Oscilloscope
Caution
The USB port is not isolated from the measurement inputs. Be careful when you measure while the device is connected to a computer: see USB and Remote Mode.
Two channels, 20 mV/div to 10 V/div, 10 ns/div to 20 s/div, with Auto, Normal and Single trigger, measurements, cursors, FFT, XY, MATH and Protocol Decoders.
| Key | Short press | Hold |
|---|---|---|
| F1 Scale | Select Scale. ↑/↓ change V/div, ←/→ change time/div. Press F1 again to switch between enabled channels | Open Probe for the selected Scale channel |
| F2 Position | Select Position. ↑/↓ move the selected channel's zero, ←/→ pan. Press F2 again to switch between enabled channels | Home view by default: center both channel zero levels and reset zoom and pan. Can be set to Reset zoom & pan to keep the channel zero positions |
| F3 Trigger | Select Trigger. Press F3 again to cycle Auto → Normal → Single. ↑/↓ adjust the level and ←/→ adjust pretrigger | Set pretrigger to 50% and set the level to the signal midpoint when a completed capture is available |
| F4 Analyze / Cursor | Open Analyze when cursors are off. With cursors enabled, select a cursor or step to the next one | Open the configured panel, Cursor settings by default |
| MENU | Open the panel for the selected control: Channel from Scale or Position, Trigger from Trigger, Analyze from Cursor | Settings |
| HOLD | Run / Stop | Save a screenshot |
| AUTO | Quick (default): start Auto Setup. Classic: reset pan. In either mode a tap cancels Auto Setup while it is running | Quick: open Zoom & Pan. Classic: start Auto Setup. Hold until the 2 s progress bar completes |
| REL | Cycle the available footer pages | Power off |
Hold MODE for 1 s to open the generator. Hold it again to return to the scope. See Controls for the full MODE cycle.
See Scope shortcuts for the AUTO, F2 and F4 settings. The trigger source is selected in the Trigger panel. If you change trigger mode while stopped, it takes effect on the next RUN and the held record stays on screen.
Pan direction in Settings → Display chooses whether ←/→ move the view or the waveform in time views. In Position and Zoom & Pan, a tap moves the visible window by one division. A held key doubles the step every eight repeats, up to 8×.
In HOLD, use Scale to set the next V/div and time/div. A calibrated held trace is redrawn at the chosen scale with the same samples. If Normal or Single is in WAIT and still showing the previous trace, the Scale card shows the next settings in the title and the displayed scale below it.
In RUN, if acquisition reuses the only record slot (for example, at 512K points on the ZT703S), panning applies to the next frame. In WAIT, panning a previous trace requires its record to still be available.
In Quick, tap AUTO and release to start. In Classic, hold AUTO until the 2 s progress bar completes. Releasing during the progress bar cancels the hold action. Successful Auto Setup resets zoom and pan. While Auto Setup is running, a tap cancels it and restores the previous acquisition settings, view and any held record. Changing V/div or time/div from Scale also cancels Auto Setup, then applies the requested step.
With a captured record in Y-T or Y-T+FFT, hold AUTO in Quick mode to open Zoom & Pan. ↑/↓ change the size of the visible window around the reference point, and ←/→ move it according to Pan direction. The zoom factor can be below 1× (for example, 1/2x). The acquisition timebase stays at its selected value.
- F4 returns to the previous control and keeps your work on the current view. F1, F2 and F3 select their respective controls. MENU opens the previous control's panel.
- The plot continues to show
ZOOMorPANafter you leave the Zoom & Pan controls while the view is adjusted. - A zoom-out step requires the full requested span to fit in the record. If a later capture is shorter than the active zoomed-out view,
ZOOM allindicates that the whole record is visible. - Hold F2 to apply your configured reset. The view returns to 1× at the selected time/div and the horizontal position returns to the trigger reference.
- Zoom and pan changes made while stopped apply to the held record. RUN restores the working view from before HOLD. An explicit F2 reset also updates the working view.
- X-Y, standalone FFT, Roll and the decoder table have their own controls.
Select Scale or Position, then press MENU to open the channel panel. Each channel has enable, DC/AC coupling, V/div, analog zero, inversion, trace brightness, probe and density settings. Hold F1 to open the probe picker directly.
The probe ratios are 1:1, 10:1, 50:1, 100:1, 500:1, 1000:1 and 10000:1. The Position card abbreviates the last one as X10K. The factor scales V/div, trigger levels, measurements and cursors. The hardware input limits still apply.
Trace brightness is adjustable from 10–100% for each channel. Use Analyze → Display → View → Grid brightness to adjust the grid from 0–100%. Both controls step by 10% and default to 100%. Trace brightness also applies to Density, Glow, Persistence, Peak Hold and the XY trace (CH1). Channel labels and trigger markers keep their contrast.
With Trigger selected, press MENU to choose the source, edge, level and mode (Auto / Normal / Single) or change Noise reject (on by default). Analyze → Rearm Single arms the next single shot. The trigger level keeps its voltage when you change V/div or the channel position. If a smaller V/div puts that voltage off the grid, the level waits at the edge and returns when you widen the range again. Set level to 50% on a signal centred on zero gives exactly 0 V. Hold ↑ or ↓ on the Trigger card to accelerate the level step up to 8×.
Pretrigger is the percentage of the record captured before the trigger. With the Trigger card selected, ←/→ change it by one percentage point per press and five per held repeat, within 0–100%. Right reserves more samples before the trigger in either Pan direction setting. In HOLD or after a completed Single, the new value applies to the next acquisition. Roll has its own continuous capture.
