GeigerMeasurement

An Arduino library for Geiger-Müller counters on ESP32 and ESP8266, implementing the core measurement algorithms of the RadPro project by Gissio, adapted for the Arduino platform.

The library is split into two independent components that can be used together:

File Purpose
GeigerTubes.h Tube sensitivity and radiation source correction data
GeigerMeasurement.h Pulse buffering, rate calculation, dead-time compensation

Features

GeigerMeasurement

  • Five averaging modes: Adaptive Fast, Adaptive Precision, 10s, 30s, 60s sliding window
  • Dead-time compensation (non-paralyzable / Type I model) with integer carry for long-term accuracy
  • 95% confidence interval (Poisson statistics with first-order correction, RadPro formula)
  • In-situ dead-time measurement from pulse buffer minimum interval
  • Live calibration against a known dose-rate source
  • Fault detection: tube alive, dead-time saturation, pulse counter overflow
  • ESP32 / ESP8266 platform abstraction (FreeRTOS spinlock vs. noInterrupts)
  • Configurable pulse buffer size via GEIGER_PULSE_BUFFER_SIZE preprocessor define

GeigerTubes

  • Sensitivity database for 6 common GM tubes (J305, M4011/J321, HH614, SBM-20, SI-3BG, LND7317)
  • 14 radiation source presets (Cs-137, Co-60, Am-241, background, and more)
  • Data from Rad Lab numerical simulations via RadPro v3.1.1
  • Compact uint8_t encoding (logarithmic scale) — 84 bytes vs. 336 bytes for raw floats

Requirements

Platform Tested
ESP8266 (NodeMCU, Wemos D1)
ESP32
Other Arduino (AVR, SAMD) Not tested — may require platform adaptation

Arduino IDE 1.8+ or PlatformIO.


Installation

Arduino IDE

Install via Library Manager:

Search for GeigerMeasurement in the Library Manager (Sketch → Include Library → Manage Libraries).

Install manually:

  1. Download this repository as a ZIP file
  2. In Arduino IDE: Sketch → Include Library → Add .ZIP Library
  3. Optionally install RunningStatistics the same way for long-term averaging support

PlatformIO

Install from PlatformIO registry:

Add to your platformio.ini:

lib_deps =
    soosp/GeigerMeasurement
    soosp/RunningStatistics ; optional, for long-term data processing

Install directly from GitHub:

lib_deps =
    https://github.com/soosp/GeigerMeasurement
    https://github.com/soosp/RunningStatistics

Manual

Copy src/GeigerMeasurement.h and src/GeigerTubes.h into your sketch folder. For long-term data processing, also copy RollingStats.h, CumulativeStats.h and ExponentialAverage.h from the RunningStatistics library.


Quick Start

#include <GeigerMeasurement.h>
#include <RollingStats.h>

// SBM-20 tube, background radiation, 60-second sliding window
GeigerMeasurement geiger(TUBE_SBM20, SOURCE_BACKGROUND);

// 128 bins × 60s = up to 2 hours of history
RollingStats<128, 60> stats;

void IRAM_ATTR geigerISR() { geiger.onPulse(); }

void setup() {
    Serial.begin(115200);
    pinMode(D7, INPUT);
    attachInterrupt(digitalPinToInterrupt(D7), geigerISR, FALLING);
}

void loop() {
    static uint32_t last = 0;
    if (millis() - last < 1000) return;
    last = millis();

    GeigerReading r = geiger.getReading();
    if (!r.valid) return;

    Serial.printf("CPM: %.1f  µSv/h: %.4f  ±%.0f%%\n",
                  r.cpm, r.uSvH, r.confidenceHalf);

    stats.addSample(r.cpm, r.timestampMs);

    if (stats.hasWindow(3600))
        Serial.printf("1h avg: %.1f CPM\n", stats.average(3600));
}

Wiring

The example above expects an active-low pulse signal on the GPIO pin (the tube module pulls the line low on each detected particle).

The wiring depends largely on the specific circuit design. Check the schematic diagram for details.

