This script generates a synthetic slanted-edge image for MTF validation in radiographic imaging systems. The edge is defined on an oversampled grid and binned to detector resolution for accurate pixel-aperture modeling, with optional Gaussian blur. Outputs include both 32-bit floating-point and 16-bit images for standard slanted-edge MTF analysis.
python numpy image-processing medical-imaging image-quality synthetic-data mtf radiography detector-characterisation modulation-transfer-function slanted-edge spatial-frequency-response iso-12233 test-image
-