Error results always say which command failed and, where it helps, what to do next. This page explains the common ones.
"Not connected to the spectrum analyzer. Use the 'connect' tool first."
No connection is open. Either set UTS_IP in the MCP client's configuration (the server then connects at startup) or call connect with a host. If auto-connect failed, the reason was written to the server's stderr (check your MCP client's server log).
host: "auto" / discover_instruments finds nothing
- Discovery uses a UDP broadcast on port 111 on each of the computer's own IPv4 subnets, so the analyzer must be powered, have LAN enabled and be on the same subnet; it does not cross routers (give the address to
connectinstead). Firewalls and VLAN or Wi-Fi client isolation block broadcasts. - With several analyzers,
connectwithautolists them and asks for an explicithost. Analyzers that are already connected under another instrument name are left out of the count.
"Unknown instrument" / "Several instruments are configured ... none is named default"
You are driving several analyzers. Call list_instruments to see the names, then pass instrument (name for connect, disconnect and identify). Calls without a name use the instrument called default (UTS_IP), or the only one configured. An instrument that is not connected fails alone; the others keep working.
connect fails ("Connection failed", timeout, refused)
- Check the address and that the analyzer is on the same network as your computer.
- Test the port:
nc -z -w 2 <ip> 5025(macOS/Linux) orTest-NetConnection <ip> -Port 5025(Windows PowerShell). - Check firewalls between you and the instrument. Raw SCPI uses TCP 5025.
- The host must be a hostname or IP address (letters, digits,
.,-,_,:); anything else is rejected before connecting. -
UTS_PORTmust be an integer 1-65535.
A call fails with "Instrument busy"
Another tool call (named in the message) held the instrument longer than UTS_BUSY_TIMEOUT_S seconds (default 120), typically a single_sweep with a long sweep time. Nothing was sent by the call that failed. Retry later, shorten the sweep time, or raise UTS_BUSY_TIMEOUT_S. Parallel tool calls are otherwise safe: they run one at a time in arrival order.
VXI-11 does not connect (UTS_TRANSPORT=vxi11)
VXI-11 needs outbound TCP to port 111 (the portmapper) and to the port the portmapper reports, which changes when the analyzer restarts. Check the firewall. The default UTS_TRANSPORT=auto falls back to raw SCPI and logs the reason to the server's stderr, and UTS_TRANSPORT=raw skips VXI-11 altogether. Over a slow routed path (VPN, Tailscale) raise UTS_VXI11_PROBE_TIMEOUT_S (total seconds, default 15, up to 120): the portmapper connect and the link are retried inside that budget, and the log line says why auto fell back. UTS_VXI11_DEVICE must be 1 to 64 characters from letters, digits and . _ , -.
"Connected but failed to initialize"
The TCP connection opened but the analyzer did not answer *IDN? in time. Check that the port really is the SCPI port and that nothing else is holding the instrument.
Once the server sends commands the analyzer is in remote mode and locks its front panel (seen on a UTS3036B V1.04.0040 over raw TCP after :KEY:LOCK?, :KEY:LED? and :SENSe:FREQuency:CENTer?; *IDN? and :SYSTem:LOCK? did not trigger it). Press Esc on the analyzer to return to local mode. :SYSTem:LOCK 0, VXI-11 device_local/device_unlock and :KEY:ESC do not work on this firmware, and the :SYSTem:LOCK?, :KEY:LOCK? and :KEY:LED? readbacks are not reliable lock indicators.
A query timed out and the next call is slow An unsupported query gets no reply and makes the analyzer unresponsive for about 2 s, so the server waits 2.5 s after any timed-out query before sending anything else. This is expected; avoid sending queries the instrument does not support (see the verified list in Instrument Behavior).
Everything times out after changing the sweep time
While a single sweep runs, and with a very long manual sweep time even in continuous mode, the instrument ignores most commands. Wait for the sweep to finish and set the sweep time back to automatic: set_sweep with sweep_time_auto: true.
single_sweep / wait_for_sweep_complete timed out
The default timeout is the sweep time plus 10 s. A sweep waiting for a video trigger can take longer; either set trigger_source: "immediate" or pass a larger timeout_s. After a timeout the instrument recovers once the sweep ends.
"adjusted by the instrument: requested X, now Y"
Not an error. The analyzer clamps out-of-range values silently, snaps RBW/VBW to 1-3-10 steps, rounds attenuation to an integer and couples some settings. The warning tells you what it actually took. Re-read with the matching get_* tool.
The reference level changed after switching the power unit
The reference level reads back in the unit currently selected, and includes the reference level offset. set_amplitude applies the unit first, then the level, in that unit.
A manual RBW/VBW/sweep time "disappeared" Setting a manual value turns its auto flag off, but changing the span or RBW can make automatic sweep time apply again. Re-read after changing them.
| Error | Meaning |
|---|---|
-113 "Undefined header" |
The instrument does not know the command or query. Stale ones from an earlier unsupported query appear about 4 s later; the server ignores them for 10 s after a query timeout |
-131 "Invalid suffix" |
A unit suffix was not understood |
-224 "Illegal parameter value" |
An invalid enum value (for example a bad detector or scale name) |
Errors are returned with their code, text and the commands that were sent, and say when settings may be partially applied. Use the matching get_* tool to see what is currently set.
"... requires confirm=true"
scpi_command refuses reset, network, lock, configuration and file-write commands unless confirm: true is passed. See Safety and Security. Commands containing newlines, control characters, non-ASCII text or unbalanced quotes are always refused.
"scpi_query only accepts queries" / "one query per call"
scpi_query takes exactly one statement ending in ?. Use scpi_command for commands and make separate calls for separate queries.
-
Limit lines, tracking generator: the tools (
get_limit_line,set_limit_line,delete_limit_line,get_tracking_generator,set_tracking_generator) are available since v0.4.0 (see the Roadmap). -
Screenshot: the analyzer stops a screen image after 144,800 bytes over raw TCP but delivers all of it over VXI-11 (issue #42). The
screenshottool therefore needs VXI-11: over raw TCP it refuses with an error telling you to reconnect withtransport: "vxi11"(orauto).
Open an issue and include the analyzer model and firmware, how you ran the server (release tarball, git tag, Docker or a clone; Node version; OS; UTS_TRANSPORT), the tool call and the exact error text. Please do not post serial numbers or private network details. For security problems use private vulnerability reporting instead (see Safety and Security).
MIT licensed. Verified on a UNI-T UTS3036B (firmware V1.04.0040); other models are untested. This wiki is generated from the repository's docs/ folder, which is authoritative: change it there, not here.