Skip to content

Remote Control

Julian Decker edited this page Nov 13, 2026 · 5 revisions

Remote control

OOS has no authentication. Everything here is open to anyone who can reach the scope. Keep it on a trusted bench network and never connect it directly to the internet.

Interface Port Purpose
Web remote 8080 live screen, keys and touch in the browser, saves screenshots and CSV
SCPI over TCP 5555 Rigol-style command subset
VXI-11 111, 9009 VISA TCPIP::<ip>::INSTR
LXI identification 8080 /lxi/identification
MCP 8080 POST /mcp, tools scpi and screenshot
Telnet 23 root shell, no password

Web remote

Open http://<scope>:8080/. Click on the picture to touch the screen, drag to swipe. It works on a phone too.

From 0.1.6 the web remote, SCPI and VXI-11 no longer slow the scope down: on a DHO924S its screen stays at 30 frames a second with two browsers open (0.1.5: 16 with one), and the browser gets about 13 new pictures a second over Ethernet instead of 4 to 5. A script that keeps an SCPI connection open costs no frames either.

From 0.1.7 System > Network > Remote services switches the web remote, SCPI, VXI-11 and telnet on or off, each on its own. From 0.1.8 the live picture can be a JPEG made by the scope's video encoder (the button JPEG next to PNG): about 75 KB a picture against 115 KB, at a third of the processor time.

Ethernet

From 0.1.12 the scope has a name: http://dho924s-204527.local:8080 (model and serial number) reaches it on any computer that speaks mDNS (macOS, Linux with Avahi, Windows 10 and later), also when it only has a link-local address on a cable straight to the computer. Computers that browse find the web remote, SCPI and VXI-11 (DNS-SD). System > Network > Remote services switches mDNS off and changes the Name, which is also the host name the scope sends to DHCP, so a router lists it. When another device has the name already, the scope takes <name>-2.

The scope takes an address by DHCP. Without a DHCP answer it takes a link-local address, 169.254.x.y (from 0.1.9; before, 192.168.177.64), and without a cable none at all. System > Network shows the speed of the link. From 0.1.6, a link that comes up at 10 Mbit/s gets one more negotiation by itself (10 Mbit retry, on by default). If your scope still ends at 10 Mbit/s after a cold start, please report it with /tmp/oos-ui.log.

WLAN

Put an RTL8188EU stick into the front USB port, then System > Network > Find WLAN networks, tap a network and enter the password on the on-screen keyboard (First Steps has how to correct inside it). Everything open on Ethernet is then open on that network too. On the DHO924S of the author the scope has joined a home network with a password and worked over it for days.

SCPI

The commands are listed in ui/src/scpi.rs. From 0.1.9 also the counter: :COUNter:ENABle|SOURce|MODE|NDIGits, :COUNter:CURRent?, :COUNter:TOTalize:CLEar.