For example, the INT output of the popular educational GM circuit from China, named RadiationD-v1.1 CAJOE can be connected directly to a digital input of a NodeMCU, since it already includes the necessary pull-up and protective resistors.

NodeMCU note: INPUT_PULLUP is not working with this circuit solution. Use INPUT as shown in the example above.


API Reference

GeigerTubes.h

Provides tube sensitivity constants and source correction data derived from RadPro v3.1.1 (Rad Lab simulations).

GeigerTube enum

Selects the GM tube type. Determines the Rad Lab baseline sensitivity used for dose-rate calculation and the field factor baseline.

enum GeigerTube {
    TUBE_J305,      // J305 107 mm — cylindrical glass, common in budget modules
    TUBE_M4011,     // M4011 / J321 — widely used in cheap commercial counters
    TUBE_HH614,     // HH614 — cylindrical glass, also common in budget modules
    TUBE_SBM20,     // SBM-20 — Soviet-era surplus, popular in the DIY community
    TUBE_SI3BG,     // SI-3BG — low sensitivity, suited for high-radiation detection
    TUBE_LND7317,   // LND 7317 — cylindrical, halogen-quenched
    TUBE_CUSTOM,    // Custom tube — no Rad Lab data; use setSensitivity() directly
    TUBE_J321 = TUBE_M4011  // Alias — J321 shares identical Rad Lab data with M4011
};
Enum Tube Cs-137 sensitivity (Rad Lab) BG sensitivity (Rad Lab) Empirical fieldFactor for BG
TUBE_J305 J305 107 mm 135.2 CPM/(µSv/h) 180.5 CPM/(µSv/h) 1.069
TUBE_M4011 / TUBE_J321 M4011 / J321 108.3 CPM/(µSv/h) 144.6 CPM/(µSv/h) 1.269
TUBE_HH614 HH614 30.2 CPM/(µSv/h) 40.2 CPM/(µSv/h) — (not measured)
TUBE_SBM20 SBM-20 106.1 CPM/(µSv/h) 106.1 CPM/(µSv/h) 1.611
TUBE_SI3BG SI-3BG 3.3 CPM/(µSv/h) 3.6 CPM/(µSv/h) — (not measured)
TUBE_LND7317 LND 7317 252.6 CPM/(µSv/h) 289.0 CPM/(µSv/h) — (not measured)
TUBE_CUSTOM any use setSensitivity()

Cs-137 and background (BG) sensitivities are Rad Lab simulation values. Empirical field factors are measured — see Empirical Field Factors for details.

TUBE_CUSTOM has no Rad Lab baseline: tubeSourceSensitivity() returns NaN, and setFieldFactor() has no effect. Use setSensitivity() to set the sensitivity directly.


GeigerSource enum

Selects the radiation source preset, which applies a correction factor to the base tube sensitivity. Use SOURCE_BACKGROUND for ambient monitoring.

enum GeigerSource {
    SOURCE_CS137,            // Cs-137          — 661 keV γ, standard calibration source
    SOURCE_CO60,             // Co-60           — 1.17 + 1.33 MeV γ
    SOURCE_TC99M,            // Tc-99m          — 140 keV γ, nuclear medicine
    SOURCE_I131,             // I-131           — 364 keV γ, medicine / fallout
    SOURCE_LU177,            // Lu-177          — 113 + 208 keV γ, nuclear medicine
    SOURCE_AM241,            // Am-241          — 59 keV γ, ionization smoke detectors
    SOURCE_RADIUM,           // Radium          — Ra-226 decay chain (mixed energies)
    SOURCE_URANIUM_ORE,      // Uranium ore     — U-238 decay chain
    SOURCE_URANIUM_GLASS,    // Uranium glass   — low-activity U-238
    SOURCE_DEPLETED_URANIUM, // Depleted uranium — U-238 dominant
    SOURCE_THORIUM_ORE,      // Thorium ore     — Th-232 decay chain
    SOURCE_XRAYS,            // X-rays at ~60 kV — diagnostic imaging
    SOURCE_K40,              // K-40            — 1.46 MeV γ, potassium in food/soil
    SOURCE_BACKGROUND        // Natural background — recommended default
};

