UserGuide
Table of Contents
- DropBot DB3-120 kit
- Install MicroDrop
- Connect DropBot
- Run the on-board hardware diagnostic tests
- Attach the Webcam
- Insert a DMF chip
- Load liquid onto the chip
- DMF chip specifications and sample mask
You should have received the following items with your DropBot kit (see here for a larger image):
- DropBot DB3-120 instrument
- +12V power supply
- Test board
- USB A/B cable
- 3x DMF chips
- Device spacer
- Webcam
- Webcam holder
- Test tube containing 0.1 M NaCl + 0.1% Tetronic 90-R4
MicroDrop is a Windows app for programming and controlling the DropBot and the latest release can be downloaded here.
MicroDrop currently supports Windows 7/8/10.
Note: If you are running Windows 7, you will also need to install this driver before connecting the DropBot.
- Plug +12V power supply into wall outlet.
- Connect +12V power supply to DropBot.
- Connect USB cable between PC and DropBot.
Once you have MicroDrop installed and have connected the DropBot, the next thing you will want to do is test that your hardware arrived in a working state:
- Launch
MicroDropfrom theWindows Startmenu. - From the
Tools\DropBotmenu, selectRun all on-board self-tests.
Note: If you have any trouble connecting to the DropBot, please see the Troubleshooting guide.
The software will pop up a wizard showing you how to insert the test board into your DropBot. These tests confirm that:
- The high-voltage source is working.
- The feedback-sensing circuit is working.
- There are no short-circuits on any of the actuation channels.
- All actuation channels are functional.
See below for an example report:
Attach the webcam to the DropBot as shown below:

Note: For (optional) advanced control over webcam settings (e.g., fixed focus), download the Logitech Webcam Software from the Logitech C920 support page.
The following steps are illustrated in the figure below (see here for a larger image):
- Pick up DMF chip by edges, being careful not to touch the surface.
- Insert DMF chip into front of DropBot, and;
- ...push forward until it hits the back of the chip guide.
- Insert device spacer under the chip guide surface.
- Push device spacer in as far as possible.
See this screencast demonstrating how to register the webcam video to align with the device drawing in the DMF device user interface.
To load liquid on the DMF chip:
-
Actuate a reservoir electrode using the MicroDrop software.
-
Manually pipette ~6 uL of liquid at edge of top plate over reservoir electrode.
The DropBot DB3-120 kit includes sample DMF chips
with a 90-pin electrode array configuration.

More of these chips are available for sale from Sci-Bots (please contact sales@sci-bots.com for a quote)
Note that custom chips may also be used with the DropBot DB3-120 system.
Please see the DropBot DB3-120 chips specifications for more information.