Notifications
You must be signed in to change notification settings - Fork 2
Mass spectra
PAMPA is designed to handle MS mass spectra, such as MALDI-TOF or MALDI-FT ICR spectra. Note that spectra must undergo mandatory deisotoping and calibration steps before being processed by PAMPA.
The program processes a batch of mass spectra simultaneously. All mass spectra files should be placed in a same folder, and the folder path is provided as parameter to -s. There are three recognized formats: CSV, MGF and mzML.
CSV format : It consists of two columns. The first column is designated for mass (m/z), and the second column records intensity (I). Columns are separated by either a comma (',') or a semicolon (';'). Each file contains one spectrum. The initial row serves as the header. Name of the file : .csv or .CSV
MGF format : This is the Mascot Generic Format. Name of the file : .mgf or .MGF
mzML format : see https://www.psidev.info/mzML. With this format, a single file can contain several spectra. Name of the file: .mzML
Any other files present in the folder will be disregarded.
The error margin is related to the resolution of the mass spectrometer and its ability to distinguish closely spaced peaks. This is the maximal deviation between a peak and the theoretical mass of the peptide marker. This option can be expressed in Dalton or in ppm (parts per million).
- If the value is smaller than 1, it is assumed to be in Da (Daltons). In this case, recommended values are 0.1 for MALDI-TOF, and 0.01 for MALDI-FTICR.
- If the value is larger than 1, it is assumed to be in ppm (parts per million). In this case, recommended values are 50 for MALDI-TOF, and 5 for MALDI-FTICR.