The source correction factor is dimensionless and relative to the Cs-137 baseline (factor = 1.0 for SOURCE_CS137). It accounts for the tube's energy-dependent efficiency. Factors are from Rad Lab simulations.


tubeSourceSensitivity()

float tubeSourceSensitivity(GeigerTube tube, GeigerSource source);

Returns the sensitivity in CPM/(µSv/h) for a tube/source combination. This is the value used internally by GeigerMeasurement and can be passed to setSensitivity(). Returns NaN for TUBE_CUSTOM or out-of-range input.


tubeSensitivity()

float tubeSensitivity(GeigerTube tube);

Returns the base Cs-137 sensitivity. Equivalent to tubeSourceSensitivity(tube, SOURCE_CS137). Returns NaN for TUBE_CUSTOM.


tubeSourceFactor()

float tubeSourceFactor(GeigerTube tube, GeigerSource source);

Returns the dimensionless source correction factor only, not multiplied by base sensitivity. 1.0 = Cs-137 baseline. Useful for diagnostics.


tubeLabel()

const char* tubeLabel(GeigerTube tube);

Returns a short human-readable name for a tube type. Suitable for Serial output, display labels, and CSV headers. Returns "Custom" for TUBE_CUSTOM or any out-of-range value.

Serial.printf("Tube: %s\n", tubeLabel(TUBE_SBM20));   // "SBM-20"
Serial.printf("Tube: %s\n", tubeLabel(TUBE_M4011));   // "M4011/J321"
Serial.printf("Tube: %s\n", tubeLabel(TUBE_J321));    // "M4011/J321"
Serial.printf("Tube: %s\n", tubeLabel(TUBE_CUSTOM));  // "Custom"

GeigerMeasurement.h

GEIGER_PULSE_BUFFER_SIZE

Preprocessor define to set the pulse ring buffer size. Must be a power of 2. Default: 256 (1 KB RAM).

#define GEIGER_PULSE_BUFFER_SIZE 512
#include <GeigerMeasurement.h>

Or in platformio.ini:

build_flags = -DGEIGER_PULSE_BUFFER_SIZE=512

The buffer holds timestamps of recent pulses. Larger buffers allow longer adaptive windows, at the cost of RAM.

Default constants (GeigerConfig namespace)

All tunable parameters have named compile-time defaults in the GeigerConfig namespace. These are used internally and can be referenced in application code:

Constant Default Description
DEFAULT_ADAPTIVE_PULSES 19.0 Target pulse count for adaptive modes
DEFAULT_ADAPTIVE_MAX_WINDOW 5.0 ADAPTIVE_FAST window cap [s]
DEFAULT_ADAPTIVE_MIN_WINDOW 5.0 ADAPTIVE_PRECISION window floor [s]
DEFAULT_DEAD_TIME_MAX_FACTOR 10.0 Dead-time compensation cap
DEFAULT_DEAD_TIME_WARN_RATIO 0.8 Saturation warning threshold (fraction of cap)
DEFAULT_CALIBRATE_MAX_CONFIDENCE 20.0 Default max CI for calibrate() [%]

AveragingMode enum

Selects how the CPM rate is averaged over time.

enum class AveragingMode {
    ADAPTIVE_FAST,       // last 19 pulses, max 5s window — fastest response, good for startup
    ADAPTIVE_PRECISION,  // last 19 pulses, min 5s window — balanced accuracy
    FIXED_10S,           // 10-second sliding window      — responsive monitoring
    FIXED_30S,           // 30-second sliding window      — general use
    FIXED_60S            // 60-second sliding window      — background monitoring
};
Mode Window Best for
ADAPTIVE_FAST Last 19 pulses, max 5s Fast response, startup
ADAPTIVE_PRECISION Last 19 pulses, min 5s Balanced accuracy
FIXED_10S 10 seconds Responsive monitoring
FIXED_30S 30 seconds General use
FIXED_60S 60 seconds Background monitoring

A common pattern is to start in ADAPTIVE_FAST and switch to FIXED_60S after 60 seconds — see GeigerBackground.ino for a complete example.


