"Select eyes to detect" (Both / Left only / Right only) is a genuine,
persistent device-level property — the same platmod-property family as the
display area — not merely a calibration input. Source of truth:
crates/tobii-protocol/src/frame.rs (op codes),
crates/tobii-protocol/src/commands.rs (EnabledEye, payload/parse),
crates/tobii-usb/src/connection.rs (get/set_enabled_eye), memory note
et5-enabled-eye-op.
| Op | Dec | Name | Request payload | Response |
|---|---|---|---|---|
0xc62 |
3170 | get_enabled_eye | empty | 00 00 02 00 00 00 04 00 00 00 0N |
0xc58 |
3160 | set_enabled_eye |
00 00 02 00 00 00 04 + BE32(value) |
ack |
SET (
0xc58) is a LOWER op code than GET (0xc62) — the inverse of the display-area and calibration convention. This is real; do not "correct" it.
Msg-class/tag 0x51 like display-area, 3 s timeout. Both are version-gated
on [dev+0x1a0] >= 0x10007; older firmware returns NOT_SUPPORTED (err 2), and
get_enabled_eye then returns None. [CODE-VERIFIED] op codes / gating
(native disasm); [CONFIRMED] live that both ACK and round-trip.
1 = LEFT 2 = RIGHT 3 = BOTH (0 folds to BOTH)
This is the C-API tobii_enabled_eye_t {LEFT=0, RIGHT=1, BOTH=2} plus 1 —
the response integer is NOT the C-API value; remap with C-API = wire − 1.
EnabledEye::{Left,Right,Both} encodes exactly 1/2/3. [CONFIRMED] —
commands.rs::to_wire/from_wire, enabled_eye_wire_round_trips.
The hardware-captured BOTH GET response is:
00 00 02 00 00 00 04 00 00 00 03
^^^^^^^^^^ BE32 value = 3 (BOTH)
parse_enabled_eye reads the trailing BE32. [CONFIRMED] —
commands.rs::enabled_eye_payload_and_parse_match_device,
connection.rs::get_enabled_eye_parses_device_response.
Enum landmine: this property's wire enum (
1=L,2=R,3=B) is a different numbering from thecal_add_point(0x408) eye argument (0=both,1=L,2=R, see Calibration). Same integer ≠ same eye across contexts. Never share a constant between the two.
Hardware-verified 2026-07-20 [CONFIRMED] (memory et5-enabled-eye-op):
- A standalone SET is acked and PERSISTS across reboot — a fresh-session GET returns the value that was set.
-
BUT a standalone SET does NOT change live detection. Both eyes keep
reporting
validity == 0in the gaze stream regardless of the setting. - The property only takes effect when the tracking model is rebuilt, i.e. via
tobii_calibration_start(enabled_eye), which happens on (re)calibration. The vendor app never SETs it standalone; it persists the choice host-side per profile and re-applies by recalibrating.
Consequence: setting enabled_eye alone stores a preference but the tracker
keeps detecting both eyes until a calibration is run. Whether a standard
calibration (which drops the eye arg on the wire, Calibration) is enough to
apply the selection is unresolved — a Windows-VM capture of toggling
Left/Right/Both settles it. [HYPOTHESIS].
Connection::get_enabled_eye() -> Option<EnabledEye>,
set_enabled_eye(eye) -> bool (acked?). CLI:
tobii enabled-eye [both|left|right] — sets (if an arg is given) then reads
back. [CONFIRMED] — connection.rs, main.rs::enabled_eye_cmd.