Skip to content

Nimrod Jaws Manager

Tom Willemsen edited this page Aug 28, 2019 · 6 revisions

Wiki > The Backend System > Specific Device IOC > Jaws and Slits > Nimrod Jaws

Nimrod jaws are a set of 6 jaws which work in concert to limit the beam of neutrons reaching the sample (see Jaws Manager for more). The 6 jaws are ordered 1 nearest the beam and 6 nearest the sample.

The parameters, with reference to the diagram in Jaws Manager, are:

Letter Parameter Value Notes
c Collimator position 0 mm No collimator so beam tapered from moderator
j1 Jaw 1 position 6458 mm Distance from moderator
j2 Jaw 2 position 8072 mm Distance from moderator
j3 Jaw 3 position 12072 mm Distance from moderator
j4 Jaw 4 position 15450 mm Distance from moderator
j5 Jaw 5 position 17235 mm Distance from moderator
j6 Jaw 6 position 19691 mm Distance from moderator
s Sample position 19691 mm Note that the sample is assumed to be at the final jawset
w Width of moderator user set
h Height of moderator user set
x Width at sample user set
y Height at sample user set

Setup

The control files for the nimrod jaws are part of the galil motor set up and are per instrument. There is an example in ...\EPICS\support\motorExtensions\master\settings\nimrod_jaws\*.cmd which would need to be copied to C:\Instrument\Settings\config\<instrument host name>\configurations\galil. IOC System tests also exist for the logic.

Jaw centres

The jaws on NIMROD are not centred with respect to the beam, i.e. all jaws being at a centre of (0, 0) will not actually let the beam through. This is due to some of the jaw sets being mounted off-centre during installation.

The expected centres of each jaw set are:

Jaw set Horizontal Vertical
1 -0.135 -1.5
2 -0.126 -1.457
3 -0.205 -9.634
4 0.159 -1.460
5 -0.303 -0.801
6 -0.378 -9.507

The numbers above correspond to the jaws being centred around the beam in "real" coordinates as opposed to motor coordinates.

Clone this wiki locally