Settings → Device → Performance → Run mode selects the core clock and the oscilloscope sample rate. The switch is instant; no restart is needed.
| Profile | Core clock | Sample rate per channel | USB data |
|---|---|---|---|
| 200/100 (default) | 200 MHz | 100 MS/s | Yes |
| 280/140 | 280 MHz | 140 MS/s | No |
| 360/90 | 360 MHz | 90 MS/s | No |
- USB (SCPI, Remote mode) works only in 200/100.
- 360/90 runs the microcontroller beyond its rated clock. It is offered for experiments; use it at your own risk.
- If a start at 360/90 is not confirmed, the next start uses 200/100 automatically and shows a message.
- While the generator streams from the DAC, the maximum sample rate is halved.
Settings → Device → Performance → Interleave CH1 samples CH1 with both ADCs in turn and doubles its sample rate. The manufacturer's firmware for the ZT703S does not use interleave; OpenOSC adds it as a new capability of this model. It is Off by default; On: CH1 alone, 10/25 ns applies at once and is saved.
| Profile | Interleaved CH1 |
|---|---|
| 200/100 | 200 MS/s |
| 280/140 | 280 MS/s |
| 360/90 | 180 MS/s |
Interleave is in force while all of these hold:
- CH1 is the only enabled channel. CH2 stays on until you switch it off.
- The timebase is 10 or 25 ns/div.
- Roll is off.
- The generator does not stream from the DAC.
The row shows what is in force:
| Row value | Meaning |
|---|---|
| On, 200 MS/s | Interleave runs, with the measured offset of this profile |
| On, 200 MS/s uncal. | Interleave runs; the offset of this profile has not been measured yet |
| On, waits: CH2 on | A condition is missing: CH2 on, 10/25 ns, Roll, GEN DAC or run mode |
The two ADCs have a small DC offset between them, which shows as a fine comb on the trace. Measure it once in each profile you use: Calibration → Interleave offset (ZT703S). Gain and timing differences between the ADCs stay uncorrected, so fast edges and high frequencies can still show a comb.
Changing the profile keeps the choice and restarts the capture; the offset of the new profile applies. A held capture keeps the rate it was taken with.
The ZT706S has its own built-in interleave, and this row is absent there.
OpenOSC is licensed under the GPL-3.0. Found a mistake on this page? Tell us in Discussions.