documents only four; the four are not one atomic snapshot). acpr gives main_power_dbm, lower_power_dbm, lower_dbc, upper_power_dbm, upper_dbc (a side that is not measured, see offset_side, reads null with a warning; the combined :FETCh:ACPower? has only three values, so the sub-results are queried); cn gives carrier_dbm, noise_dbm, cn_db; toi gives lower_third_order and upper_third_order (frequency_hz, amplitude_dbm, dbc, toi_dbm), lower_tone and upper_tone (frequency_hz, amplitude_dbm) from four queries (the documented :FETCh:TOI:IPT? gets no reply and is not used; each product carries its TOI); harmonics gives fundamental (frequency_hz, amplitude_dbm), harmonics (a list of order 2..n with frequency_hz and amplitude_dbc, n = harmonics_count; the instrument pads its lists to 10 values and the tool drops the padding) and thd_percent, thd_dbc. With several queries per result only the first uses READ (it runs the measurement and waits), the others FETCh that result, so they are not separate measurements. A reply that is not a number list of the expected length (nan right after entering a measurement, before its first result) is an error naming the command.

exit_measurement

restore_settings (default true). Sends :CONFigure:SANalyzer, checks that spectrum analyzer mode answers again and, when configure_measurement saved settings, puts the center, span, RBW, VBW, sweep time and continuous state back and verifies them by reading every value and auto flag back: a field that differs is sent once more (the warnings say which), and if it still differs the call is an error naming each field with its actual and saved value (never a success). After the analyzer settings it restores the tracking generator output saved by configure_measurement (level and offset, then output on; readback verified, one resend, warning on a retry, error naming enabled, level_dbm or level_offset_db if it still differs; a failed analyzer restore does not skip it). With restore_settings=false a generator that was on stays off and a warning says so. A measurement entered another way has nothing saved: it only exits and a warning says the span, RBW and sweep time keep the measurement's values and the generator output was not restored. In spectrum analyzer mode with nothing saved it does nothing.

measure

One shot that leaves the instrument as it found it: from spectrum analyzer mode it saves the state (including the tracking generator output, level and offset when on), enters measurement (with preset and the parameters), reads the result with READ (FETCh after a 1.5 s settle over raw TCP), and always, even after an error, exits to spectrum analyzer mode, restores the saved state (frequency settings and the tracking generator output, which entering a measurement switches off) and verifies it by readback (try/finally; a mismatch is retried once, then reported as an error naming the field). With the requested measurement already active it only reads and puts any parameters you passed back to their previous values. It refuses when the instrument is in neither spectrum analyzer mode nor the requested measurement (use exit_measurement). Returns measurement, result, restored, warnings; if the read fails the error says the instrument was restored, and if the restore fails the error says to check with get_measurement/exit_measurement.

Screenshot

screenshot

Read-only, no parameters. Captures the analyzer screen (1280 x 800) and returns it as an MCP image content block (image/png) followed by a text line with the size (about 90 KB, taking about 3 s). It reads the picture format (:SYSTem:PICture:FORMat?, BMP or PNG), selects PNG, queries :DISPlay:DATA? as a # block, checks the PNG signature, and always restores the original format (try/finally, also after a failed transfer). If the restore fails the result is an error that says so and still carries the image.

Needs a transport whose large replies arrive complete (completeLargeBlocks: VXI-11). Over raw TCP the instrument stops after exactly 144,800 bytes (issue #42), so the tool refuses at once, sends nothing, and the error tells you to reconnect with transport: "vxi11" (or auto) or set UTS_TRANSPORT. Errors name the command: a timeout names :DISPlay:DATA?, an unreadable format reply names :SYSTem:PICture:FORMat?, an empty or non-PNG reply is rejected, and an instrument error while selecting the format carries its code. The tool does not save files; the client receives the image. The screen shows whatever the instrument displays (including its system information page if open).

