Skip to content

Conversation

@xesxen
Copy link
Contributor

@xesxen xesxen commented Nov 5, 2025

The Agilent DSO-X 2022A with MSO extension features 8 digital channels rather than 16. Querying for >8 results in Query UNTERMINATED notices on the oscilloscope and freezes the application for ~26 seconds on startup. Leaving the channels enabled breaks the normal acquisition mode as it will not retrigger after one cycle.

The second commit addresses a weird situation where my scope replies with +0 in response to :WAV:PRE?. This only happens on the 1st channel once the scope has finished acquiring data, the 2nd channel didn't present the same issue. Retrying the same command once if the string is unexpectedly small fixes this behavior.

@azonenberg azonenberg added this to the v0.1.1 milestone Dec 9, 2025
@azonenberg azonenberg merged commit 1d958f9 into ngscopeclient:master Dec 9, 2025
Sign up for free to join this conversation on GitHub. Already have an account? Sign in to comment

Labels

None yet

Projects

None yet

Development

Successfully merging this pull request may close these issues.

2 participants