sFSC and DRI — anisotropic 3D resolution

With a volume open in Fiji, select Plugins → srFRC → sFSC. This tool provides a lateral resolution and a DRI-inferred axial resolution using directional FSC measurements.

Input volumes

With If SingleFrame=true, the active 3D volume supplies a complementary pair generated internally. With false, select the two matching opened volumes via 3D Stack 1 (paired mode) and 3D Stack 2 (paired mode); follow the plugin's channel-selection guidance if applicable.

All voxel-spacing input is in nm. Pixel Size (XY) and Pixel Size (Z) may be initialized from Image → Properties but are editable; the values entered in the dialog are used.

Main and advanced settings

Label Default Description
If SingleFrame true Single-volume pair or two acquired volumes
Pixel Size (XY) Image properties or user input XY spacing, nm
Pixel Size (Z) Image properties or user input Z spacing, nm
Threshold 1/7 hard threshold Alternative 3-sigma
Z Sampling false Manually sample odd/even z-planes
Low Angle 15 Degrees from XY plane; recommended >10°
High Angle 60 Degrees from XY plane; recommended <80°
Number of Cones 5 At least 5 recommended
Plot Fitting Curve false Optional extra DRI fitting figure
xyMode 'slice' '3D' optional for low-SNR input

The last settings are accessible under Advanced Settings. Angles are degrees in this implementation.

Reading the output

The results table shows lateral and axial resolutions in nm and E_DRI in percent alongside settings. The optional DRI fit figure displays the model's agreement with sampled directions. The full grading guidance should be interpreted as fit consistency, not a ground-truth error bar.

Diagnosing an overly small axial resolution

First check Z pixel spacing and volume structure; then review the non-paraxial cutoff curves. Compare Z Sampling OFF and ON manually when appropriate. Changing fitting angles should be recorded in the final report. Directional extrapolation under fixed 1/7 is considered only when no sampled direction offers a measured crossing.