Skip to content

Multimeter

Val edited this page Nov 24, 2026 · 12 revisions

Caution

The USB port is not isolated from the measurement inputs. Be careful when you measure while the device is connected to a computer: see USB and Remote Mode.

DC/AC voltage, millivolts, Type K temperature, DC/AC current, milliamps, resistance, continuity, diode and capacitance, with auto and manual ranges, HOLD modes, REL, statistics, trend and logging.

Keys

Key Action
F1 Voltage group (each press: next function of the group)
F2 Resistance / continuity / diode / capacitance group
F3 Current group
F4 V/Ω jack: DC mV → AC mV → Type K. mA jack: DC mA → AC mA
AUTO / RANGE Short press: manual range, then the next range. Hold 2 s: auto range
↑ / ↓ Change the manual range
← / → Primary / statistics / trend / analog dial view
HOLD Hold, according to the HOLD mode below
REL Use the current reading as the reference; press again to clear
MODE Tap to open the generator. Hold for 1 s to go to the oscilloscope
MENU Panel: range, display, statistics, HOLD, relative, CSV logging, serial output, layout and colors

Each key press sends exactly one command to the meter; the function shown is always the one the meter reports. In diode and continuity, AUTO/RANGE is ignored (no ranges). Diode mode has no beep; a mod of the multimeter chip can add it (Diode Beep).

HOLD modes

Set in MENU → Measurement → Hold:

Mode HOLD press Shows
Manual (default) Freezes the reading; again: live HOLD
Delayed Counts down, then holds the next reading DELAY n, then HOLD
Stable Waits for a stable reading and holds it once WAIT, then HOLD
Auto Holds every new stable reading, again and again A-WAIT, A-HOLD

Options: delay (1–60 s), stable time (0.2–10 s), stable band and rearm threshold in digits of the last displayed position, minimum number of readings. Pressing HOLD again cancels a countdown or releases the held value. Statistics keep running while a value is held.

Display

  • Digital or analog dial view, rearrangeable (MENU → Layout), with custom colors.
  • MENU → Appearance → Preset: Custom (your own look) or Hybrid (dark background, white scale, orange needle, amber digits). Choosing Hybrid keeps your Custom look, and choosing Custom brings it back; changing the look in Hybrid makes that your new Custom look. Reset element and Reset all confirm with a message.
  • With Bar graph and Statistics off (MENU), the reading grows to 72 or 80 px and takes every free line.
  • The function row shows the function's sign; units are spaced ("2.5 kΩ"), and REC appears while CSV logging runs.
  • Bar graph and dial scaled to the meter: 25,000 counts on ZT703S, 66,000 on ZT706S. Analog dial labels sit clear of the arc and tick marks, including on wide Type K ranges.
  • In diode test, the dial and bar use a 0–3.3 V scale on both models from 0.21.6. The scale appears on the first reading or OL.
  • Statistics: MIN, MAX, AVG; reset from the panel. Trend chart, drawn from the first reading.

Type K temperature

  1. Connect a Type K thermocouple to COM and V/Ω.
  2. Press F4 until the function shows the thermometer sign (DC mV → AC mV → Type K).
  3. In MENU → Measurement → Junction source, choose how to estimate the cold-junction temperature.
  4. Temp units picks °C or °F.
Junction source Temperature used
Manual (default) Manual junction, initially 25.0 °C. Enter the temperature where the thermocouple wires meet the copper leads. ←/→ step by 0.1 °C and ↑/↓ by 1 °C, from −50.0 to 100.0 °C
MCU at startup The STM32 die temperature captured when the firmware starts
MCU live The current STM32 die temperature, updated about once a second

The die temperature can differ from the temperature at the terminals as the instrument warms up. The manual value and the chosen source are saved independently.

OpenOSC measures the thermocouple voltage in DC mV, adds the EMF of the cold junction and converts the sum with the NIST ITS-90 tables.

  • HOLD, REL, the trend, MIN/MAX/AVG and CSV logging work in temperature. The CSV value column is in °C; the unit and text columns follow the chosen unit.
  • The dial picks a scale segment that fits the reading: 0–100, 100–300, 300–600, 600–1000 and further, and −100–0, −300–−100 below zero.
  • Changing the unit or the cold junction starts a new period: HOLD, REL and the statistics are reset.
  • The range follows the DC mV range. To change it, go back to DC mV, select the range and switch Type K on again.
  • An open thermocouple is not detected in this mode.

Link to the meter chip

If the meter chip stops answering, OpenOSC restarts its serial link up to 3 times, then power-cycles the chip with growing pauses and selects the function again. Readings taken before the function is active are dropped.

Reading states

Badge Meaning
LIVE Normal reading
NO VALUE The meter shows dashes while it waits or changes range. Left out of statistics
OVERLOAD OL on the display
INVALID A reading that cannot be interpreted
NO DATA / STALE No data from the meter, or old data

Logging

CSV logging in the panel writes readings to /openosc/LOGS/DMMnnnnn.CSV (Screenshots and Captures).

Clone this wiki locally