Calibration

Val edited this page Nov 22, 2026 · 8 revisions

Settings → Device → Scope calibration adds your own corrections on top of the device's built-in calibration data. The built-in data is only read. Your corrections are stored as files in /openosc/CALIBRATION/ and can be switched off at any time. The multimeter has its own calibration and is not affected.

Overview

Row What it does Prepare the inputs
Hardware zero (PWM) Adjusts the analog offset of both channels so that a grounded input sits exactly on zero, on every range and position Ground both inputs
Software zero Adds a small software correction to the zero of both channels Disconnect both inputs
Gain (sub-page) Corrects the scale of one range of one channel against a DC voltage you know Ground, then a DC reference on the probe tip
Gain → Match to other channel Matches the gain of one channel to the other on a common signal Both probes on one stable square wave
Interleave offset (ZT703S) Measures the DC offset between the two ADCs for CH1 interleave, per profile Ground CH1
Interleave gain & timing (ZT703S) Measures the gain and timing match of the two ADCs for CH1 interleave A clean sine from a generator on CH1
Interleave offset (sub-page, ZT706S) Matches the two interleaved ADC halves of CH1 A signal on CH1
Reload calibration JSON Re-reads all calibration files, for example after editing them on a computer —
Use factory zero Switches off both zero corrections; the measured values stay in the files —

Use Hardware zero or Software zero, one at a time. While Hardware zero is in use, Software zero is refused with Hardware zero in force: use factory zero first.

Before you start

  • Let the device warm up for a few minutes after switching on.
  • Every procedure asks for confirmation first, and the cursor starts on Cancel.
  • The generator is switched off and stays off afterwards.
  • While a procedure runs, the status bar shows CALIBRATING, Auto Setup, saving captures, USB Remote and the generator are blocked, and auto power off waits.
  • F4, MENU or MODE cancel at any time. The channel settings, timebase, trigger and RUN/HOLD state are restored. A held capture comes back as it was.
  • A result is saved only after the whole procedure has succeeded. A cancelled or failed run leaves the previous correction in force.

Hardware zero (PWM)

Each input has an analog offset, set by a PWM signal. Hardware zero moves this offset until a grounded input reads exactly the zero that the built-in calibration expects. This is the most complete zero correction and is the one to use in most cases.

  1. Connect both inputs to ground (probe tip to its ground clip, or a shorted BNC).
  2. Select Hardware zero (PWM). Confirm GROUND BOTH SCOPE INPUTS?
  3. Keep both inputs grounded while the progress panel runs. It shows the range, the percentage and the position being measured. A run usually takes about 15 s.

What it does:

  • Both channels together, DC coupling, ×1 probe, 1 ms/div.
  • All 11 ranges from 20 mV/div to 10 V/div, at 7 vertical positions (20, 40, 60, 80, 100, 120 and 140 of 160; 80 is the centre). That is 77 points per channel.
  • At each point it measures, corrects the offset and measures again, up to 12 times, until the zero is within a quarter of an ADC code. Each measurement waits 50 ms after the change and uses 32768 samples per channel.
  • The correction may move the zero by up to 40 ADC codes (about 1.5 divisions) from the built-in value. A larger error stops the run and nothing is saved.

Any DC voltage on an input is taken as an offset. If an input is not really grounded, the run either fails or stores a wrong zero. Run it again with the inputs grounded to fix it.

Running it again gives the same result; corrections do not add up. The result is stored in scope_pwm.json.

Software zero

Software zero leaves the analog offset as it is and corrects the zero in the measured data. It suits a small drift of the zero.

  1. Disconnect both inputs.
  2. Select Software zero. Confirm DISCONNECT SCOPE INPUTS?

It measures the same 77 points per channel with 262144 samples each and stores the difference between the measured and the expected zero. Between the 7 positions the correction is interpolated. The correction may be up to ±16 ADC codes; an unstable input, a larger error or noise on the input stops the run. The result is stored in scope_zero.json.

Gain

Gain corrects the scale (volts per division) of one range of one channel. You need a stable DC voltage whose value you know, for example a regulated 3.3 V or 5 V supply measured with a good multimeter.

