openSciLab 0.1.0b1

Pre-release
Pre-release

Choose a tag to compare

@github-actions github-actions released this 06 Oct 17:20
· 49 commits to main since this release

The first release of openSciLab, the open measurement and control lab that PiPiLogicAnalyzer
has become - a beta: every function works with the simulators and is covered by about 2000 tests,
but much of the new firmware has not been checked on real hardware yet.

Highlights

  • A lab instead of one instrument: several instruments at once, each with a device card -
    Pico boards (firmware protocol 8 with GPIO, PWM, ADC, pattern generator and UART/SPI/I²C),
    Arduino boards with a firmware of their own, DSLogic analyzers, the Rigol DHO900 oscilloscope
    (with a bridge app), measuring devices on other computers (openscilab_device) and simulators
    of all of them.
  • Flows: a visual editor (and a Python DSL) for *.flow.yaml - devices, GPIO, generators,
    every sigrok decoder, measurements, signal processing, control, data, views and reports; real
    time or deterministic virtual time; every node type has an example.
  • Panels as the front panel of a measuring station, reports in HTML or PDF with checks,
    and openscilab run to use a flow as a test.
  • Time: the samples of different instruments on one time axis (sync signals, latency
    measurement, drift, NTP/PTP).
  • The logic analyzer of PiPiLogicAnalyzer as the data view, with streams, analog and state
    captures, trigger sequences and large captures arriving progressively.
  • A new window: start page with every example project as a starting point, projects,
    templates, sidebar, command search, settings.
  • The repository is now deckerjulian/openSciLab, with a new wiki as the user guide.

The details follow.

Application

  • openscilab run <name> in a project folder runs the project's flow of that name, as the
    command line documentation says; openscilab --debug-driver also logs the drivers that run in a
    device process of their own.

  • Only the simulated boards with the openSciLab Arduino firmware are offered with the Arduino
    protocol (the simulated DAQ no longer shows up among them); messages and help texts name the
    current menus.

  • Every node checked, every node in an example. A review of all node types and simulators
    found and fixed these; a test now fails when a node type has no example and when an example's
    flow has any warning (a misspelt parameter was silently ignored before):

    • Streams: convert.edges and dsp.debounce no longer lose edges and short pulses where two
      blocks meet; dsp.resample keeps its sample grid across blocks; dsp.math combines the
      signals at their common times (blocks of a stream once, not block k with block k-1);
      measure.count keeps each wire apart. FIR filters and moving averages are placed without
      their delay; empty blocks no longer give two samples; impossible cutoffs are refused.
    • Numbers: measurements no longer round the sample rate to whole hertz; measure.setup_hold
      pairs data and clock by time; dsp.integral starts at 0 (trapezoids), dsp.derivative is the
      same streamed and whole; the FFT no longer doubles the Nyquist line; min(a) and max(a) work
      in dsp.math, which keeps the unit and makes a comparison of signals digital.
    • Units: control.compare, control.limit, dsp.threshold and control.sweep take
      quantities in the unit of the signal (20 mA against a reading in mA, 3.3 V, a tolerance of
      50 mV); a missing value to compare with is an error.
    • State machine: values that came before a state was entered no longer lead out of it, and
      values for other states no longer hold up their sender; a way out without after or input
      is refused. A wired start of a sequence runs it for every value; waiting for an input
      without a wire is refused.
    • Devices: unknown pins are named (uno has no pin 'd7' - D7?); a refused capture says the
      device's limits (the rate is at most 100 kHz); pre must be fewer than samples; levels
      may be high/low, a duty cycle 50 %, an I²C address 0x48; remote.set reads false
      as false; channels named CH1 fit digital and analog inputs.
    • Generators: gen.arbitrary plays a CSV file or a signal at their own rate; gen.tx_*
      send the data parameter for an event without data and say done when the transmission is
      through; gen.output plays an .sdl file on a pin and rate of your choice; a square-only
      output gets a square over its whole range by default.
    • Decoders: text takes the first data row that has annotations (only MOSI wired: its data)
      and of I²C only the data bytes.
    • Data, views, reports: bundles are written as numbers, not Python text; data.file writes
      signals as captures, bundles as CSV, refuses a table as a capture file, and its numbered files
      go on after those already there (data.file_write, a duplicate of it, is gone);
      data.logger writes only the new rows of a sliding table, one line per column, and keeps
      analog samples too; data.buffer keeps when its values came. view.scope joins only blocks
      that follow each other; view.strip_chart draws every channel of a capture and events at
      their times; view.number and view.led show signals and texts like on; diagrams of a
      report skip values that are no numbers; structure.unbundle finds numbered fields.
    • Flows: an unknown parameter is named with a suggestion (has no parameter 'level' - did you mean 'threshold'?); errors no longer repeat the node's name; a conversion inserted on a wire
      gets its parameters (a number to a truth value: at or above 0.5).
  • Captures and streams in virtual time see what the flow does meanwhile: the flow's time runs
    on while a simulator's capture waits for its trigger, so a stimulus after armed - also after a
    wait - is captured, and a stream ends where its stop came. Two examples that had to run in real
    time no longer do; captures of USB devices (Pico, DAQ) have the flow's time in virtual time too.

  • State mode in flows: device.capture with a clock channel takes one sample per clock edge
    and sends them at states (convert.state_bit takes a channel of them); a pattern trigger may
    have don't-care bits (1x11).

  • The simulators behave like the devices: the Uno's captures hold 448 samples up to 100 kHz
    (pins only, at most 2 s after the trigger) and it streams 11.5 kB/s, as its firmware; the Uno R4,
    the Pico (800 kB/s) and the DAQ (20 kS/s for all inputs) likewise; the oscilloscope and a
    multi-board set no longer offer a stream. A generator that stops keeps its time in the signal
    (a capture across the stop sees both parts); Safe, a restart and the watchdog stop generators
    and analog outputs too; pull resistors no longer stick; an RC low pass follows a generator or a
    DAC; sources swing to the board's logic level (5 V on an Uno); the Arduino board stays free
    after a stopped capture; the state mode waits for its trigger; an edge right at the start of a
    capture counts. The simulated DAQ on the network measures a 50 Hz sine on AI0 as its demo does.

  • New examples: State mode, Signal processing (derivative, integral, resampling, averaging,
    logic to volts), Send SPI, Waveform file, Replay a capture; Bundle takes its bundle apart
    again.

  • Fixed: questions and messages had a large empty space left of their text (the icon took the
    width meant for the text); the question when closing names a temporary project properly.

  • Edit a node in a window of its own: a double-click on a node opens its parameters in a
    window (a double-click on the data shown in it still opens them in a view). Mappings and long
    lists - the states of a state machine, the signals of a simulator - get a big text field in YAML,
    one entry a line, in the inspector too.

  • Fixed: YAML read on: and off as true and false (YAML 1.1), so the transitions of a state
    machine were lost and the Traffic light example stopped in its first state. Flows, panels and
    projects are read as YAML 1.2 now (only true and false are truth values); a state machine
    names keys it does not know and states it cannot go to.

  • A tidier window. The command search sits right after the logo; Examples is called
    Templates (menu and header). Tabs have their close button on the left. The node palette is no
    sidebar section any more: it shows beside a flow whose graph is edited (View → Node palette,
    Ctrl+Shift+N). The activity bar has the start page and the settings at the bottom and is
    arranged with a right click (show, hide, order, names under the icons); the parts of the
    sidebar sections (open documents, project, data, recent files, connected and available devices)
    open and close with a click on their header and show how many entries they hold. The overview
    of a flow draws its nodes as blocks in the colours of their kinds, with the visible part outlined.

  • The rows of a data view start at the top with their height instead of stretching to fill the
    view; the overview below the waveform shows analog channels too (as an envelope) and fast
    digital signals clearly.

  • Nodes of a flow look the part: a title bar in the colour of their kind (devices, measurements,
    views, data, timing, ...) with an icon, a soft shadow, port names brighter when wired, parameters
    as a small table. Wires back (feedback, e.g. to the next step of a sweep) run round below the
    nodes instead of through them.