From 0.21.4, reaching either trigger level limit pulses the level arrow at the right edge for 0.5 s. A small warning triangle stays next to it until you move away from the limit. A fresh press against the limit repeats the pulse. The Trigger panel's Level row shows (min) or (max) at the corresponding limit.
Center trigger (trigger panel, or Analyze → Display → View) puts the trigger mark in the middle of the grid, clears the pan and sets Pretrigger to 50%. In HOLD it centres the displayed record and prepares the next capture. To adjust the pan numerically, use Horizontal position in View. The value is in thousandths of the record, from −1000 to +1000. ←/→ step by 1, ↑/↓ by 25, and holding ←/→ speeds up to 25.
- Between samples. When the trace is drawn between samples (Y-T or Y-T+FFT, zoomed in), it is placed where the drawn curve crosses the trigger level, between two samples. Fast signals such as 20 MHz at 100 MS/s then stay still from frame to frame. The samples and the exported data are unchanged.
- Auto on a flat signal. Auto accepts an edge only close to the real trigger event. A noise crossing far from it leaves the frame free-running, so a flat trace stays still.
- Hold F4 with the default Cursor settings shortcut. Select the channel and enable Time cursors, Voltage cursors or both.
- Press F4 for Back. The Cursor card opens on Time A, or Level a when only voltage cursors are on.
- Use ←/→ to move time or frequency cursors and ↑/↓ to move level cursors. Tap F4 to select the next enabled line. MENU opens Analyze.
Standalone FFT labels the first frequency cursor Freq A.
When F4 shows Cursor, press MENU from the Cursor card to open Analyze.
| Row | What it does |
|---|---|
| Measurements | Up to four measurements per channel: frequency, period, Vpp, RMS, AC RMS, mean, min, max, duty, pulses per frame and the hardware frequency counter |
| Cursors | Off, Time, Level or Time+Level (Frequency in FFT). Shows Δt, 1/Δt or ΔV |
| Display | Sections as sub-pages, Back returns to the parent menu. Mode (YT or X-Y), samples (Peak / HD mean / Auto), interpolation (linear, sin(x)/x, spline; Sinc by default for a new or reset theme), averaging, persistence (including infinite), grid, MATH, a reference waveform and the frequency counter overlay |
| Layout | Order of the waveform, the measurements and the soft-keys |
| Colors | Color schemes: Current, High contrast, Daylight, Amber, or custom RGB per element |
| Acquisition | Record depth and sample rate |
| Decoder | Protocol Decoders |
| Rearm Single, Save RAW, Save CSV | Actions; files go to /openosc/CAPTURES (Screenshots and Captures) |
| FFT spectrum | Up to 2048 points; Rectangular, Hann, Hamming or Blackman window; dBFS or linear |
| Center channels, Reset pan, Trigger 50% | Quick actions |
| Bode (FRA) | Gain and phase of a circuit against frequency (Bode Analyzer) |
- Persistence (Analyze → Display → History): earlier frames fade in three brightness steps by age and disappear after the chosen time, in Y-T and X-Y. The trail keeps fading while no new frames arrive. HOLD freezes the accumulated view; Infinite keeps the trail until Clear history and reference.
- Density (per-channel Channel panel): Colors maps more frequent hits to different colors, and Mono varies the brightness of a single color. Density color picks that color or uses the channel color. Density works in Y-T and Y-T+FFT. With Persistence off it keeps a 1 s history. Infinite persistence retains the history until you clear it or change acquisition settings that reset it. HOLD freezes it. Use Clear history and reference to clear it manually.
- Glow (Analyze → Display → Capture rendering): a soft halo around the trace in Y-T and X-Y. It costs some speed, for example 39 instead of 46 frames per second.
- Capture CHn reference and Clear history and reference confirm the action with a message. A reference that cannot be taken says why, for example Reference needs a visible Y-T channel.
Digital gain in Analyze → Display → View sets per-channel trace gain, with presets of 1%, 25%, 50%, 75%, then 25% steps from 100% to 6400%. 100% is unity. If a custom theme has an in-between value, stepping snaps to the next preset in the direction you pressed.
The footer shows two measurements per channel by default, frequency and Vpp. With two channels on, tap REL to reach readouts 3–4 if configured. With one channel on, it can show all four readouts on one page. Each value keeps its place, and the unit prefix changes only after a clear change of the value.
-
FC is the live hardware frequency counter. It keeps updating in HOLD.
FCfmeans the count was filtered (bounce or narrow pulses removed);FC?means the count is not confirmed. - f~ / T~ are estimated from the spectrum of the record when there are fewer than eight points per period.
- A value that cannot be measured is shown as missing.
Analyze → Display → Frequency counter shows the hardware counter over the waveform:
| Row | |
|---|---|
| Counter | Off (default), CH1 or CH2 |
| Readings averaged | 1, 3 or 10 settled readings |
| Clock correction | ±1000 ppm, to match the counter to a reference |
The number of digits follows the averaging, up to 7 digits.
With Rolling → Auto in Acquisition and Auto trigger selected, timebases of 200 ms/div and slower run in continuous Roll mode. The newest samples fill the rightmost part of the plot. Completed columns keep their samples as the trace scrolls.
Set the record memory budget in Settings → Device → Performance → Record memory. The overview bar in the top row shows which part of the record is visible.
The sample rate depends on the profile: 100, 140 or 90 MS/s per channel.
Interleave CH1 doubles the CH1 sample rate when CH1 is the only channel: up to 200, 280 or 180 MS/s. It is experimental on the ZT703S; on the ZT706S Auto keeps the built-in rules (Performance Profiles).
OpenOSC is licensed under the GPL-3.0. Found a mistake on this page? Tell us in Discussions.