GeigerReading struct

Returned by getReading(). Always check valid before using numerical fields.

struct GeigerReading {
    float    cpm;                   // dead-time compensated rate [CPM]
    float    uSvH;                  // dose rate [µSv/h]
    float    confidenceHalf;        // 95% CI half-width [%]
    uint32_t pulseCount;            // raw pulses in averaging window
    uint32_t compensatedPulseCount; // dead-time compensated pulse count (integer carry)
    float    windowSec;             // actual window duration [s]
    uint32_t timestampMs;           // measurement time [ms] — use for addSample()
    bool     valid;                 // false during startup / insufficient data
    bool     saturated;             // dead-time factor > 80% of cap
    bool     tubeAlive;             // false if no pulse within timeout
    bool     counterSaturated;      // totalPulses reached UINT32_MAX
};

Field notes:

  • confidenceHalf: 95% Poisson CI using the RadPro formula. Value of 269.0 means N≤1 (maximum uncertainty). Smaller is better.
  • compensatedPulseCount: more accurate than cpm * windowSec / 60 for long-term dose integration because fractional remainders are carried forward across calls.
  • timestampMs: pass this to RollingStats::addSample() — do not use millis() directly, which has processing skew.
  • uSvH: NaN if TUBE_CUSTOM and setSensitivity() has not been called.
  • tubeAlive: checked via overflow-safe unsigned subtraction (millis() - lastPulseMs < timeout), correct across the ~49-day millis() rollover.

Constructor

explicit GeigerMeasurement(
    GeigerTube    tube       = TUBE_SBM20,
    GeigerSource  source     = SOURCE_BACKGROUND,
    AveragingMode mode       = AveragingMode::FIXED_60S,
    float         deadTimeUs = 0.0f
);
Parameter Default Description
tube TUBE_SBM20 GM tube type
source SOURCE_BACKGROUND Radiation source preset
mode FIXED_60S Averaging mode
deadTimeUs 0.0 Dead time [µs]. 0 = disabled.

Getters — tube, source, mode

GeigerTube    getTube()   const;  // current tube type
GeigerSource  getSource() const;  // current source preset
AveragingMode getMode()   const;  // current averaging mode

onPulse()

void IRAM_ATTR onPulse();

Records a pulse. Must be called from the GPIO interrupt handler:

void IRAM_ATTR geigerISR() { geiger.onPulse(); }
attachInterrupt(digitalPinToInterrupt(PIN), geigerISR, FALLING);

On ESP32, onPulse() acquires the same FreeRTOS spinlock used by getReading(), ensuring mutual exclusion across both cores. See Thread Safety for details.


getReading()

GeigerReading getReading();

Computes and returns the current measurement. Call from loop(), typically once per second. Thread-safe — uses a critical section to snapshot ISR data before computing.


Pulse counters

uint32_t totalPulses();     // pulses since last reset() — saturates at UINT32_MAX
uint32_t lifetimePulses();  // pulses since power-on — never reset, saturates at UINT32_MAX

Both are read under a critical section for consistency on ESP32.

Note on saturation: For an SBM-20 tube, the saturation time of the uint32_t counters is 408 years under continuous operation with background radiation (20 CPM), but even with continuous extreme high radiation levels (10,000 CPM), it is 298 days.


reset()

void reset();

Clears the pulse buffer and totalPulses. lifetimePulses is preserved.


Tube and source setters

void setTube(GeigerTube tube);
void setSource(GeigerSource source);
void setTubeAndSource(GeigerTube tube, GeigerSource source);
void setSensitivity(float cpmPerUsvH);  // direct override

setTube(), setSource(), and setTubeAndSource() recompute sensitivity automatically from GeigerTubes.h data and reset fieldFactor to 1.0. No reset() required.

setSensitivity() sets the sensitivity directly — also updates _fieldFactor for consistency. For TUBE_CUSTOM only this calibration method can be used.


setMode()

void setMode(AveragingMode mode);

Changes the averaging mode. Clears the pulse buffer. totalPulses and lifetimePulses are preserved.


