Galvo rotation and scan lens installation

Setting up the scanner rotation

Excitation path

The scanner should be rotated in its mount so that the beam is on axis if 0V are applied. When doing this step, is important to check again and again that the scanner is truly at 0V. Often, galvo scanners will move to a non-zero position when switched off.

Leveling the holder 30 that will carry the alignment collimator / laser. Excitation path

Connect the Galvo scanner to a source with 0V / short the input to GND: Excitation path

Put the alignment collimator into the holder 30 and stop the beam down to <10 mm diameter using an adjustable iris. Make sure that the beam hits the scan mirror in the center using the manual linear stage. Excitation path

Attach a spare 60 mm cage plate and a few rods to be able to insert a 60 mm alignment target. Rotate the galvo scanner so that the beam is centered on the alignment target. Excitation path

Depending on how you set up the system, do the same for the other excitation path: Excitation path

Mount the Nikon excitation lenses in their 60 mm cage plates

Attach the cage plate to the scanner mount using two ER posts (each composed of an ER2 and an ER05). The posts are secured to the scanner mount using set-screws.

Making the excitation lens mounts parallel to the main rail

Excitation path Excitation path Excitation path This is done using the digital indicator by indicating against the surface of the cage plate. Here, there is only 0.007 mm difference across the length of the cageplate - usually everything below 0.02 mm is ok.