  • Better arranging, and Auto-arrange: Arrange (now in the tool bar of a flow) lines wires up
    port to port, routes long wires past the columns between and orders the nodes of a column with
    as few crossings as it finds; Auto arranges the flow again whenever nodes or wires are added or
    removed. Every example and template is arranged anew (none overlaps).

  • Fixed: the capture settings opened from the tool bar of a data view showed on a white background.

  • Every example is a starting point; the start page shows them all. The start page lists every
    project of the example library as a tile, by category, scrollable and with a search; a click opens
    it as a temporary project. The first category, Start a project, holds the plain starting points:
    Empty lab, Logic Analyzer (now an ordinary project that opens a data view with the connected
    device: open: [data]), Data acquisition (analog inputs of a measuring box recorded until Stop,
    strip charts, a file, an output set from the panel), Synchronized instruments (a logic analyzer
    and a DAQ with a drifting clock aligned by a sync signal) and Remote measuring device (a device on
    another computer with the script to run there). Characteristic curve, Test bench and
    Long-term logger are examples of Measurement and Data and reports. Project → New project…
    shows the start page; there are no separate templates any more.

  • Streams of analog inputs in flows: device.stream hands on analog channels as blocks of
    volts (channels: [AI0, AI1]), also without digital channels; analog-only streams of the
    simulators no longer fail. Inputs named AI0 or AIN0 are analog ports.

  • A local DAQ simulator (sim:daq, Simulation: USB DAQ): a measuring box on USB in the class
    of an NI USB-6001 - eight analog inputs, two analog outputs, twelve digital lines - whose blocks
    arrive as over USB and whose sample clock runs 30 ppm fast, with a digital loopback (P0.0 → P0.1)
    for the latency, AO0 wired to AI2 and P0.7 free for a sync signal. Simulator profiles take a
    clock_drift (ppm); Signals → USB link and clock changes it.

  • Smaller text and tighter controls by default: 10 px text; padding, icons and the rows of the
    waveform follow the text size (Settings → Appearance; 12 px gives the earlier spacing).

  • Simulated remote devices with the timings to try. The device list has the remote DAQ with its
    own drifting clock, with a clock that follows UTC (as with PTP) and with a shared sample clock;
    the Simulation: timing box on its Remote tab changes the network's latency and jitter and the
    drift of its clock while it runs, and starts it again with another clock. The address keeps the
    settings (remote-sim:daq?latency=20&jitter=2&drift=300).

  • A higher priority for reading devices on macOS and Linux (Settings → Devices, off by
    default): openSciLab asks when it starts and raises itself and its device processes with the
    password dialog of the system; on Linux it can be allowed for good. Windows needs no rights.

  • Fixed: a black box in the top left corner of some device cards (an empty event list).

  • A Timing tab on the device card. How the samples of the device are placed and how well (now
    also for captures of the data view, not only of flows), its sample clock - own, or shared /
    external, which timing.align (new default drift: auto) follows -, Measure the latency
    with a loopback right there (an output wired to a channel, rate, stream or capture; kept like
    that of timing.calibrate), the measured latencies, a sync output on a pin (the signal of
    timing.sync until stopped) and this computer's NTP/PTP clock.

  • Wires and the USB link of a simulator on its Signals tab: Add wire… connects a pin to a
    channel (Wire them on the Timing tab adds the loopback at once), USB link emulates latency,
    jitter and frames (or not); both are kept per simulator. In real time a simulated capture
    records what the outputs do until it ends, as the device does.

  • The device card has everything about a device in one place. The extra Device information
    window is gone: the Details tab shows the device's functions (capture settings, self-test,
    network, bootloader, firmware, Reconnect, Copy the details), the capture limits, the
    capabilities, board, firmware and connection. New there: Restart device for devices that report
    RESTART (the simulators; the Pico firmware has no restart command yet - Reconnect restarts an
    Arduino, Restart into the bootloader a Pico). Profiles are in the card's header (load,
    save, standard profiles), which also names the settings of the next capture.

  • Disconnect all devices at once (Devices menu, the button above the connected devices):
    one question names every device and what disconnecting it interrupts.

  • A fresh start when nothing was saved. The session brings back saved files and the project;
    device cards, data views of devices and the connected hardware no longer come back, and devices
    are only connected again when Settings → Startup says so (off by default).

  • Views of a flow's data do not offer to capture. The data view of Run data and of a scope
    node shows search, measure, panels and save - no capture bar, no device, no Open.

  • The simulators are one group of the device list again: also the simulated Arduinos and
    oscilloscopes that speak the real protocols. They are no longer listed as connected hardware.

  • Connected hardware says how firmware gets onto a board: three steps above the list (plug in,
    with BOOTSEL for a board without firmware; find it; Install firmware in its row).

  • Tabs look alike everywhere (console, device card, dialogs, analysis panels): flat, with an
    accent line under the current one; the white line under the tabs is gone.

  • Plugins add devices without changing openSciLab. A .py file or package in the plugins
    folder of the settings directory (or a folder of OPENSCILAB_PLUGINS, or an installed package
    with an openscilab.plugins entry point) registers a kind of device with its driver:
    kinds.register("mydevice", open_mydevice, detect=..., process=True). The device list then shows
    its devices, flows open it with address: mydevice:..., and so do openscilab-cli and
    api.open(); with process=True it is read in a device process of its own, Simulate uses the
    simulator profile it names. A plugin with user interface of its own (a backend that asks for an
    address) brings it in setup_ui(). A plugin that fails is reported (status bar, Help →
    Plugins…
    , openscilab-cli plugins), the others load. examples/plugins/counter_device.py is a
    complete example; see docs/drivers.md. An address of an unknown kind (foo:...) now says so
    instead of being tried as a Pico.

  • Tool bars that fit the window and can be arranged. The header, flow, panel and data view bars
    (including the view bar above the tracks) give way when the window is narrow: the less important
    entries first show only their icon, then move into a More menu (⋯) at the end of the bar,
    where they work as in the bar (check boxes, device and other choices, settings menus). Capture and
    Run always stay. A right click on a bar opens Customize toolbar… (show, hide and order entries,
    some extra ones such as the align tools of a flow), the display (Automatic, Icons only, Icons
    and text
    ) and Reset toolbar; the arrangement is kept per bar. The data view's two bars became
    one, its view bar is a tool bar as well, so a data view fits narrower windows. Buttons have a flat
    tool look with clear hover and checked states.

  • Devices on USB are read in a process of their own, so nothing the application does - drawing,
    flows, scripts, its garbage collection - can make a stream overflow (the Pico firmware's ring lasts
    22 ms at 6 MB/s). A Pico, Arduino or DSLogic runs in a device process: its driver and facets live
    there, its samples come through shared memory, progress and completion come back in order (a busy
    application lags behind, the reading never waits). The device process stamps the start command and
    every block, which places the samples more closely. A device process that dies fails its capture
    with a message and disconnects the instrument; one whose application went away stops the device.
    On by default (Settings → Devices); simulators, remote devices and the Rigol stay in the
    application, the Python API opens in a device process with process=True. Also: drivers hand
    their events to a notifier thread of their own, the startup heap is frozen for the garbage
    collector, serial ports get a large receive buffer (Windows). The simulated Pico emulates the
    firmware's ring: tools/benchmark.py stream shows its stream overflowing under load in the
    application and not in a device process.

  • Decoding never stalls recording and synchronisation. The protocol decoders run in a process
    of their own (started once, the samples through shared memory, the annotations back in small
    parts); a long decode in the application used to make the threads that receive a stream or a
    remote device wait up to about 100 ms, now as long as when nothing else runs. Cancelling ends it
    at once; a decoder process that died is started again. Annotation segments are tuples the
    garbage collector stops looking at (no long pauses after a large decode). The sync path is
    vectorized: pairing edges (15x), finding the sample of a time (500x), analog sync inputs (5x,
    and noise no longer makes edges before the signal shows its levels). tools/benchmark.py
    measures all of it.

