Smaller estimated resolution (nm) indicates finer resolvable detail, conditional on valid sampling, sufficient correlation signal, and the chosen threshold. Do not read a numerical resolution value as a guarantee of image correctness or as the physical PSF FWHM.
| Result | Practical meaning |
|---|---|
| Global sFRC resolution | Effective XY resolution of the analyzed image or frame/pair |
| sFSC lateral resolution | Estimated XY frequency support expressed as a resolution |
| sFSC axial resolution | Estimated Z resolution inferred by DRI from directional FSC data |
E_DRI |
Relative fitting inconsistency of the directional-support model |
| srFRC local map | Spatially varying local resolution in nm |
NaN or unavailable |
No trustworthy value produced by that analysis branch |
srFRC map value 0
|
Masked background or unavailable local result, not measured 0 nm resolution |
The user-facing choices are 1/7 hard threshold and 3-sigma. sFRC and sFSC default to 1/7; srFRC defaults to 3-sigma. A correlation cutoff is separate from the background intensity threshold applied to local maps. Changing the criterion can change the reported resolution; document the chosen threshold when making comparisons.
For global sFRC/sFSC, curve extrapolation is conditional and limited to the fixed 1/7 criterion; it should be understood as a model-based estimate rather than an observed Fourier-frequency crossing. In the DRI directional stage, extrapolation is attempted only when none of the measured directions has a usable crossing. Local srFRC windows do not undergo that global-curve extrapolation workflow; calibrated internal fallback thresholds may be used where implemented.
E_DRI is a cosine-weighted relative RMS fitting error between measured directional cutoffs and the DRI model. Some APIs store a fraction, while result summaries display percent.
E_DRI |
Interpretation in the current software guidance |
|---|---|
| Below 10% | Good fit consistency under the DRI criterion |
| 10–20% | Interpret with caution; review angular cutoffs and fit |
| Above 20% | The axial estimate is not acceptable under the stated fit-quality criterion |
| Not available | DRI fitting or assessment failed |
A small fitting error does not independently validate an axial result; a smooth but systematically distorted directional profile can still bias DRI. Compare fit plots, Z sampling, and the acquisition context where necessary.
- The original physical pixel/voxel spacing must match the acquired data.
- XY diagonal subsampling changes the effective analysis-grid spacing; enter the original acquisition spacing, not the internally subsampled one.
- For sFSC, check the order of the voxel-spacing vector in the corresponding language.
- For srFRC, foreground mean/min/max may be calculated on the native map before upsampling and may not equal statistics recomputed from an interpolated exported map.
- Quantitative Float32 maps preserve resolution values; coloured RGB previews are display products. When each frame uses its own colour scale, identical colours do not imply identical nm values across frames.
- Avoid treating an absent crossing,
NaN, or a masked-zero pixel as extreme image resolution.
State the image modality, frame/volume pairing policy, physical spacing and units, 1/7 versus 3-sigma threshold, any correction or extrapolation, and (for sFSC) angular interval, cone count, Z sampling and E_DRI. For srFRC record window width, step, background-selection mode, analysis frame(s), and whether statistics came from native or interpolated maps.