Skip to content

Troubleshooting

Val edited this page Nov 19, 2026 · 12 revisions

The device does not start the firmware

Hold F2 while switching on to open Recovery, then install the firmware again (Updating). If Recovery opens by itself, no valid firmware is installed.

The Recovery file disappears from the drive, but nothing is installed

Rename the file to ZOYI-703S-FW-V143.ZTK and copy it again. Some ZT703S units (Boot v1.0.15) install only files with this kind of name (Installation).

"IMAGE REJECTED" in Recovery

The file is for the other model or damaged. Check the model in the file name and the SHA-256 from the release page.

The computer does not see a USB drive

The storage appears as a USB drive in Recovery. The running firmware appears as a USB instrument (USB and Remote Mode).

USB disconnects or is not detected

  • Use the 200/100 profile; USB data does not work in the others.
  • Connect directly to a computer port.
  • After switching between oscilloscope and multimeter, reopen the connection.

Settings were lost after a reset

Settings are saved at an orderly power-off, restart or Recovery entry (Saved at power-off in Settings). Switch the device off with REL/Power before you remove the battery or reset it.

Settings look wrong after editing a file

A file that cannot be read is renamed to .bad; settings then come from the previous version or the copy in /openosc/SETCOPY/. Compare it with the new file, fix it and use Reload settings JSON (Files on the Device). Settings → Device → Restore settings returns to defaults.

The oscilloscope zero is off

Run Hardware zero with both inputs grounded (Calibration). Use factory zero switches your corrections off.

The multimeter shows NO VALUE, OVERLOAD or INVALID

See the reading states on the Multimeter page. NO VALUE during range changes is normal.

The file system is damaged

Recovery → CHECK / REPAIR FAT. Format only as a last resort.

Reporting a bug

Open a bug report with:

  • the model and the version (Settings → Device → About);
  • the profile;
  • what you did, what you expected and what happened;
  • a screenshot (hold HOLD/SAVE), if the problem is on screen;
  • for USB problems: operating system, PC software, hub or not.

Please do not attach your boot block backup; it contains your unit's calibration and is not needed.

Clone this wiki locally