  • The time of samples, for every device (docs/timing.md). Captures and
    streams of openSciLab's own instruments are placed between the command that started them and the
    arrival of their data (the lower envelope of the arrivals removes the jitter of USB) instead of at
    the moment the flow asked; the device card shows the method and the accuracy (Timing). Better:

    • a sync signal from openSciLab: timing.sync drives a pin with edges at irregular intervals;
      timing.align places an instrument on a reference's time by it (offset and drift, or only the
      offset for a shared sample clock), remote.sync a remote device - also one that records the
      signal on an input of its own (a spare channel of a DAQ), with the edges placed again with all
      its arrivals. Accuracy: a sample period of the slower recording.
    • a measured latency: timing.calibrate switches an output wired back to an input and keeps
      the latency per instrument and rate; later captures and streams use it without the wire.
      Device scripts do the same (input.calibrate(), latency=) and stamp their blocks from when
      they arrive (input.start()).
    • a time scale (PTP, GPS, NTP): a device whose clock follows UTC or TAI is converted without
      measuring when this computer's clock follows it too (Settings → Time); drivers of local
      instruments that stamp their samples report it the same way.
    • The simulated Pico emulates USB (frames, latency, jitter), so all of it works without hardware;
      remote-sim:daq is a DAQ behind USB on another computer (?shared_clock=1, ?timescale=utc);
      examples/remote/ni_daq.py is a template for a real NI DAQ; four examples in the category
      Time.
  • Standard profiles for the common buses, loaded without importing: every profile file in the
    new profiles folder (profiles/ in the settings directory, Profiles → Open the profiles
    folder
    ) is there at once. openSciLab puts its standard profiles into it – UART, MIDI, DMX512,
    LIN, I²C, SPI, 1-Wire, CAN, PWM, WS2812, I²S, parallel, SWD, JTAG, IR (NEC), USB and the C64
    expansion port (moved from examples/c64-expansion-port-profile.json to
    examples/profiles/c64-expansion-port.json) – each with its channels named, a trigger, the
    decoders and the wiring in its notes. Profiles → Standard profiles loads one with a click; a
    profile needing more channels than the device has is off (with the reason), and rate and length
    are fitted to the device when a profile is loaded. Profiles from the folder are edited and
    deleted in their own file; a deleted standard profile stays deleted.

  • Flows and their panels find each other. The flow editor's Panel button has a menu of the
    panels that use the flow (open ones and the panel files of the project or the flow's folder) and
    New panel for this flow; the flow's inspector lists them too. A panel opens its flow with
    Open flow (tool bar and inspector).

  • The simulators of the device list are one group that opens and closes with a click (it
    remembers how it was left).

  • The example library is part of the packaged application (the Examples menu was empty there).

  • Capturing happens in the data view. Its capture tool bar chooses the device (connected
    instruments, or Connect a device…) and holds Capture (F5, also the header's button),
    Again, Stop, the quick settings, Settings… and Profiles; the Capture tab shows the
    next capture's settings and channels. The template Logic Analyzer asks which connected device
    to use. The device card no longer captures: Capture… opens its data view, and the functions of
    the device (information, self-test, test signals, network, bootloader, firmware) are buttons on
    its Details tab instead of a Device menu.

  • A tool bar instead of the device names. The top bar starts with the new logo of openSciLab
    (a lab flask holding a digital signal; also the application icon) and a button for each basic
    element – Flow, Panel, Data view, Waveform, Project (from a template) – then Open,
    Save and Examples; the connected devices are in the device list and the status bar.

  • The selection on the time axis shows over every track (digital, analog, pinned, buses), and
    Fit selection (view bar, the ruler's menu, Ctrl+E) shows exactly the selected samples.

  • Time axis and tracks line up again: with annotation rows (or many channels) the ticks of the
    time axis drifted away from the lines of the tracks by the width of the tracks' scroll bar.

  • Fixes of the window: menus that open from a full tool bar (or from its buttons) and the scroll
    buttons of the document tabs were transparent; the tabs drew stray white lines and cut names to
    a few letters; on macOS the menus of the application (Examples, Devices, …) were replaced by those
    of a data view once it had the focus.

  • Remote devices over IP. A Python script on another computer becomes a device: with the
    package openscilab_device (standard library only, Python 3.8+) it describes its inputs (numbers,
    truth values, sampled signals, events), outputs, commands and sync signals and connects to
    openSciLab (Settings → Remote devices, with a token; a beacon lets devices find openSciLab).
    The device appears in the device list with a Remote tab (latest values, outputs, commands, how
    well its clock is known) and in flows as remote:<name>: remote.receive, remote.set,
    remote.call, remote.sync; gpio.write, gpio.read and device.monitor work with it too.
    Time stays right by itself: values are stamped on the device; openSciLab measures offset and
    drift of its clock all the time (a fraction of a millisecond on a wired network), outputs are sent
    ahead with the time they should happen, values of several devices are handed on in the order of
    their time. A sync signal (an output of the device that a logic analyzer records too) aligns
    it to microseconds. Demo devices (python -m openscilab_device.demo climate), simulated remote
    devices (remote-sim:climate, audio, echo) behind an emulated network, four examples, a
    Raspberry Pi template and docs/remote.md (with the protocol).

  • Examples menu with a library of 63 examples, sorted by topic like the examples of the Arduino
    IDE: Basics, Digital I/O, Analog, Protocols, Measurement, Signal generation, Control, Data and
    reports, Panels, Python and Advanced. Each one is a small project with simulated devices (logic
    analyzer, Arduino Uno, Pico, oscilloscope with function generator) that runs without hardware
    and explains itself; it opens as a copy, so the library stays as it is. The examples are also in
    the command palette; examples/library/README.md lists them.

  • Fixes found with the examples: a counter missed the edges between the blocks of a stream
    (200 ms of 1 kHz counted 180); a buffer joins the captures of a stream (it kept only single
    channels); channels with a dot in their name (B2.GP5 of a multi device) can be wired; a
    simulated capture in virtual time sees outputs the flow sets after it was armed (a pulse train);
    values read better on wires and in logs (the value of a single event, bytes in hex, the fields of
    a bundle, numbers without a unit without an SI prefix).

  • The data of a run in the graph and in data views. While and after a flow runs, its nodes
    show their data below their parameters: a trace of the latest capture, the latest value with its
    course, the rows of a table; a view node shows what arrives at it. Double-click a node (or Show
    data
    in its menu, or the corner of its data) to open its data in a data view – a table opens as
    a table. Run data in the flow's toolbar shows all data of the run in one data view on one time
    axis: the channels of every capture, sampled signals, measurements as lines that step from value
    to value, checks as digital lines. Open data views follow a running flow and the next run.
    Show data in the nodes and Clear run data are in the Flow menu. Nodes keep room for their
    data, so an arranged flow stays in order when data arrives.

  • Flows are arranged along their wires. Nodes without a place are laid out from left to right
    in columns, with their real height (no more nodes on top of each other) and with feedback wires
    (a sweep waiting for its measurement) not turning the flow around. Arrange (Flow menu,
    Ctrl+Shift+L) does it for the selected nodes or the whole flow, as one undo step.

  • The inspector offers what the device has. Pins come as a list of the pins that can do what
    the node needs (PWM pins for a PWM node, outputs for Write pin, reserved pins left out),
    channels and analog inputs are ticked in a menu, generator outputs are chosen from the device's
    outputs. A decoder's channels are chosen one by one from the channels of the capture wired to
    it; a report image offers the columns of its table. The values come from the instruments open in
    the device list and from the simulator profiles.

  • Completions while typing YAML and Python. The YAML view of a flow offers the sections, node
    types after type:, the parameters of a node's type that are not set yet, the values a
    parameter can have (choices, the pins and channels of the device – a node typed just now uses the
    flow's only device) and the ports in the edges (outputs before ->, inputs after it). The code
    of a Python node is highlighted and completes ctx., np., signals., units., the names of
    the code and its port names in ctx.emit("…"), ctx.receive("…") and port == "…". The Python
    console completes the attributes of its names. The list opens by itself or with Ctrl+Space;
    Enter or Tab takes a completion.

  • Values that do not fit are found while editing. A word where a number belongs, a frequency
    given in seconds, a choice that does not exist, a pin the device does not have or that cannot do
    PWM, a channel the capture does not record: each is a warning at its node, in the inspector and
    in Problems (the flow still runs – a project may use another device).

