Repository navigation
Performance Profiles
Settings → Device → Performance → Run mode selects the core clock and the oscilloscope sample rate. The switch is instant; no restart is needed.
| Profile | Core clock | Sample rate per channel | USB data |
|---|---|---|---|
| 200/100 (default) | 200 MHz | 100 MS/s | Yes |
| 280/140 | 280 MHz | 140 MS/s | No |
| 360/90 | 360 MHz | 90 MS/s | No |
- USB (SCPI, Remote mode) works only in 200/100.
- 360/90 runs the microcontroller beyond its rated clock. It is offered for experiments; use it at your own risk.
- If a start at 360/90 is not confirmed, the next start uses 200/100 automatically and shows a message.
- While the generator streams from the DAC, the maximum sample rate is halved.
Settings → Device → Performance → Interleave CH1 samples CH1 with both ADCs in turn and doubles its sample rate.
| Model | Choices | Default |
|---|---|---|
| ZT703S (experimental) | Off, On: CH1 alone | Off |
| ZT706S | Off, On: CH1 alone, Auto: factory rules | Auto |
The manufacturer's firmware for the ZT703S does not use interleave; OpenOSC adds it as a new capability of this model. On the ZT706S, Auto keeps the built-in rules: interleave at 280 MHz and 10 or 25 ns/div.
On works whenever CH1 is the only enabled channel, in every profile and on every timebase, Roll included. CH2 stays on until you switch it off. The fastest rates:
| Profile | Interleaved CH1 |
|---|---|
| 200/100 | 200 MS/s |
| 280/140 | 280 MS/s |
| 360/90 | 180 MS/s |
Slower timebases use slower rates as usual. While the generator streams from the DAC, the rate is halved.
The row shows what is in force:
| Row value | Meaning |
|---|---|
| On, active | Interleave runs, with the measured offset of this profile |
| On, active uncal. | Interleave runs; the offset of this profile has not been measured yet |
| On, waits: CH2 on | A condition is missing: CH2 on, trial or run mode (in Auto also 10/25 ns, Roll or GEN DAC) |
- DC offset. The two ADCs read a few codes apart, which shows as a fine comb on the trace. On the ZT703S, measure it once in each profile you use: Calibration → Interleave offset (ZT703S). The trace applies it in 1/256 code steps.
-
Gain and timing (ZT703S). With a profile for the unit, a short filter matches the second ADC to the first and lowers the spur at
Fs/2 − f. Without a profile, gain and timing differences remain, and fast edges and high frequencies can still show a comb. Measure it with Calibration → Interleave gain & timing (ZT703S). - Clipping. When a capture clips (at least 32 samples in eight 128-point windows), the next capture uses the next wider input range while the V/div grid stays the same. The trigger level follows. The selected range returns when you change V/div, the position or the coupling, or leave interleave. Roll and the offset measurement keep the selected range.
- Trigger. Near the trigger event, the trigger prefers a crossing that both ADCs confirm, which reduces jumps to a ringing edge. Short pulses still trigger as before.
Changing the profile keeps the choice and restarts the capture; the offset of the new profile applies. A held capture keeps the rate and the corrections it was taken with.
OpenOSC is licensed under the GPL-3.0. Found a mistake on this page? Tell us in Discussions.