Summary
This release updates the polarization workflow around PPL/XPL, adds a reusable polarizer calibration baseline, and sets validated default acquisition exposures for routine PPL/XPL imaging.
Added
- Polarization mode naming is now standardized to
PPL/XPLacross the app, sequence flow, and documentation. - Polarizer-angle calibration now performs a coarse scan plus a fine scan around the darkest XPL angle.
- Calibrated polarizer baseline values are saved and restored on later app launches.
- Reachable orthogonal-angle handling for PPL derivation from the calibrated XPL angle.
scripts/analyze_xpl_exposure_sweep.pyfor reviewing RAW16 XPL exposure-sweep datasets.- Release packaging now stages files before zipping to avoid transient file-lock failures during archive creation.
Changed
- Default validated acquisition baseline:
PPL:18000 usat polarizer5 degXPL:400000 usat polarizer95 deg
- Default polarizer calibration exposure:
200000 us
PPLis no longer chosen from a brightness search; it is derived from the reachable orthogonal angle relative toXPL.- Sequence output folders now use
pplandxpl. - Distribution and implementation docs now match the current PPL/XPL baseline and calibration flow.
Sequence Behavior
- Live camera preview is paused before sequence capture and restored afterward.
- Sequence capture remains
XI_RAW16only. - Sequence images are saved as Bayer RAW
uint16 TIFF. - Sequence outputs include:
{SampleID}_log.csv{SampleID}_metadata.json
- CSV logs now include calibration signal metrics used during XPL-angle selection.
Polarizer Calibration
- The calibration scan searches for the darkest
XPLangle first. - A coarse scan identifies the minimum region, then a fine scan refines the result around that angle.
PPLis assigned asXPL + 90 degorXPL - 90 deg, whichever stays inside the reachable polarizer range.- The resulting baseline is saved for reuse on later runs.
Conversion Tool
- Converts saved RAW16 TIFF datasets into RGB preview images.
- Processes the selected folder and its immediate child folders.
- Saves results into a sibling
_rgbfolder. - Uses demosaic plus fixed white balance plus gamma for preview generation.
Validation
python -m py_compile apis\\config.py apis\\io.py apis\\sequence.py apis\\utils.py app\\main.py app\\workers.py scripts\\analyze_xpl_exposure_sweep.py tests\\test_ximea_raw16_sequence.pypython -m unittest tests.test_ximea_raw16_sequence tests.test_mock_serial.\build\build.ps1 -Version 0.1.3
Notes
- Sequence capture is still
RAW16-only in this release. - Live view and Snapshot behavior are unchanged.
- Reusing the same