Skip to content

Performance Profiles

Val edited this page Nov 20, 2026 · 5 revisions

Settings → Device → Performance → Run mode selects the core clock and the oscilloscope sample rate. The switch is instant; no restart is needed.

Profile Core clock Sample rate per channel USB data
200/100 (default) 200 MHz 100 MS/s Yes
280/140 280 MHz 140 MS/s No
360/90 360 MHz 90 MS/s No
  • USB (SCPI, Remote mode) works only in 200/100.
  • 360/90 runs the microcontroller beyond its rated clock. It is offered for experiments; use it at your own risk.
  • If a start at 360/90 is not confirmed, the next start uses 200/100 automatically and shows a message.
  • While the generator streams from the DAC, the maximum sample rate is halved.

CH1 interleave

Settings → Device → Performance → Interleave CH1 samples CH1 with both ADCs in turn and doubles its sample rate.

Model Choices Default
ZT703S (experimental) Off, On: CH1 alone Off
ZT706S Off, On: CH1 alone, Auto: factory rules Auto

The manufacturer's firmware for the ZT703S does not use interleave; OpenOSC adds it as a new capability of this model. On the ZT706S, Auto keeps the built-in rules: interleave at 280 MHz and 10 or 25 ns/div.

On works whenever CH1 is the only enabled channel, in every profile and on every timebase, Roll included. CH2 stays on until you switch it off. The fastest rates:

Profile Interleaved CH1
200/100 200 MS/s
280/140 280 MS/s
360/90 180 MS/s

Slower timebases use slower rates as usual. While the generator streams from the DAC, the rate is halved.

The row shows what is in force:

Row value Meaning
On, active Interleave runs, with the measured offset of this profile
On, active uncal. Interleave runs; the offset of this profile has not been measured yet
On, waits: CH2 on A condition is missing: CH2 on, trial or run mode (in Auto also 10/25 ns, Roll or GEN DAC)

Corrections

  • DC offset. The two ADCs read a few codes apart, which shows as a fine comb on the trace. On the ZT703S, measure it once in each profile you use: Calibration → Interleave offset (ZT703S). The trace applies it in 1/256 code steps.
  • Gain and timing (ZT703S). With a profile for the unit, a short filter matches the second ADC to the first and lowers the spur at Fs/2 − f. Without a profile, gain and timing differences remain, and fast edges and high frequencies can still show a comb. Measure it with Calibration → Interleave gain & timing (ZT703S).
  • Clipping. When a capture clips (at least 32 samples in eight 128-point windows), the next capture uses the next wider input range while the V/div grid stays the same. The trigger level follows. The selected range returns when you change V/div, the position or the coupling, or leave interleave. Roll and the offset measurement keep the selected range.
  • Trigger. Near the trigger event, the trigger prefers a crossing that both ADCs confirm, which reduces jumps to a ringing edge. Short pulses still trigger as before.

Changing the profile keeps the choice and restarts the capture; the offset of the new profile applies. A held capture keeps the rate and the corrections it was taken with.

Clone this wiki locally