馃悰 Fixed

  • A bench that was already connected could not be reconnected. After a
    server update the scope and both supplies refused to connect while the
    server was still holding all three open and serving readings from them.
    connect always opened the resource again, so pressing Connect asked
    libusb to claim an interface the server itself was holding; it answered
    [Errno 16] Resource busy and the API returned 500 -- a refusal that reads
    like broken hardware and is really the server colliding with itself. An
    instrument that is already open is returned as it stands, and an entry that
    is present but no longer connected is dropped first, since there a real
    open is wanted.
  • A websocket stream retried a command the instrument does not have. A
    scope asked for get_readings logged "Unknown command" twice a second,
    indefinitely. Every pass sends the same command, so what fails once fails
    always; the stream stops now.
  • A timestamp in a broadcast disconnected the client. send_json uses a
    plain json.dumps, which cannot encode the datetime in every readings
    payload, and the failure was raised inside the per-connection loop -- so a
    serialisation bug was read as the client hanging up and the websocket was
    dropped. The message is encoded once, before any connection is touched.