Skip to content

Bode Analyzer

Val edited this page Nov 21, 2026 · 7 revisions

The Bode analyzer measures the frequency response of a circuit on the device: gain and phase against frequency, without a computer. It drives the circuit with the built-in generator and compares CH2 (the circuit output) with CH1 (its input). ZT703S only.

Bode plot: gain in dB above, phase in degrees below, 100 Hz to 100 kHz

Connecting

  1. Connect the generator output to the circuit input and to CH1.
  2. Connect the circuit output to CH2.
  3. On the oscilloscope, in RUN, enable both channels.
  4. Open Analyze → Bode (FRA).

CH1 is the live reference, so any change of the generator level with frequency divides out.

Setup

Row
Start, Stop The frequency range, from 1 Hz. When the sweep starts, the top is limited by the DAC rate and the drive level
Points/decade 5, 10, 20 or 50 points per decade, log-spaced
Settle Wait after each frequency change before measuring: 0 to 2000 ms. Give slow circuits more time
Averages 1, 3 or 10 frames per point
Drive The level: the generator's own level (GEN), Dedicated, Low, High or Custom
Custom low code, Custom high code For Custom: DAC codes 0–4095, low below high. ←/→ step by 1, ↑/↓ by 100
Run sweep Starts the sweep
Show plot Back to the plot
Results Peak, both −3 dB points, bandwidth, gain and phase margins
Save CSV Saves the sweep

At low frequencies one capture can take several seconds.

Keys

Key Action
F1 Start or stop the sweep
F2 Setup; Back returns to the result
F3 Switch between the plot and the table; from Results, back to the plot
F4 Close the Bode analyzer
MENU Switch the active cursor, A or B
← / → Move the active cursor from point to point
↑ / ↓ Jump between markers on the plot; page through the table

The top line shows the frequency, gain and phase at the cursor. The Δ B−A line shows the difference between the cursors. Missing values show as --.

Table and results

  • The table lists each point with its signal level and the input range of each channel.
  • part avg marks a point with fewer good frames than Averages asked for.
  • low CH2 marks a point where CH2 was too small and the gain is an estimate.
  • Results shows the peak, both −3 dB points relative to the peak, the bandwidth between them, and the gain and phase margins with their crossover frequencies. Crossings are interpolated on a log frequency scale between good points. The gain margin uses the −180° phase crossing, the phase margin the 0 dB gain crossing; with several crossings the smallest margin is shown.

CSV files

Setup → Save CSV writes a finished or stopped sweep to /openosc/FRA/FRAnnnnn.CSV (Files on the Device). The file starts with # key=value lines that describe the sweep, including the drive level and averaging, followed by one row per point: index, frequency, gain, phase, signal level, input ranges and quality flags.

During a sweep

  • The capture uses 64K points, 32K per channel.
  • The generator and capture commands over SCPI wait until the sweep ends.
  • When the sweep ends, or you stop it or leave, the capture depth and the generator settings come back as they were.

Clone this wiki locally