Repository navigation
Controls
The same keys work the same way on every screen. Each screen gives F1–F4 its own meaning; the labels above F1–F4 always show what each key does right now.
| Key | Short press | Hold |
|---|---|---|
| MODE | Oscilloscope → Multimeter → Generator → Oscilloscope. Closes an open panel | Hold for 1 s to switch between Oscilloscope and Generator. From Multimeter, go to Oscilloscope |
| MENU | Opens the panel of the current screen; in a panel or menu, takes the selected row | 1 s: Settings |
| HOLD / SAVE | Oscilloscope: Run/Stop. Multimeter: HOLD. Generator: output on/off | 1 s: screenshot of the screen |
| AUTO / RANGE | Oscilloscope Quick: Auto Setup or cancel. Classic: reset pan or cancel. Multimeter: manual range / next range | 2 s: oscilloscope Quick opens Zoom & Pan; Classic starts Auto Setup. Multimeter: auto range |
| REL / Power | Oscilloscope changes the measurements page. Multimeter toggles relative measurement | The "Hold to power off" bar starts at 0.5 s. The device switches off and saves settings at the configured hold time (1, 1.5 or 2 s, default 2 s). Release earlier to cancel |
| F1–F4 | Screen-specific, see each instrument page | In the oscilloscope, F1 opens Probe, F2 resets the view, F3 sets pretrigger to 50% and centres the trigger level on the signal, and F4 opens a configurable panel. Oscilloscope keys |
| Arrows | Change the selected parameter; in panels, move between rows | Repeat (after 0.4 s, every 80 ms) |
A hold never also triggers the short-press action. Keys held while the device starts must be released first. The MODE shortcut is available from OpenOSC 0.21.4 and uses the same panel-closing behavior as a tap. USB Remote locks MODE.
The instrument panels (MENU on a screen) and Settings (hold MENU) work the same way. Each row is a list, a switch, a number, an action or a sub-page.
| You are on | Keys |
|---|---|
| The rows | ↑/↓ select a row. MENU opens a list, flips a switch, starts editing a number, runs an action or opens a sub-page. → also opens a sub-page, selects the next choice or flips a switch. Choices and switches take one step per press. F4 or ← goes back one level |
| An open list | ↑/↓ choose (the current value is marked with <). MENU takes it. F4 or ← closes without a change |
| A number being edited | ←/→ (and ↑/↓) step the value. MENU keeps it. F4 restores the value it had before you started |
When a change takes effect:
- Oscilloscope and generator panels: field by field. A list choice or a switch applies at once; a number applies when you press MENU. Stepping a number does not restart the capture or the generator.
- Multimeter panel: changes apply at once.
- Settings: choices apply when selected, and numbers are kept with MENU. Brightness previews as you step. F4 restores the original value.
- Settings and the theme are written to the storage when you switch the device off, restart it or enter Recovery. A battery pull, a crash or a reset loses the changes made since then.
- If the device refuses a change (for example, switching off the last enabled channel), the row shows the value actually in use and a message explains why.
MODE, entering USB Remote mode or powering off while a number is being edited works like F4: the unfinished edit is dropped.
The top line shows the run state, trigger, the FPS or sample rate (Settings → Display → Top bar counter), record memory, USB and battery. On the oscilloscope the run state reads STOP in HOLD, TRIG'D while frames trigger and AUTO while Auto mode runs free; the multimeter shows LIVE, HOLD and REC while CSV logging runs.
When a menu continues on another page, More ↓ appears under the rows.
OpenOSC is licensed under the GPL-3.0. Found a mistake on this page? Tell us in Discussions.
Getting started
Instruments
Device
- Menu Map
- Settings Menu
- Calibration
- Screenshots and Captures
- Files on the Device
- Performance Profiles
Hardware Mods
PC connection
Help
Contributors