  • A panel from the graph. Panel in the flow's toolbar (and the Flow menu) makes a front
    panel for the flow – also for the unsaved flow the application starts with, which the panel then
    uses as it is edited; saving the panel saves the flow first. A new panel starts with suggested
    widgets: a start button of a sweep, a slider for the duty cycle of a PWM, inputs the flow needs,
    numbers and LEDs for measurements and checks, scopes and charts for what the flow shows.

  • The template flows are laid out and the test bench passes. Every node of the templates has
    its place; the test bench captures long enough to find the UART text wherever the capture
    starts (it failed in real time) and writes its report when the flow ends, with both checks. The
    empty template shows its device as a node.

  • Pico boards with firmware 8 are full instruments. The device card, flows, scripts and the
    command line get their pins (the board's pin table, reserved pins with the reason), outputs,
    pulses and pulse trains timed by the board, PWM with the frequency the board makes, the monitor,
    analog inputs A0–A2 in captures and streams, the pattern generator, the square wave of GP22,
    UART/SPI/I²C sent by the board and the live state stream. A pulse can be triggered while a
    capture waits for its trigger on the same board. A loaded pattern and analog channels share the
    capture memory, so the capture settings offer fewer samples then.

  • Arduino boards (arduino:<port>) with the openSciLab Arduino firmware: Uno, Nano, Mega 2560,
    Uno R4 Minima, ESP32 and ESP32-S3. They are found by their USB identifiers and offer stream,
    buffer and live state captures with analog channels, GPIO, the monitor, the pattern generator,
    square and arbitrary waveforms (DAC on the R4 and the ESP32) and UART/SPI/I²C. A simulated board
    speaks the same protocol (arduino-sim:uno, Simulation: Arduino Uno (firmware protocol)).

  • Rigol DHO900 oscilloscopes (rigol:<address>): D0–D15 and CH1–CH4 in one capture with edge,
    pattern or no trigger and adjustable logic thresholds, and the built-in generator from the
    waveform document (standard waveforms, arbitrary points, sweep, burst). With the openSciLab
    bridge app on the instrument a large capture is shown at once and its samples follow tile by
    tile, the visible range first; an interrupted transfer resumes; the Cache tab of the device card
    lists the captures the app keeps. Oscilloscopes with the app appear in the device list by
    themselves. Without the app the driver reads the instrument directly, more slowly.
    rigol-sim:bridge and rigol-sim:direct simulate one. The SCPI commands are provisional until
    checked on an instrument.

  • tools/dho_la_probe.py records what a DHO900 answers to the commands the driver needs.

  • Pulses take a count and a period (device card, flow node gpio.pulse, simulators).

  • Outputs stay as they were set. The driver keeps a device's watchdog satisfied while it is
    connected – not the device card: an output a flow or a script sets no longer falls back to an
    input after a second (it did on the simulated Arduino). Disconnecting asks while any pin is an
    output, whoever set it.

  • Sending UART, SPI and I²C with the device's own hardware. Devices that can (capabilities
    TX_UART, TX_SPI, TX_I2C) send through the generator facet and give back what a target
    answers: the bytes on MISO, or the bytes an I²C device returns after a write. The nodes
    Send UART/SPI/I²C use it (new output received, I²C parameter read, SPI parameter miso)
    and play a pattern on devices without it; the device card has a new tab Send. The simulated
    Pico, Uno and Uno R4 can, with a temperature sensor at I²C address 0x48; the Uno got a pattern
    output (D2–D7) and a square wave output (D9).

  • Devices get their facets (GPIO, monitor, analog inputs, generator) from what their driver
    offers, the same way in the device list, in flows, on the command line and in scripts.

  • A second check of the review. Saving a capture file keeps the form it has (a file of
    the original software was silently packed by Save). The dialog shown while a file is
    written can no longer be closed (closing it ended in an error and left the document
    "unsaved"); data edited or captured while a file is being written stays marked as unsaved.
    Files are written through links, keep their permissions, respect write protection and may
    have long names again. A flow gives the Ctrl+C handler of the program around it back. A
    decoder channel assigned to a capture channel that a profile or the capture does not have is
    kept instead of being dropped by saving the profile or pressing OK in the decoder options.

  • Command line and smaller corrections.

    • capture checks the ending of the output file and its options before it captures (a
      capture was thrown away for -o x.txt). Errors of damaged files, flows that cannot be
      read and devices are one line instead of a traceback (OPENSCILAB_TRACEBACK=1 shows it).
    • capture --duration 5m is 5 milliseconds now, as in flows and run --duration
      (it was 5 minutes there and 5 ms here); minutes are 5 min.
    • Analog to digital with hysteresis is computed without a loop (2 million noisy samples:
      0.6 s → a few milliseconds).
    • Patterns (SDL): // inside a string is no comment, \\n is a backslash and an n, and a
      repeat count that would make a pattern longer than 64 million samples is refused instead
      of hanging.
    • Removed what could not be reached or was not used: the help menu of a data view (the
      window has it), a few helper functions.
  • Scripts and the command line without the user interface. No module of the core, the
    drivers or the lab loads Qt any more (the colours of the waveform moved to the user interface,
    a region holds its colour as four numbers): import openscilab.api takes 110 ms instead of
    195 ms, and the lab is loaded when a script first uses it.

  • Start-up. The theme Follow the system asks the application for the colour scheme (it
    was always dark on Linux and started a helper program on macOS); the style sheet is built
    and applied once instead of three times. Restarting after a change of the theme opens the
    documents again (the session was saved a second time, empty).

  • Drivers: what happens when two things happen at once (checked against simulated
    transports, not on hardware):

    • Pico: asking a board for its capabilities or details while a capture starts no longer
      writes the question into the capture and reads samples as the answer; a stopped capture
      leaves no reader behind that could take the first bytes of the next one. A set of boards
      sends the bootloader command to every board (it stopped at the first that did not answer)
      and reports the result of a capture without holding its lock.
    • DSLogic: every capture has its own stop flag, so a capture started right after one was
      stopped is neither read nor ended by the reader of the old one. A stream the computer
      cannot keep up with ends with a message instead of filling the memory.
    • Software trigger: a second capture started while one waits for its trigger is refused
      without taking away the watch of the first (its endless stream would never have stopped);
      analog channels with a rate of their own are cut at the right place.
    • Simulator: stopping a capture no longer interrupts the transfer of the capture before it.
    • Scripts: Device.capture waits until all samples of a capture that arrives piece by piece
      (an oscilloscope over the network) are there; saving or decoding it at once saw zeros.
    • The window no longer stutters every few seconds while the serial ports are listed (read in
      a thread now), and writing a firmware with avrdude, bossac, esptool or adb no longer
      freezes it for minutes.
  • The data view stays quick with large captures (measured with 20 million samples on 24
    channels and 47,000 annotations):

    • A change of the data – an edit, an undo, a renamed channel – blocked the window for 2.1 s;
      now 20 ms. Moving a channel: 2.2 s → 60 ms. The index of the edges and the values of the
      buses are kept for the channels whose samples did not change, the rows of the channel
      column are updated instead of built again, and the statistics of the Measure tab are
      read from the edge index, only while the tab is shown.
    • Annotation rows with more entries than pixels are drawn as blocks: 640 ms → 5 ms per
      picture, moving a cursor over them 65 ms → 3 ms. An entry that began far before the view
      (a frame around many bytes) is drawn; it could be missing.
    • Bus rows zoomed out: 245 ms → 7 ms per picture, pixel for pixel the same.
    • Dragging a cursor no longer computes minimum, maximum and RMS of the analog channels over
      the whole range at every step.
    • The search runs in the background and can be cancelled; the filter of a long annotation
      list waits for a pause in typing, and an entry that spans the whole row no longer makes
      every lookup search the row from its start.
    • Undo keeps as many steps as fit into 2 GB of replaced samples (4 for that capture, 50 for
      small ones) instead of 5 steps of any size.
    • The preview of Play as signal is drawn from the runs of the pattern, not from every
      sample.
    • States shown on their real time: the regions and markers of the states wait for the way
      back instead of marking wrong times, and saving writes the states, not the picture of
      them. A state analysis no longer carries the regions of the timing capture.
  • Capture files: small, quick, any size.