Selecting an instrument (all tools)

Every tool except discover_instruments and list_instruments takes an optional instrument argument (a name from list_instruments: 1-32 letters, digits, _ or -; connect, disconnect and identify call it name). The tool then runs on that instrument's own connection, queue and exclusive session, so calls to different instruments proceed in parallel while calls to the same instrument stay serialized.

  • Omitted: the default instrument (UTS_IP), or the only configured instrument whatever its name. With one instrument nothing changes for existing users.
  • An unknown name is an error that lists the configured names and sends nothing; so is an omitted name when several instruments are configured and none is called default.
  • An instrument that is not connected fails with the usual not-connected message; the other instruments are not affected.

Error handling (all tools)

Tools never throw; failures are isError results. A malformed or empty reply names the command and quotes the reply; a timeout names the command; setters report instrument errors with their code and the commands sent, and say when settings may be partially applied (also when the readback after a successful send fails). single_sweep failures say when the instrument may have been left in single sweep mode. A dropped connection is reported as Connection to the spectrum analyzer was lost (<reason>) while sending|querying <command> and the next call says to use connect to reconnect; the command in flight when a setter fails is named as possibly applied. See AGENTS.md "Error handling standard".

Concurrency (all tools)

An MCP client may send several tool calls at once. The analyzer has one command stream, one global error queue, and ignores every command while a single sweep runs, so interleaved calls would corrupt each other (one call's *CLS clearing another's errors, a getter reading half-applied settings, queries lost during a sweep). The server therefore runs tool calls one at a time per instrument, in the order they arrive (first in, first out): every tool is registered through defineTool (src/tools/util.ts), which runs the whole handler, from the first command to the last readback, as one exclusive session on the connection (runExclusive in src/connection.ts, src/sessionLock.ts). Parallel calls are safe and each one gets its own result and its own errors. This works the same on every transport.

  • A long operation blocks the others. While single_sweep (or start_sweep / wait_for_sweep_complete) holds the instrument for the sweep time, other calls wait their turn.
  • Instrument busy. A call that waits longer than the limit fails with Instrument busy: waited <n> s for another tool call (<tool>) to finish and gave up...; nothing was sent. Retry later or raise the limit. The limit is the UTS_BUSY_TIMEOUT_S environment variable, in seconds (default 120; an invalid value uses the default).
  • Released on every path. A session ends when its tool returns, fails, times out, or the connection is closed or lost; closing or losing the connection also fails every waiting call (Connection closed or Connection to the spectrum analyzer was lost (...)). disconnect does not wait for a running call.
  • Invalid input is refused first. Argument errors that need no instrument (schema errors and the tools' pure validation) are returned before waiting for the lock.
  • Limit. A sweep started with start_sweep wait=false releases the lock when that call returns, so another call can run while the instrument is still sweeping (and be lost); use single_sweep or start_sweep with wait=true unless you call wait_for_sweep_complete next.

Environment variables

Variable Description
UTS_IP Analyzer hostname or IP, or auto (find it on the local network); enables auto-connect and is the default for connect
UTS_PORT TCP port of the raw transport, default 5025
UTS_TRANSPORT auto (default: VXI-11 if the instrument registers it with the portmapper, else raw), raw or vxi11; default for the connect tool's transport parameter. Setting UTS_PORT without UTS_TRANSPORT means raw
UTS_VXI11_DEVICE VXI-11 logical device name, default inst0
UTS_VXI11_PROBE_TIMEOUT_S Total seconds (1 to 120, default 15) auto and vxi11 may spend reaching VXI-11: up to three attempts to reach the portmapper on port 111 and open the link, with doubling pauses, all inside this budget. Raise it for a slow routed path (VPN, Tailscale)
UTS_INSTRUMENTS Additional named instruments, name=host[:port],... or JSON; each is connected at startup on its own connection (own queue and lock). See list_instruments
UTS_BUSY_TIMEOUT_S How long a tool call waits for another one to finish before failing with Instrument busy, in seconds (default 120); see Concurrency