Dead-time setters

void  setDeadTime(float us);           // dead time [µs], 0 = disabled
void  setDeadTimeMaxFactor(float f);   // compensation cap (default: 10.0)
void  setDeadTimeWarnRatio(float r);   // saturated threshold (default: 0.8)
float getDeadTime()          const;    // current dead time [µs]
float getDeadTimeMaxFactor() const;    // current compensation cap
float getDeadTimeWarnRatio() const;    // current saturation threshold

Typical dead time for common GM tubes is 50–200 µs. The compensation cap prevents runaway correction at extreme count rates. saturated in GeigerReading is set when the compensation factor exceeds warnRatio × maxFactor.

As a starting point, here are some commonly used dead-time values found online:

Tube Value (µs) Notes
SBM-20 190 according to the datasheet
J305 ~90 The conventional value is 50-100 µs for both 90 and 107 mm variant.
M4011/J321 ~90 The conventional value is 50-100us.
SI3BG 200 Value recommended by testers on varios online forums.
HH614 15 According to the datasheet.
LND7317 40 According to the datasheet, but based on measurements elsewhere, up to 270µs is recommended.
// Option 1: set a known value
geiger.setDeadTime(150.0f);

// Option 2: estimate from the pulse buffer (upper bound)
// Collect at least 500 pulses first for a useful estimate
float dt = geiger.measureDeadTime();
if (dt > 0) geiger.setDeadTime(dt);

Adaptive window setters

These clear the pulse buffer when called.

void  setAdaptivePulses(float n);        // target pulse count (default: 19)
void  setAdaptiveMaxWindow(float s);     // ADAPTIVE_FAST cap [s] (default: 5)
void  setAdaptiveMinWindow(float s);     // ADAPTIVE_PRECISION floor [s] (default: 5)

Adaptive window getters

float getAdaptivePulses()     const;     // current target pulse count
float getAdaptiveMaxWindow()  const;     // current ADAPTIVE_FAST cap [s]
float getAdaptiveMinWindow()  const;     // current ADAPTIVE_PRECISION floor [s]

Tube-alive timeout

void     setTubeTimeoutMs(uint32_t ms);  // 0 = auto (RadPro formula, default)
uint32_t getTubeTimeoutMs() const;       // current override [ms], 0 = auto

When set to 0 (default), the timeout is computed from tube sensitivity:

timeout = 12 000 000 / sensitivity_CPM_per_uSvH [ms] + 1000 ms

Less sensitive tubes need a longer timeout because background pulses are rarer. For TUBE_CUSTOM before setSensitivity() is called, a 2-minute fallback is used.


Dead-time measurement

Two complementary methods are available. Both return an upper bound on τ — the true dead time may be shorter, but no interval shorter than τ was observed.

getMeasuredDeadTime()
float getMeasuredDeadTime() const;

Returns the minimum inter-pulse interval observed across all pulses since the last tube change in µs — a continuously improving lifetime estimate. The value is updated by every pulse in onPulse() and is never reset by reset(). It is reset only by setTube() and setTubeAndSource(), since dead time is a physical property of the tube.

Prefer this method for normal use. It never forgets a previously observed minimum and requires no explicit trigger.

// Apply the lifetime estimate whenever it improves:
float dt = geiger.getMeasuredDeadTime();
if (dt > 0) geiger.setDeadTime(dt);
measureDeadTime()
float measureDeadTime(uint32_t sampleCount = GEIGER_PULSE_BUFFER_SIZE);

Scans a recent window of the pulse buffer and returns the minimum interval found. Useful when a time-limited window is needed:

  • Diagnosing whether tube behaviour has changed recently (ageing, HV drift)
  • Measuring dead time only during a high-rate event
  • Limiting the scan to a specific number of recent pulses via sampleCount
Parameter Description
sampleCount How many recent pulses to scan (default: full buffer)
returns Estimated dead time [µs], or 0.0 if fewer than 2 pulses in window

For an SBM-20 at background levels (~20 CPM), 500 pulses take ~25 minutes. A check source speeds this up considerably.