    • .lac files hold the samples packed (8 per byte, compressed) instead of a number per
      sample: a capture of 5 million samples on 24 channels is a file of 0.05 to 20 MB instead of
      360 MB, saved in 0.1 s instead of 5 s and opened in 0.1 s instead of 7 s with a gigabyte
      less memory. Captures recorded to disk that did not fit into memory can be saved and
      opened now. Every file written so far still opens.
      The original LogicAnalyzer software cannot open the packed files: Save as → Captures
      for the original LogicAnalyzer (*.lac)
      (in scripts Capture.save(path, compatible=True))
      writes its form, a number per sample. A file that has that form keeps it when it is opened
      and saved again; only Save as changes the form.
    • Opening, saving and exporting run in the background with a small dialog; the window stays
      usable and a second save cannot start meanwhile. A file that is damaged or cut off is
      reported as such in every case (some errors ended silently).
    • CSV and VCD exports are written block by block: 2 million samples on 16 channels in 0.06 s
      instead of 3.2 s (CSV) and 0.75 s instead of 1.9 s (VCD), without a copy of the capture in
      memory. A channel name with a comma is quoted in the CSV; a VCD never contains nan, and a
      capture without samples is refused with a message.
    • .lac.gz is compressed at level 6 (level 9 took much longer for a few percent).
  • A running flow no longer floods the window.

    • What the engine reports is collected and shown 25 times a second: the last state of every
      node, the values of every output as a batch (charts of panels and the sparklines of the
      wires still get every value), the log, the views. A flow in virtual time sends hundreds of
      thousands of values a second; each one used to be a signal, a search through all wires and
      a repaint of the scene and the minimap.
    • A value in a chart, number or table no longer rebuilds the menus, the inspector and the
      project lists of the window (it did so for every value); neither does an undo step or a
      keystroke in a document. The data list of the project is read when the folder changes and
      when a run ends.
    • The scope of a flow keeps the zoom and position while more of the signal arrives.
    • Charts: thousands of points are drawn as their envelope (two points per pixel column);
      a value that is no number (a measurement without a result) no longer blanks the chart of a
      panel; zooming in very far no longer freezes the window; numbers, LEDs and charts of a
      panel show the last sample of a block of samples (an LED was always on for one); tables
      add their new rows instead of being filled again.
    • Flow editor: the inspector is replaced, not collected (every click left one behind), and
      typing the name or description of the flow keeps the cursor; the ports of the nodes are
      only built again when they change; search popups, the command palette and detached
      windows are freed; closing a document whose flow still runs no longer prints errors.
    • A control of a panel bound to a node the flow no longer has is skipped and named in the log
      instead of stopping the start of the run.
  • Flows that run for days.

    • Memory no longer grows with the run: the engine keeps what every view shows now instead of
      everything it ever showed (the Long-term logger template took 750 MB after 60,000 values,
      now a constant 80 MB, and runs twice as fast), and the last 10,000 lines of the log.
    • Timers and the monitor keep their period (the n-th tick at n × interval; the time a tick
      took no longer adds up), a flow in real time wakes a hundred instead of a thousand times a
      second, and duration also ends a flow whose stream is still open.
    • The logger names the source of every row (the pin, e.g. A0, instead of its own name),
      writes to disk after flush rows or five seconds, logs the samples of analog and digital
      blocks, and the rows of a growing table once each.
    • Reports render each section, table and diagram once when they are written instead of on
      every value; diagrams of long captures show their envelope; PDF reports contain the
      diagrams
      (they were missing).
    • Two flows could not be stopped: a sequence that only sets, emits or logs with repeat: 0,
      and an endless timer with interval 0 (now refused). Nodes that wait for each other to take
      their values end the flow with the error deadlock instead of "finished".
    • Pins, PWM, voltages and the monitor of a real device are called in a thread, one call at a
      time per device: a device that is slow or does not answer no longer holds the other
      nodes, Pause and Stop.
    • Smaller: text parameters of several lines keep their line breaks exactly when saved;
      parameters may be called position or node_id; a function node with a params
      argument validates; a capture rate below 1 Hz is refused with a message; Browse… stores
      the path the flow resolves to the chosen file, also in subfolders of the project.
  • Decoders: about eight times faster, and safe to close.

    • Waiting for an edge or a level is looked up in an index of the channel's edges instead of
      searching 65,536 samples each time: a UART capture of 2 million samples decodes in about
      3 s instead of 26 s, with identical results.
    • Closing a data view (or the application) while it decodes no longer risks a crash: the
      decoders are told to end and the view lets go of them. A new decode requested while one
      runs ends the old one instead of waiting for it.
    • An edit of the samples or an undo decodes once, not twice; the options dialog can be used
      while a decode runs without mixing old and new settings.
    • Flows and data views share one set of decoders (the decoder folders of the application
      apply to flows as well), loaded once. The node palette imports the decoders only when its
      Decoders group is opened or searched, so the window starts faster.
  • Review of the code base: nothing is lost or silently wrong.

    • Decoders keep reading their channels when the channels are reordered (dragging a channel
      made them decode another one), also after the next capture and with profiles that capture
      only some channels. A decoder assigned to a channel the capture does not have says so.
    • A flow that ends with an error, with Ctrl+C or by Stop still writes what it measured: the
      rows of a logger, the recorded samples and the report (which then says what failed). Stop
      pressed right after Run is no longer lost.
    • Subflows with a required input run (Make subflow produced flows that could not start),
      and a subflow may use the other subflows of its flow.
    • Captures, profiles, waveforms, projects, panels, flows and reports are written to a
      temporary file that replaces the old one only when it is complete: a crash or a full disk
      no longer destroys the file that was there. A settings file that cannot be read is kept as
      <name>.bad instead of being overwritten, and one unreadable profile no longer takes the
      others with it.
    • Markers and pinned channels are saved in the capture file and come back when it is opened.
    • A selection in the ruler is cleared when other data is shown (cut or delete acted on the
      old positions); shifting channels of different length, inserting samples with an empty
      channel and search hits after an edit are right.
  • Simulators: several at once, multi device sets, and what they simulate. Every simulator
    can be connected again and again (sim:uno, sim:uno#2, …), each an instrument of its own.
    Simulated multi device… combines 2 to 5 simulated boards into one device captured together
    (sim:pico*2: 48 channels), like a multi device set of real boards. The new Signals tab of
    a simulator's device card says what it simulates: a UART, SPI, I²C, all three, a counter, the
    C64 bus (a Commodore 64 at its expansion port, on the channels of the profile C64
    expansion port
    ; needs 48 channels), the channels of a capture file, or nothing – and any
    source on a single channel (double-click it). It applies at once, names the channels for the
    next capture, is kept per device and comes back with the session. In a flow the device node
    takes it as signals. A pin of a simulator shows its signal again after it was an output.

  • Device handling: more reliable and faster.

    • Opening a device no longer freezes the window: boards, network devices and the DSLogic are
      opened in the background with a small dialog (Cancel stops waiting), also when the session
      reconnects its devices at startup; the power of a network device is asked in the background.
    • Stopping a capture that waits for its trigger ends it everywhere (card, data view, header);
      stopping never blocks the window.
    • A device that stops answering (a failed capture or no answer over the network) is marked
      with an error and its card offers Reconnect; a network board that vanished is noticed
      through TCP keep-alive instead of a capture waiting forever.
    • Pico driver: a board still streaming from a program that crashed is resynchronised on open;
      stopping and starting again at once no longer lets the old capture end the new one; a board
      that does not answer the capability question is not asked again for every property (each
      time 5 s); device details are read once; serial ports are opened exclusively; the timeouts
      of the identification follow the connect timeout.
    • Flows leave the devices of the device list tidy: a capture or stream they started is
      stopped when the flow ends early, generators stop also after an error, and simulators wired
      for the run are wired as before.
    • The simulator: a failing source ends the capture with an error instead of leaving the
      device "capturing" for good; wiring loops are refused; stopping a capture that waits for a
      trigger works at once (it could take 10 s) and no longer spins a core; endless streams with
      analog channels use constant memory; state captures sample thousands of times faster;
      sources are read consistently while outputs change in another thread.
    • Less work in the background: firmware images and simulator profiles are read once, serial
      ports are listed once per look, the overview of the connected hardware only scans while it
      is shown, and the pin values of a fast monitor update 30 times a second.
    • A listener of the device hub that fails no longer breaks adding or removing devices; a
      removed device takes its capture controller, handlers and timers with it; closing a device
      card leaves the monitor of a running recording alone; closing the hardware overview while
      a version is being read no longer risks a crash.
  • Polish. Shortcuts are written the way the platform does (⌘K on macOS); keys without
    Ctrl/Cmd only act in a data view that has the focus; Measurements moved to Ctrl+Shift+M
    (Cmd+M minimises on macOS); Esc means All outputs safe on the device card's Pins tab only;
    the shell's shortcuts and the document's menus work in documents opened in a window of their
    own; device card and panel commands are in the command palette; Help → Keyboard shortcuts
    is a searchable list with the keys and gestures of the views. The data view's view menu is
    called Display; file dialogs say Captures; saving with Compressed captures adds
    .lac.gz; code and logs use the system's fixed-width font. The device card, the data view
    and the side panels need less width, so a card and its data view fit side by side.

