On May 5 (AM), both CTDs on B2 stopped reporting. Suspect the Inductive Couple Cable (ICC) or fuse to the SBE Inductive Modem within the canister. Code that cycles power to the modem has been activated and does not remedy the CTD blackout. CTDs are set to record internally so no data is being lost.
Things to test on next visit.
-
test for continuity on ICC plug to canister (pins 2 and 3) -- will indicate health of the ICC cable below
-
connect to CR1000 (locally) via RF modem or Diagnostic cable and use talk-thru to determine health of the SBE Inductive Modem
-
Depending on above, findings (a) replace fuse, (b) replace SBE inductive modem, (c) place another ICC cable on the buoy which would require divers and very calm day