Settings → Device → Scope calibration adds your own corrections for the oscilloscope. They are stored as files in /openosc/CALIBRATION/ and can be switched off at any time. The device's built-in calibration data is only read, never changed. The multimeter calibrates itself and is not affected.

Row What it does Prepare the inputs
Hardware zero (PWM) Measures and corrects the zero of both channels on every range and position Ground both inputs
Software zero A software zero correction (refused while Hardware zero is in use) Disconnect both inputs
Gain Corrects the scale of one range of one channel against a known DC reference Apply the reference to the probe tip
Interleave offset ZT706S only: the offset between interleaved halves of CH1 —
Reload calibration JSON Re-read the calibration files —
Use factory zero Switch off both zero corrections and use the built-in zero —

Every procedure asks for confirmation first (the cursor starts on Cancel). The generator is switched off during a procedure and stays off.

Zero

Hardware zero checks both channels, all 11 ranges from 20 mV/div to 10 V/div and seven vertical positions, and takes about 15 s. Run it after the device has warmed up, with both inputs grounded, not just open.

Gain

  1. Choose Channel, V/div and Probe.
  2. Apply a stable DC reference to the probe tip and enter its value in Reference at tip. Expected span shows what the measurement expects to see.
  3. Start measurement, then save the result.

Gain in force shows whether a correction is active for that range; Use factory gain switches it off.

Files

scope_zero.json, scope_pwm.json, scope_gain.json and (ZT706S) scope_interleave.json. Each file is bound to this unit's calibration data, so a file from another device is not applied. A previous version is kept as *.previous.json.