  • Capture settings in pages; accessibility. The capture settings are pages – Sampling,
    Channels, Trigger – with a line below that says what the capture will be; a problem opens
    the page where it is fixed and marks it. Every button that shows only an icon has a name for
    screen readers (from its tooltip), the flow graph and the waveform describe what they show,
    and the status dots are sharp on high resolution screens.

  • Panels arranged with the mouse, charts that zoom. The panel editor draws its grid with free
    rows below the widgets: drag a widget from the palette onto a cell, drag widgets to move them
    and their corner to resize them (a taken place is shown red and refused); every step can be
    undone. The Port list offers the ports of a type the widget shows, Tab and Group are chosen
    from the existing ones (or typed new), a widget whose port does not exist is marked, and Edit
    and Operate are toggles in the toolbar. Charts and the plots of panels zoom with the wheel or a
    pinch (Shift: values only), move with two fingers or by dragging, and go back to the automatic
    range on a double-click or smart zoom.

  • Undo in the data view, channels in any order. Deleting, inserting, pasting and shifting
    samples, derived channels, renaming, colours, hiding, the order of channels, markers, regions
    and buses can be undone and redone (Edit → Undo/Redo, Ctrl+Z); large captures keep fewer
    steps. Cut, copy, paste and delete of the selected samples are in the Data menu with their
    usual shortcuts, as are Insert samples… (now also for files and test signals) and Shift
    channels…
    . Channels are reordered by dragging their name or with Move up/down; the order
    is saved with the capture. The button above the channel names says how many channels are
    hidden and shows them again one by one or all at once. Regions can be renamed and recoloured
    (Markers → Edit…). Data → Use these channel names for the next capture hands names given
    in the data view to the device.

  • The last session comes back. openSciLab opens the documents of the last session again –
    files, device cards, data views of a device, the start page, the connected hardware – in their
    groups and side by side, with the active one in front, flows and captures zoomed as they were,
    and the console as it was; the devices are connected again. Files that are gone are left out
    (noted in Log), devices that are not there are named in the device list. Unsaved documents
    are not kept. Settings → Startup turns this off and chooses whether the start page shows. A
    window saved on a screen that is no longer there opens on one that is.

  • Settings and a light theme. Project → Settings… (Cmd+, on macOS) sets the theme (dark,
    light, like the system; applied after a restart the dialog offers), the text size, what happens
    at startup, whether the mouse wheel zooms or moves and in which direction, the folder for data
    and new projects, the number of recent entries, how often devices are looked for and whether
    simulators are listed. The light theme covers the shell, the documents, the flow graph,
    comments, charts, LEDs and the pin strip; the waveform display stays dark like an instrument
    screen. Text keeps a contrast of at least 4.5:1 in both themes.

  • A friendlier flow editor. While a wire is dragged, the ports it fits light up and the others
    fade; it snaps to a fitting port nearby or to the right port of the node it is dropped on. A
    wire dropped on the empty canvas offers the nodes it fits (placed there and wired), a wire that
    does not fit says why above the graph and offers the conversion node. Right-click a wire to
    insert a node into it or remove it, a port to add a connected node or disconnect it. Copy,
    cut, paste, duplicate (Ctrl+D) and select all work on nodes with their wires; new nodes no
    longer land on top of each other; Tab or Add node in the toolbar opens the node search, and
    an empty flow says how to begin. Required inputs without a wire are marked, nodes show a badge
    for their problems and their type next to their name, the breakpoint dot shows on hover.
    Adding the first device wires the nodes waiting for one (the status bar says what was wired).
    Group frames carry the nodes inside them, are grabbed by their title and edge and resize at
    their corner (comments too). The inspector marks required parameters, lists the node's
    problems, offers the pins of the wired device and … to choose a file, says when changes
    apply only to the next run, keeps multi-line descriptions and checks seed and duration. Less
    used tools in the flow toolbar show their icon only; the inspector wraps instead of cutting
    text off.

  • Clearer ways through the application. The start page comes first with Try a simulator
    (a simulated device, its card open; its first capture shows data at once) and Connect a
    device
    ; with every document closed the area offers a new flow, connecting, opening a file or
    the start page. Devices → Connect… lists detected boards, network and multi devices and the
    simulators in one dialog; the Devices menu reaches the card, the data and Disconnect of every
    open device, Install or update firmware… and outdated firmware lead to Connected hardware,
    and a device whose firmware was installed there opens again by itself. Disconnecting asks while
    a capture runs, outputs are driven or a flow uses the device; an unplugged or disconnected
    device card offers Reconnect. The first capture opens its data view beside the device card,
    the header's Capture / Capture again / Stop work for device cards and data views, and the
    card shows the state of the capture. Connection errors, failed captures and flows that cannot
    run appear as a banner where they happened (a run without its device offers Run with
    simulators
    ); Check and running a flow open the console, the status bar counts the problems
    and a double-click on a problem shows its node. Use settings (formerly Apply) keeps the
    capture settings, Enter in a channel name moves to the next name, and the card's Profiles is
    the only profile menu. The Project menu has New project from template…, Open project… and
    recent projects; the Project sidebar lists the flows, panels and waveforms of the project.
    Untitled documents are numbered (Untitled flow 2), a data view names its device, Ctrl+Tab
    switches tabs, and Split needs a second document. The Devices menu's mnemonic is Alt+V
    (Alt+D is the data view's Data).

  • Data, device and project fixes. Every data view has its own decoders (adding one no longer
    changes the others). The data view of the Logic Analyzer template receives the first capture.
    Opening a file or test signals over unsaved data asks first; renaming channels, colours,
    markers, regions and buses count as unsaved changes; a capture never replaces data the user
    changed (it opens in another data view); Capture again uses the settings of the last capture,
    not samples edited since; a capture whose data view was closed opens in a new one. The device
    card shows the settings of the next capture, its capture bar works for simulators, and
    loading a profile sets the next capture of this device only (and updates the bar). A status
    change keeps switches and values in the pin table, a declined switch goes back, and the
    Monitor box follows the monitor (also when a recording starts or stops it); a new rate
    applies at once. Devices found by autodetect, over the network or as a multi device carry
    their real address, and unplugged devices are shown as disconnected. Choose an image… no
    longer restarts the port of an open device behind the application's back.

  • Projects and closing. Closing the last document of a temporary project asks whether to
    keep it; its files no longer fill the recent files. Save as in a temporary project saves the
    project first, and Save project as lets you name the project folder. Closing the window asks
    once for everything unsaved, and Cancel closes nothing. Opening a project without flows or
    panels says so instead of doing nothing. The status bar shows real or virtual time of the
    active flow or panel and which documents are unsaved; the window title marks unsaved changes.
    The sidebar stays as you left it at startup. Messages point to the right menus (Data → Align
    boards
    , Devices → Connected hardware).

