Oscilloscope

Val edited this page Nov 19, 2026 · 16 revisions

Two channels, 20 mV/div to 10 V/div, 10 ns/div to 20 s/div, with Auto, Normal and Single trigger, measurements, cursors, FFT, XY, MATH and Protocol Decoders.

Keys

Key Action
F1 Scale Arrows change V/div (↑/↓) and time/div (←/→). Press again to switch CH1/CH2
F2 Position ↑/↓ move the channel's zero, ←/→ the horizontal position. Press again to switch CH1/CH2. Hold F2: center both channels
F3 Trigger ↑/↓ move the trigger level. Each press steps the trigger mode Auto → Normal → Single, shown on the trigger card; the source is set in the trigger panel. Hold F3: level to 50 % of the signal
F4 Analyze / Cursor Opens Analyze. With cursors on, F4 steps through the cursor lines and the arrows move the active one; MENU then opens Analyze
MENU Channel panel: enable, coupling, probe, invert, position
HOLD Run / Stop
AUTO Short press: reset the pan. Hold 2 s: Auto Setup for both channels (a bar shows the progress); with no signal it reports this at once

←/→ move the view in steps of one division or one pixel of the shown window; a held key doubles the step every 8 repeats. Settings → Display → Pan direction chooses whether the arrows move the view or the waveform.

In HOLD, ↑/↓ scale the held frame and ←/→ choose the time/div of the next capture; the held frame is redrawn at that scale at once. A held capture stays on screen until you press HOLD again.

Channel panel (MENU)

Enable, coupling (DC/AC), probe (×1, ×10, ×50, ×100, ×500, ×1000), invert and position, per channel. The probe factor is used for V/div, trigger levels, measurements and cursors.

Trigger

Source, edge, level and mode Auto / Normal / Single in the trigger panel, plus Noise reject (on by default). Analyze → Rearm Single arms the next single shot. The trigger position (pre/post-trigger) is adjustable. The trigger level keeps its voltage when you change V/div or the channel position.

Analyze (F4)

Row What it does
Measurements Up to four measurements per channel: frequency, period, Vpp, RMS, AC RMS, mean, min, max, duty, pulses per frame and the hardware frequency counter
Cursors Off, Time, Level or Time+Level (Frequency in FFT). Shows Δt, 1/Δt or ΔV
Display Sections as sub-pages, Back returns to the parent menu. Mode (YT or X-Y), samples (Peak / HD mean / Auto), interpolation (linear, sin(x)/x, spline), averaging, persistence (including infinite), grid, MATH, a reference waveform and the frequency counter overlay
Layout Order of the waveform, the measurements and the soft-keys
Colors Color schemes: Current, High contrast, Daylight, Amber, or custom RGB per element
Acquisition Record depth and sample rate
Decoder Protocol Decoders
Rearm Single, Save RAW, Save CSV Actions; files go to /openosc/CAPTURES (Screenshots and Captures)
FFT spectrum Up to 2048 points; Rectangular, Hann, Hamming or Blackman window; dBFS or linear
Center channels, Reset pan, Trigger 50% Quick actions

Readouts

The footer shows two measurements per channel (by default frequency and Vpp); with one channel on, its readouts fill both panels. Each value keeps its place, and the unit prefix changes only after a clear change of the value.

  • FC is the live hardware frequency counter. It keeps updating in HOLD. FCf means the count was filtered (bounce or narrow pulses removed); FC? means the count is not confirmed.
  • f~ / T~ are estimated from the spectrum of the record when there are fewer than eight points per period.
  • A value that cannot be measured is shown as missing.

Frequency counter overlay

Analyze → Display → Frequency counter shows the hardware counter over the waveform:

Row
Counter Off (default), CH1 or CH2
Readings averaged 1, 3 or 10 settled readings
Clock correction ±1000 ppm, to match the counter to a reference

The number of digits follows the averaging, up to 7 digits.

Roll mode and record depth

Timebases of 200 ms/div and slower run in continuous Roll mode. Record memory is set in Settings → Device → Performance → Record memory. A record always fills the selected depth; a bar at the top shows where the visible part lies.

Performance profile

The sample rate depends on the profile: 100, 140 or 90 MS/s per channel.

On the ZT703S, the experimental Interleave CH1 option doubles the CH1 sample rate at 10 and 25 ns/div when CH1 is the only channel: 200, 280 or 180 MS/s (Performance Profiles).