The Gain page:

Row
Channel CH1 or CH2
V/div The range to correct
Reference at tip The voltage at the probe tip, with 1 mV resolution; any sign
Reference step The step of the arrows while editing the reference: 1 mV, 10 mV, 100 mV (default), 1 V or 10 V
Probe The probe ratio of the channel, taken from its channel panel (Oscilloscope)
Expected span How many ADC codes the reference gives at the built-in scale
Gain in force The correction now active for this range (1.000 = none)
Start measurement Starts the procedure
Factory gain Shown: the factory gain of CH1 and CH2 from the built-in calibration record
Match to other channel Matches the selected channel to the other one, see below
Use factory gain Switches the gain correction off; the measured values stay in the file

Expected span must be between 40 and 110 codes, roughly 1.5 to 4 divisions. Otherwise Start measurement is refused: choose another V/div or another reference.

  1. Set Channel, V/div and the probe on the channel, then enter Reference at tip.
  2. Start measurement. Confirm GROUND THE CHn PROBE TIP? and keep the tip grounded. OpenOSC measures the zero (65536 samples).
  3. The panel asks Apply +3.300 V to the CH1 tip (with your values). Connect the reference and select Reference applied. Without an answer the procedure ends after 300 s.
  4. OpenOSC measures the reference and checks that it is stable and within the ADC range.
  5. A review shows the measured span, the reference, the proposed gain and the gain in force. Select Save to keep it or Cancel to discard it.

The gain is the ratio between the voltage you entered and the voltage measured with the built-in scale, and must be between 0.8 and 1.25. It is measured from the difference between the reference and ground, so a zero offset does not affect it. Running it again replaces the value; nothing accumulates. The probe ratio is used once, so you can calibrate with a ×10 probe.

A new gain applies to new captures; a held capture keeps the scale it was taken with. Each of the 11 ranges of each channel has its own value, stored in scope_gain.json.

Matching the two channels

The two channels can read the same signal slightly differently because their gain differs. Hardware zero and Software zero correct only the zero, so they cannot align the channels: a zero shift changes Vrms, and Vpp stays the same. Match to other channel corrects the gain of one channel so that it reads the same as the other.

  1. Connect both probes to one stable square wave, for example the 1 kHz calibration output (Signal Generator). Set the correct probe ratio on both channels.
  2. Enable both channels, DC coupling and RUN, and set the same V/div on both. With ×10 probes on the calibration output, start with 1 V/div, 500 µs/div and 64K. The signal must span at least 60 ADC codes and stay inside the ADC range.
  3. Open Settings → Device → Scope calibration → Gain, set Channel to the channel to correct and select Match to other channel → Measure.
  4. The review shows:
    • Correction: the new gain of the channel;
    • Resolution: how much the ADC steps can affect the result;
    • Edge delta CH1/CH2: how far each channel deviates near the edges, a hint for probe compensation.
  5. With equal probe ratios, Swap probes asks you to swap the probes between the channels and measures again. The two measurements together remove the difference between the probes.
  6. Save stores the correction of the selected channel for the current range and applies it to new records.
  • The measurement uses your own RUN records; the calibration output and the generator keep running.
  • A signal that is still moving, too small, clipped or gives a correction outside 0.8–1.25 is refused with a hint. A change of the position, probe, V/div or the saved gain stops the measurement.
  • The correction is per range: repeat it on the ranges you use.
  • After Use factory gain, Save is refused when it would switch other saved ranges back on.
  • Match to other channel aligns the channels with each other. For an absolute gain, use Start measurement with a known DC voltage.

On a ZT703S with both ×10 probes on the calibration output, the difference between the channels dropped from 0.47 % to 0.09 %.

Interleave offset (ZT703S)

With the experimental Interleave CH1 option (Performance Profiles), CH1 is sampled by both ADCs in turn. The second ADC reads a few codes away from the first, which shows as a fine comb on the trace. This row measures that DC offset and stores it for the profile that is running.

  1. Switch on Settings → Device → Performance → Interleave CH1 and make sure it is in force: CH1 alone, Roll off, trigger mode Auto, RUN. 10 or 25 ns/div is recommended.
  2. Connect CH1 to ground.
  3. Select Interleave offset and confirm GROUND CH1?. Settings stay open on the row, which shows Measuring....
  4. After 4 captures a message shows the result, and the row shows it, for example +9.8 codes. Not measured means this profile has no value yet.
  5. Switch to the other profiles you use and repeat.
  • MENU on the row cancels the measurement. A change of V/div, coupling, position, timebase, channels or trigger, switching interleave off, a profile change or HOLD interrupts it.
  • Input not quiet (the input moves by more than 10 codes), Input at the ADC limit, an offset beyond ±16 codes or a timeout of 5 s stop the run.
  • A stopped, cancelled or interrupted run leaves the previous value in force.
  • A new value applies to new captures; a held capture keeps its offset.
  • Only the DC offset is corrected. Gain and timing differences between the ADCs remain.
  • On the test unit the offset was about +9.8 codes at 200/100, +7.1 at 280/140 and +6.6 at 360/90.