// Snapshot the current window after enough pulses:
if (!measured && geiger.lifetimePulses() >= 500) {
    float dt = geiger.measureDeadTime();
    if (dt > 0) geiger.setDeadTime(dt);
    measured = true;
}

Calibration and field factor

The field factor corrects for the difference between Rad Lab simulation values and measured real-world tube sensitivity. It is tube- and source-specific: automatically reset to 1.0 by setTube(), setSource(), and setTubeAndSource().

bool  calibrate(float knownUsvH, float maxConfidencePct = 20.0f);
void  setFieldFactor(float factor);   // direct: sensitivity = radlab × factor
float getFieldFactor() const;         // read back current factor
float getRadlabSensitivity() const;   // Rad Lab baseline (before correction)
void  resetFieldFactor();             // restore to 1.0 (Rad Lab default)

calibrate() computes fieldFactor = (cpm / knownUsvH) / radlabSensitivity and updates both _fieldFactor and _sensitivity atomically.

Returns false if the reading is not yet valid, knownUsvH ≤ 0, confidence exceeds maxConfidencePct, or TUBE_CUSTOM is in use (setFieldFactor() also has no effect for TUBE_CUSTOM).

// Live calibration against a known reference:
while (!geiger.calibrate(0.116f, 15.0f)) { delay(1000); }
float ff = geiger.getFieldFactor();  // e.g. 1.269 — save to EEPROM/flash

// Restore on next boot without re-calibrating:
geiger.setFieldFactor(1.269f);

// Empirical values from parallel background measurements (Pannonhalma, 2026):
// TUBE_M4011:  setFieldFactor(1.269f)
// TUBE_SBM20:  setFieldFactor(1.611f)
// TUBE_J305:   setFieldFactor(1.069f)  // 107 mm variant
// See Empirical Field Factors section for measurement details.

// Custom tube (no Rad Lab baseline):
GeigerMeasurement geiger(TUBE_CUSTOM, SOURCE_BACKGROUND);
geiger.setSensitivity(130.1f);  // J305 90 mm: empirical value from 43.5h measurement

Thread Safety and Interrupt Latency

How shared state is protected

onPulse() runs in interrupt context and writes to several shared variables (_timestamps, _head, _count, _totalPulses, _lifetimePulses). getReading() reads all of them from the main loop. Without synchronisation, a pulse arriving mid-read could produce an inconsistent snapshot — for example, _head already incremented but the corresponding timestamp not yet written.

The library uses two complementary critical-section macro pairs:

Macro Used in ESP32 ESP8266
GEIGER_ENTER/EXIT_CRITICAL() task context (getReading, reset, …) taskENTER/EXIT_CRITICAL(&mux) noInterrupts() / interrupts()
GEIGER_ENTER/EXIT_CRITICAL_ISR() ISR (onPulse) taskENTER/EXIT_CRITICAL_ISR(&mux) no-op

Both pairs share the same portMUX spinlock on ESP32, so they mutually exclude each other across both cores. On ESP8266, the single-core architecture means an executing ISR already excludes the main loop — no additional locking is needed inside onPulse().

Can a pulse be lost during the critical section?

Yes — but the probability is negligible at background levels.

getReading() acquires several short critical sections in sequence: the main buffer snapshot (256 × uint32_t), plus smaller snapshots in _windowDurationUs(), _applyDeadTime(), _applyDeadTimeCarry() and _tubeAliveCheck(). Each individual section is brief (~0.2–1.3 µs), and their combined ISR-blocking time per getReading() call is approximately 2 µs.

At background radiation levels, pulses arrive on average every 3 seconds for an SBM-20. The probability of a pulse coinciding with any critical section is:

2 µs / 3 000 000 µs ≈ 0.00007%

At background levels this corresponds to roughly one missed pulse every 55 years of continuous operation. Even at 10 000 CPM (extreme radiation), the overlap probability is only ~3% — comparable in magnitude to the tube's own dead-time loss, which the dead-time compensation already corrects for.

