Hold F2 while switching on to open Recovery, then install the firmware again (Updating). If Recovery opens by itself, no valid firmware is installed.
The file is for the other model or damaged. Check the model in the file name and the SHA-256 from the release page.
The storage is shown as a drive only in Recovery. In the running firmware the device is a USB instrument instead (USB and Remote Mode).
- Use the 200/100 profile; USB data does not work in the others.
- Connect directly to a computer port, not through a hub.
- After switching between oscilloscope and multimeter, reopen the connection.
A file that cannot be read is renamed to .bad and replaced with defaults. Compare it with the new file, fix it and use Reload settings JSON (Files on the Device). Settings → Device → Restore settings returns to defaults.
Run Hardware zero with both inputs grounded (Calibration). Use factory zero switches your corrections off.
See the reading states on the Multimeter page. NO VALUE during range changes is normal.
Recovery → CHECK / REPAIR FAT. Format only as a last resort.
Open a bug report with:
- the model and the version (Settings → Device → About);
- the profile;
- what you did, what you expected and what happened;
- a screenshot (hold HOLD/SAVE), if the problem is on screen;
- for USB problems: operating system, PC software, hub or not.
Please do not attach your boot block backup; it contains your unit's calibration and is not needed.
OpenOSC is licensed under the GPL-3.0. Found a mistake on this page? Tell us in Discussions.