  • Flow and panel editor fixes. A double-click in the node palette adds one node (not two);
    a new device node starts with the default device instead of failing when the flow runs; a click
    on a node no longer marks the flow as changed, and two moves are two undo steps; saving a flow
    opened from a Python script saves it as .flow.yaml instead of overwriting the script; YAML
    with an error is kept when switching views (keep editing or discard), and the edits of one visit
    of the YAML view are one undo step; an error tooltip goes when the node is idle again; Run is
    available again as soon as a run has ended; renaming a node keeps the ports of a subflow; the
    palette shows the node types of the active flow's project. The panel inspector keeps the
    cursor while typing, steps of one setting are one undo step, Redo is enabled after Undo, and a
    panel runs the current flow (the open flow document with its changes, or the file read again
    when it changed). Save dialogs without a project start in ~/Documents/openSciLab.

  • Trackpad and pinch in the flow graph. Two fingers move the canvas instead of zooming it,
    pinch zooms at the fingers, a two-finger double tap fits the flow, Cmd + swipe zooms; the mouse
    wheel zooms at the pointer (Cmd + wheel sideways, Shift + wheel up and down). The zoom stops at
    its limits instead of ignoring the step, far nodes stay reachable, Flow → Zoom in / Zoom out /
    Fit (Ctrl+0) / Actual size
    are new, and switching to YAML and back keeps zoom and position.
    The overview can be dragged and passes wheel and pinch on to the canvas.

  • The same navigation in every part of the data view. Analog tracks and buses follow the
    trackpad instead of zooming on every swipe, pinch works over the ruler, buses and analog tracks,
    Cmd + swipe zooms the waveform, and the overview below the waveform follows the trackpad. The
    panel editor scrolls also with the pointer over a widget.

  • Clearer device card. The card has tabs: Capture (settings of the next capture, Start,
    the channels with their names), Pins (pin, channel, name, mode, value, actions; PWM and DAC
    controls on lines of their own), Events (simulators) and Details. Device information,
    self-test, network, bootloader and firmware moved into the Device menu of its header. A
    declined or failed switch goes back to Off.

  • Apply in the capture settings. All settings… → Apply keeps the settings without capturing;
    the channel names appear in the card's channel and pin tables at once.

  • Connected hardware. Devices → Connected hardware lists every board and device on the
    computer with its firmware and whether it is current. Install/Update firmware picks the image
    that fits the board, restarts it into its bootloader, writes the image and waits until it is
    back; Choose an image… keeps the manual dialog.

  • PiPiLogicAnalyzer is now openSciLab. The Python package is openscilab, the commands are
    openscilab, openscilab-cli and openscilab-gui, the settings live in the folder
    openSciLab and the environment variables are OPENSCILAB_SETTINGS_DIR and
    OPENSCILAB_DECODERS. The old names are gone (no aliases, settings are not taken over). The
    packaged applications are called openSciLab-<version>-<platform>. The Pico firmware keeps
    its name until the next firmware release.

  • New lab shell. The application opens in the openSciLab shell: an activity bar and sidebar
    (project, devices, search), documents in tabs that can be split side by side or below each
    other and opened in a window of their own, an inspector on the right, a console below
    (problems, execution, log and a Python console with access to the shell) and a command palette
    (Ctrl+K) that reaches every menu entry, with fuzzy search and the recently used commands first.
    A start page lists recent files and the Logic Analyzer template. Layouts can be saved and
    restored (View → Layout).

  • The logic analyzer is a document. Everything the main window did is unchanged inside an
    analyzer document: its Capture, Device, Analyze and Profiles menus appear while it is
    active, its view menu is called Waveform. Several captures can be open at once. A capture
    with unsaved changes is marked in its tab and closing it asks first. Files given on the command
    line open in analyzer documents.

  • Several devices at once. Opened devices are instruments in the device hub: the sidebar
    section Devices lists the open ones with their status and the ones that can be opened, the
    header shows a chip per device, and each device has a device card (capabilities, pins,
    connection). An analyzer document captures with a device of the hub; closing the document
    keeps the device open for others, Disconnect closes it.

  • For scripts and later steps: typed signals (openscilab.core.signals: digital, analog,
    scalar, bool, event, states, capture, table, with time bases and conversions), instruments
    with facets (openscilab.core.instrument) and the hub with trigger routes and time base
    offsets (openscilab.core.hub).

  • Flows. A measurement program is a flow: nodes (devices, timers, sweeps, tables, files,
    views) and the wires between them, stored as *.flow.yaml or written in Python with the DSL
    of openscilab.lab (with flow(...) as f: n.device.capture(...) >> n.data.file(...)); both
    convert into each other unchanged. The engine runs flows in real time or, with simulators, in
    virtual time (fast and deterministic), with back pressure, breakpoints and single steps.
    Projects are folders with project.yaml, flows/, data/ and own nodes in nodes/.

  • openscilab run <flow> [--sim] [--fast] [--device name=address] [--report out.html] runs a
    flow without the window, openscilab sim lists the simulator profiles; openscilab capture
    and the other commands of openscilab-cli work as openscilab <command> too.
    api.run_flow() does the same from scripts.

  • Simulated instruments (sim:free): a 16 channel logic analyzer with test signals
    (counter with clock, UART, SPI, I²C, square wave), buffer captures with edge and pattern
    triggers and streams, with the limits of its profile. api.open("sim:free") opens it.

  • Visual flow editor. A flow document has three views: Graph (nodes and wires on a canvas,
    wires coloured by type, minimap, zoom and pan), YAML (the file itself, editable) and Python
    (the same flow as a DSL script). Nodes come from the palette in the sidebar (search, drag and
    drop) or by typing on the canvas; the inspector edits their parameters. Wires whose types do
    not fit are refused, with an offer to insert a conversion node. Undo/redo for every edit,
    comments, groups, alignment, subflows (opened as documents of their own). Running a flow shows
    the state of every node and the latest values on the wires (a sparkline for numbers);
    breakpoints and single steps work in the graph, scope nodes open analyzer documents.

  • Node library: measurements (frequency, period, pulse width, duty cycle, min/max/mean/RMS,
    peak to peak, counts, setup/hold), signal processing (threshold with hysteresis, debounce,
    FIR/IIR filters, FFT, derivative, integral, averaging, resampling, expressions), every sigrok
    decoder as a node (decode.uart, decode.i2c, ...), control (sequences without code, state
    machines, Python nodes, compare, limit, counter), data (tables, buffers, reading and writing
    captures and CSV, long-term logger on disk), views (scope, strip chart, XY chart, spectrum,
    table, number, LED, log – each opens a document while the flow runs), bundles. Projects add
    their own nodes as Python functions in nodes/.

  • Analog channels. Devices with analog inputs capture them together with the logic channels;
    the analyzer shows them as tracks below the digital channels (envelope zoomed out, line and
    points zoomed in), with their value at the pointer and at the cursors, min/max/mean/RMS/peak to
    peak in the measurements, and a digital channel from a threshold with hysteresis. .lac files
    keep them (older files load unchanged), sigrok sessions for PulseView carry them as analog
    traces, CSV exports them in volts and VCD as real.

  • Large captures arrive progressively: an overview of the whole capture at once, the samples
    in full resolution after it, the part on screen first; saving, export and decoders wait for the
    rest; an interrupted transfer can be resumed.

  • Live: roll mode (the view follows the newest samples or holds still), analog channels and
    losses while streaming.

  • State captures with times: the time of every state is kept; the view shows the states one
    by one (numbered axis, cursors with state and time) or on their real time. State mode on a
    stream (live state) and software triggers on states; the clock input is chosen from the pins
    that can clock.

  • Simulator: analog sources (sine, triangle, ramp, noise, sums), state captures, the provisional
    profile sim:dho924s (4 analog and 16 logic channels, progressive transfer).

  • Controlling pins from the device card. Devices with GPIO show their pins as a strip and a
    table: mode per pin (input, pull-up, output, PWM, analog), switches, pulses of a set width, PWM
    and voltages of analog outputs; reserved pins are greyed out with the reason. The first output
    on a pin asks once about its level. All outputs safe (Esc) turns every output back into an
    input. While outputs are driven the card keeps the device's watchdog fed. The event log lists
    what the device did.

  • Monitor: the card reads the inputs periodically (levels and voltages with a short history);
    Record records the monitor as a slow capture in a new analyzer document.

  • Markers for actions: everything set on a device while a capture runs is a marker in that
    capture (at once in a live display, placed when it completes otherwise, marked with ≈ when the
    position is estimated).