A pulse lost during the critical section is simply absent from the ring buffer, causing a slight underestimate of CPM — identical in character to dead-time loss. The simpler and safer design (full buffer copy under lock) is therefore preferred over a lock-free ring buffer, which would be significantly more complex for no meaningful practical gain.


Empirical Field Factors

The Rad Lab simulation values are theoretical (Cs-137 reference geometry). In practice, the tested real GM tubes measuring background radiation show deviations — typically 1.0–1.6× — due to wall material, energy response, and geometry differences vs. the simulation model.

Use setFieldFactor() to apply these corrections, or calibrate() to determine the value for your specific tube and location.

Measured values

Measurement setup: four custom boards built under the principles of CAJOE module, with NodeMCU, equipped with different GM tubes, running in parallel — BOSEAN FS-5000 as reference

Conditions:

  • Location: Pannonhalma, Hungary — indoor, ~1 m above floor
  • Hardware: four custom boards built under the principles of CAJOE module, with NodeMCU, equipped with different GM tubes, ~380–400V HV, running in parallel
  • Source setting: SOURCE_BACKGROUND
  • Duration: ~43.5 hours parallel run (lifetime average, Poisson error ~1%)
  • Reference: BOSEAN FS-5000 with J321 tube, RadPro 3.1.1 (factory calibration) displaying ~16.6 CPM → 0.115 µSv/h, consistent with regional OMSZ/HM monitoring network data (~0.087–0.115 µSv/h in the Győr–Pápa region)

Results:

Tube CPM (measured) fieldFactor Notes
M4011 21.11 1.269
SBM-20 19.66 1.611 Largest deviation — steel wall energy filtering(?)
J305 107 mm 22.19 1.069 Close to Rad Lab prediction
J305 90 mm 14.96 Use setSensitivity(130.1f) with TUBE_CUSTOM

Key ratios vs. Rad Lab predictions:

Ratio Measured Rad Lab Deviation
M4011 / SBM-20 1.074 1.363 −21.2%
J305 107 mm / SBM-20 1.129 1.701 −33.6%
J305 107 mm / M4011 1.051 1.248 −15.8%

These are single-location measurements on individual tube samples. Your values may differ due to tube-to-tube manufacturing tolerances, HV supply accuracy, and local radiation field composition.


Example Sketches

Sketch Description
GeigerMinimal.ino Minimal single-file example
GeigerMeasurement.ino Basic example displays CPM, µSv/h, EMA values, confidence, and window
GeigerMonitor.ino Full-featured monitor with RunningStatistics
GeigerBackground.ino Background monitoring with adaptive→fixed mode transition
GeigerCompare.ino Parallel tube sensitivity comparison — CSV output, lifetime average, accumulated dose
GeigerPulseIndicator.ino Demonstrates per-pulse LED flash and passive buzzer click indicators

Optional: RunningStatistics

For long-term averaging (30-minute, 1-hour, 24-hour windows), install the companion library: RunningStatistics

#include <GeigerMeasurement.h>
#include <RollingStats.h>  // from soosp/RunningStatistics

RollingStats<128, 60> stats;  // 128 bins × 60s = 2 hours

// In loop():
stats.addSample(r.cpm, r.timestampMs);

if (stats.hasWindow(3600))
    Serial.printf("1h avg: %.1f CPM\n", stats.average(3600));

See GeigerMonitor.ino and GeigerBackground.ino in examples for more complete examples.


Development note

This library was designed and developed by Péter Soós in collaboration with Claude (Anthropic AI assistant). The iterative design process — including algorithm selection, architecture decisions, code review, and documentation — was conducted through an extended conversation with Claude.


Acknowledgements

Core algorithms adapted from RadPro v3.1.1 (MIT License). Sensitivity data originally computed by Rad Lab, a separate simulation tool by the same author. Special thanks to Gissio for both projects.


License

MIT License — see LICENSE for details.

About

An Arduino library for Geiger-Müller counters on ESP32 and ESP8266, implementing the core measurement algorithms of the RadPro project by Gissio, adapted for the Arduino platform.