Interleave gain & timing (ZT703S)

The two ADCs also differ slightly in gain and sampling time, which shows as a spur at Fs/2 − f with interleave on. Interleave gain & timing measures this on the device and saves a profile: a 9-tap filter then brings the second ADC to the gain and timing of the first, for the trace, the measurements and the exports. This calibration is experimental.

You need a generator that gives a clean sine up to 66 MHz.

  1. Measure Interleave offset for the run mode you use.
  2. Switch on Interleave CH1, switch CH2 off, set the trigger to Auto and the timebase to 10 or 25 ns/div. Keep RUN and switch the OpenOSC generator off.
  3. Select Interleave gain & timing. The page shows the Capture rate and the Reference frequency to apply, for example 2 MHz.
  4. Feed a clean sine of that frequency into CH1, below clipping, and select Measure tone.
  5. Repeat for every frequency the page asks for. Six tones fit the filter (Progress Fit 1/6 … 6/6), and five more check it (Check 1/5 … 5/5). The frequencies lie between 2 and 66 MHz.
  6. The review shows Check tones and the Median improvement of the spur in dB. Select Save profile, then restart the device to apply it.
  • A noisy, clipped or wrong-frequency input is refused with a hint (Need a clean sine, Check generator frequency, Reduce sine level or change CH1 scale), and you can measure the tone again. A profile that fails the checks is not saved.
  • A change of the capture settings stops the calibration. Cancel leaves it.
  • The profile is stored in /openosc/CALIBRATION/scope_interleave_match.json and read at startup. It is bound to this unit and holds up to four profiles, each for one run mode, sample rate, input range (before the probe factor), coupling and zero position. All four profile slots are used means the file is full; remove the file to start again.
  • Without a matching profile, in Roll, during zero calibration and after an automatic range step, only the DC offset applies. Record edges and clipped stretches also keep the DC offset only.
  • ACQ:INTerleave:MATCh? returns 1 when a profile is in use. A held capture keeps the coefficients it was taken with.
  • To switch the correction off, remove the file and restart.

Interleave offset (ZT706S)

On the ZT706S, CH1 can be sampled by two ADCs in turn. A small offset between them shows as a fine comb on the trace. This page lets you set that offset by hand.

Row
Offset The offset in ADC codes, ±16, in 0.1 steps. When you finish editing with MENU, it applies to new captures at once as a preview
In force The value used now
Saved The value in the file
Save Writes the value in force to the file
Cancel Returns to the saved value
Reset to factory Returns to the built-in value

A preview that you leave without Save stays until Cancel, Reload calibration JSON or a restart. The value is stored in scope_interleave.json. This page has not been tested on a ZT706S yet.

Files

File Contents
scope_pwm.json Hardware zero: the PWM values per channel, range and position
scope_zero.json Software zero: the corrections per channel, range and position
scope_gain.json Gain: one value per channel and range
scope_interleave.json The interleave offset: ZT703S per profile, ZT706S one value
scope_interleave_match.json ZT703S: gain and timing profiles for interleave, from Interleave gain & timing
  • Each file is bound to this unit's built-in calibration data and applies only to this unit. A file from another device is renamed to *.bad.NNNN.json and the built-in values are used. At startup a notice lists such files.
  • When a file is saved, the previous version is kept as *.previous.json. A damaged file is set aside as *.bad.NNNN.json, and the previous version is used if it is valid.
  • Use factory zero and Use factory gain write "enabled": false and keep the values. A new run switches the correction on again.
  • After editing a file on a computer, use Reload calibration JSON.

The firmware recognizes Long, Compact and MKR factory calibration records. Compact records use the factory offset tables and the default gain for both channels. The firmware and Recovery included in 0.20.0 and later recognize Compact records with extra bytes after the offset-table CRC.

Recovery shows the built-in calibration data (VIEW CALIBRATION), read only. Update Recovery itself to refresh its calibration viewer.