  • Stimulus and capture: the card arms a capture with any connected instrument (also another
    device) and then pulses the selected pin.

  • GPIO nodes for flows: gpio.write, gpio.pulse, gpio.pwm, gpio.read, gpio.dac and
    device.monitor.

  • Simulator sim:uno: an Arduino Uno with GPIO latency, exact pulses, a watchdog, D7 wired to D3,
    the PWM of D9 through an RC low pass to A0 and a bouncing push button on D2; faults can be
    injected (disconnect, delay, restart). Buffer captures of the simulator wait for their trigger
    in real time, so a stimulus after arming is captured.

  • Signal generation. A waveform document builds what a generator output plays: standard
    shapes (sine, square with duty cycle, triangle, ramp, pulse, DC, noise) with sweep and burst,
    arbitrary waveforms from a formula or a CSV file, and digital patterns as SDL per pin with a
    preview of the tracks. It lists the generator outputs of all connected instruments, explains
    what does not fit an output (frequency, voltage, points, rate – a too fast pattern can be
    limited to the output's rate) and starts and stops it. Waveforms are saved as *.wave.yaml;
    *.sdl files open too.

  • Play as signal (Analyze menu of the analyzer): digital channels of a capture as a pattern,
    or an analog channel as an arbitrary waveform, between the cursors if they are set.

  • Generator nodes for flows: gen.waveform, gen.arbitrary, gen.pattern, gen.tx_uart,
    gen.tx_spi, gen.tx_i2c, gen.replay and gen.output; their sync output marks the start.
    The capture node reports armed once it waits for its trigger.

  • Simulator: sim:free has a pattern generator on D0-D7 (looped back to the inputs) and a
    trigger input on D15, sim:dho924s a 25 MHz generator output on CH1 with its sync on D15;
    trigger routes of the hub between simulated instruments are wired (a sync output triggers a
    capture on another instrument).

  • Panels. A panel document (*.panel.yaml) is the front panel of a measuring station:
    switches, push buttons, sliders, inputs, choices, numbers, LEDs, charts, an embedded scope and
    labels on a grid with tabs and groups, each bound to a port of a flow (inspector). In
    Operate the controls send their values into the running flow and the displays show its
    outputs; F11 shows the panel full screen as an application of its own. Undo/redo for every
    change of the layout.

  • Project data: the project section of the sidebar lists the captures, tables, logs and
    reports of the project's data/ folder; captures open in the analyzer, two of them can be
    compared, one exported as CSV; tables and reports open in their system application.

  • Reports: nodes report.section, report.table, report.image (XY diagrams and traces as
    SVG), report.check (limits or an expected value, passed or failed) and report.write (HTML,
    or PDF). openscilab run --report puts the same parts into its report, and its exit code is 1
    when a check failed.

  • New template Characteristic curve (start page): a PWM duty sweep on the simulated Uno
    measured behind an RC low pass, with a panel, a report and a check of the slope. Templates are
    projects copied to a folder you choose; recent projects open from the start page.

  • data.table keeps the values of each column in order, so a column that runs ahead no longer
    overwrites values before their row is complete; control.sweep can be started by a value at
    trigger (a start button).

  • Simulators complete: profiles sim:uno, sim:uno_r4 (DAC on A0 wired to A1), sim:pico
    (24 channels, depth by the number of channels, ADC, pattern generator), sim:dho924s and
    sim:free, as JSON files in examples/sim/ (own profiles in the settings folder sim). The
    device list of the sidebar offers every profile (Simulation: …); simulators opened there run
    on the clock of the hub. Devices → Simulation injects faults (disconnect, delay, overflow,
    restart), scripts use device.inject(); api.instrument("sim:uno") gives an instrument with
    all its facets. A project wires simulated circuits further (simulation.wiring in
    project.yaml), the source type pulse is new, and the same seed gives the same samples in
    virtual time. The profile keys, sources, wires and faults are described in docs/simulator.md.

  • Templates on the start page: Empty lab, Logic Analyzer, Characteristic curve (now with
    the simulated Uno and DHO924S: the scope measures the RC voltage), Test bench (checks with a
    report, a test on the command line) and Long-term logger (CSV and strip chart).

  • Panels run in virtual time if wanted (Virtual time); a project wires two simulated devices to
    each other (simulation.wiring: {from: "uno:A0", to: CH1}).

  • Documentation: the README starts with the lab, the analyzer is a chapter; docs/lab.md
    (window, instruments, flows, all nodes, panels, generation, reports, projects),
    docs/simulator.md, docs/protocols.md, facets in docs/drivers.md.

  • Devices are nodes of a flow. A device.instrument node stands for an instrument (a
    simulator, a device address, or an instrument already open in the device list, which is used as
    it is); its device output is wired to the nodes that use it (capture, stream, monitor, GPIO,
    generator). The palette offers the open instruments, the project's devices and the simulators,
    the device list Add to the flow; a new node is wired to the flow's only device at once. The
    devices: section and the device: parameter of flow files are gone.

  • The analyzer document is a data view. It shows, analyses, decodes and edits captures (also
    files and test signals) and no longer connects devices or holds their settings. Capturing is the
    device card's business: its capture bar (rate, length, trigger, Settings…, Start F5,
    Capture again, Stop), Profiles, Test signals of the board, Device information,
    Self-test, Network, Bootloader and Update firmware; the result appears in the device's
    data view, whose toolbar names the source (a click opens its card) and can capture again or
    stop. Devices are opened only in the device list, which also notices boards without firmware
    and offers the update of outdated ones. Forget multi device sets moved to the Devices menu.

  • A flow from the start: the application opens with an untitled flow beside the start page; it
    writes into a scratch folder until it is saved. Templates open as temporary projects without
    asking for a folder; the first Save (or Project → Save project as…) asks where to keep the
    project and moves its open documents along.

  • New dependency: PyYAML.

Firmware

  • openSciLab Pico, protocol 8. The Pico firmware is now openSciLab Pico in firmware/pico/
    (source files without the old prefix), released with the application (0.1.0b1) and speaking
    protocol 8. It identifies itself as
    OPENSCILAB_PICO_<BOARD>_V8_0, its USB product name is openSciLab Pico (VID/PID unchanged),
    the images are called openSciLab-pico_<BOARD>[_Turbo].uf2. Boards with PiPiLogicAnalyzer 7
    are recognised and offered the update; the application only works with protocol 8.
  • The capability strings of the Pico firmware come from capabilities.h; docs/protocols.md
    specifies the whole vocabulary (capabilities, the facets they unlock, pin capabilities, the
    wire protocols) and a test checks the firmware header and the application against it. The
    firmware reports the same capabilities as before.
  • firmware/build_all.sh builds by firmware (--pico [boards], --arduino, --bridge); a dev
    container (.devcontainer/) has the Pico SDK, the ARM toolchain, picotool, PlatformIO and the
    Android tools. The release file with all Pico images is called
    openSciLab-firmware-pico-uf2.zip.
  • Devices → Install or update firmware… and the device card update the firmware of a Pico. For
    the other boards the update tools (avrdude, bossac, esptool, adb) are prepared in the
    application, but no driver uses them yet: Arduino boards are flashed with PlatformIO or
    avrdude, the bridge app with adb.
  • openSciLab Pico 8 does more (protocol 8): pin table, pin modes, outputs, PWM, pulses
    exact to a clock cycle, the monitor, the watchdog (outputs are released 1 s after the host is
    gone), analog channels by DMA, the pattern generator on up to 24 pins, UART/SPI/I²C transmitters
    and the live state stream. Host tests of the frame parser, pattern data and pin tables run in CI.
  • openSciLab Arduino firmware (firmware/arduino/, PlatformIO) for Uno, Nano, Mega 2560,
    Uno R4 Minima, ESP32 and ESP32-S3, with native tests and a benchmark project.
  • Bridge app for the Rigol DHO900 (firmware/bridge-android/): passes SCPI through, keeps
    captures in a cache, sends overviews and compressed tiles and announces itself on the network.
  • None of the three has been tried on hardware yet.

Based on and extending the LogicAnalyzer by Agustín Giménez Bernad (gusmanb). Thank you!